Performance Evaluation of Calypso® 4D Localization and Kilovoltage Image Guidance Systems for Interfraction Motion Management of Prostate Patients

被引:11
作者
Ogunleye, Tomi [1 ]
Rossi, Peter J. [1 ]
Jani, Ashesh B. [1 ]
Fox, Tim [1 ]
Elder, Eric [1 ]
机构
[1] Emory Univ, Sch Med, Dept Radiat Oncol, Atlanta, GA 30322 USA
关键词
Calypso; prostate; OBI; tracking; motion; electromagnetic localization; real-time; RADIOTHERAPY; CANCER;
D O I
10.1100/tsw.2009.61
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Prostate cancer represents a model site for advances in understanding inter- and intrafraction motion for radiotherapy. In this study, we examined the correlation of the electromagnetic transponder system/Calypso (R) 4D Localization System with conventional on-board imaging (OBI) using kilovoltage imaging. Initially using a quality assurance (QA) phantom and subsequently using data of seven patients, the vector distances between Calypso- and OBI-recorded shifts were compared using the t-test. For the 30 phantom measurements, the average differences between the measured Calypso offset and the calculated OBI shift were 0.4 +/- 0.4, 0.2 +/- 0.3, and 0.4 +/- 0.3 mm in the lateral, longitudinal, and vertical directions, respectively (p = 0.73, p = 0.91, and p = 0.99, respectively), and the average difference vector for all sessions was 0.8 +/- 0.4 mm. For the 259 patient measurements, the average differences between the measured Calypso offset and the calculated OBI shift were 0.7 +/- 0.5, 1.1 +/- 0.9, and 1.2 +/- 0.9 mm in the lateral, longitudinal, and vertical directions, respectively (p = 0.45, p = 0.28, and p = 0.56, respectively), and the average difference vector for all sessions was 2.1 +/- 1.0 mm. Our results demonstrated good correlation between Calypso and OBI. While other studies have explored the issue of Calypso/OBI correlation, our analysis is unique in our use of phantom validation and in our performing the patient analysis on an initial population prior to routine setup using Calypso without OBI. Implications for Calypso's role as a QA tool are discussed.
引用
收藏
页码:449 / 458
页数:10
相关论文
共 14 条
[1]   Marker based IGRT in prostate radiotherapy - Merits and pitfalls [J].
Ashish, K. ;
Bansal, V. ;
Patel, V. ;
Bhavsar, D. ;
Jani, K. ;
Bhonde, M. ;
Gajar, S. ;
Abraham, A. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01) :S353-S353
[2]  
*CAL MED TECHN INC, 2008, LBL0008004 CAL MED T
[3]  
CHRISTOPHER FS, 2006, INT J RADIAT ONCOL, V65, P1585
[4]  
Fox Timothy H, 2006, J Am Coll Radiol, V3, P38, DOI 10.1016/j.jacr.2005.08.019
[5]   In-vivo comparison of an electromagnetic system to standard kV X-rays for treatment setup during external beam radiotherapy of patients with prostate cancer [J].
Jani, S. ;
Weinstein, G. ;
Kupelian, P. ;
Willoughby, T. ;
Enke, C. ;
Solberg, T. ;
Flores, N. ;
Liu, D. ;
Mahadevan, A. ;
Djemil, T. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03) :S78-S78
[6]  
JOHN MS, 2005, INT J RADIAT ONCOL, V63, P800
[7]   Multi-institutional clinical experience with the Calypso System in localization and continuous, real-time monitoring of the prostate gland during external radiotherapy [J].
Kupelian, Patrick ;
Willoughby, Twyla ;
Mahadevan, Arul ;
Djemil, Toufik ;
Weinstein, Geoffrey ;
Jani, Shirish ;
Enke, Charles ;
Solberg, Timothy ;
Flores, Nicholas ;
Liu, David ;
Beyer, David ;
Levine, Lisa .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (04) :1088-1098
[8]  
LAKSHMI S, 2008, INT J RADIAT ONCOL, V70, P892
[9]   Quantifying respiratory-induced prostate motion using continuous real-time tracking technology [J].
McDonald, T. L. ;
Ku, L. ;
O'Donnell, K. M. ;
Kaurin, D. ;
Gagnon, P. J. ;
Thomas, C. R. ;
Hung, A. Y. ;
Fuss, M. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01) :S334-S334
[10]   Clinically useful observations of prostate motion using the Calypso four-dimensional localization system [J].
Shelton, J. ;
Ogunleye, T. ;
Jani, A. B. ;
Elder, E. ;
Rossi, P. J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01) :S299-S300