Spontaneous disinsertion of a haptic from a 3-piece acrylic foldable intraocular lens after surgery

被引:7
作者
Fujishima, K
Yoshitomi, F
Oshika, T
机构
[1] Univ Tsukuba, Inst Clin Med, Dept Ophthalmol, Tsukuba, Ibaraki 3058575, Japan
[2] Yoshitomi Eye Ctr, Fujishima Eye Clin, Fukuoka, Japan
关键词
D O I
10.1016/S0886-3350(02)01219-1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
A 49-year-old man with cataracts had uneventful phacoemulsification, continuous curvilinear capsulorhexis, and implantation of an MA30BA acrylic foldable intraocular lens (IOL) in the bag. One week later, 1 haptic spontaneously dropped off and the optic dislocated into the anterior chamber. The IOL was surgically removed and replaced with a new acrylic foldable IOL. Examination of the explanted IOL revealed that the loop hole was not created in the middle of the side face, leading to in vivo detachment of the haptic postoperatively.
引用
收藏
页码:1296 / 1298
页数:3
相关论文
共 17 条
[11]   BROKEN SEMIFLEXIBLE INTRAOCULAR-LENS IMPLANT [J].
MAGUEN, E ;
NESBURN, AB ;
JACKMAN, N ;
MACY, JI .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1985, 99 (02) :170-172
[12]   Alterations to the AcrySof intraocular lens during folding [J].
Milazzo, S ;
Turut, P ;
Blin, H .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 :1351-1354
[13]   Effect of folding on the optical quality of soft acrylic intraocular lenses [J].
Oshika, T ;
Shiokawa, Y .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 :1360-1364
[14]   INVIVO FRACTURE OF AN EXTRUDED POLYMETHYLMETHACRYLATE INTRAOCULAR-LENS LOOP [J].
PARK, SB ;
KRATZ, RP ;
OLSON, PF ;
PFEFFER, BR ;
APPLE, DJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1987, 13 (02) :194-197
[15]   Stress fractures after folding an acrylic intraocular lens [J].
Pfister, DR .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1996, 121 (05) :572-574
[16]  
SHAMMAS HJF, 1981, AM INTRAOCULAR IMPLA, V7, P46
[17]   Forceps-induced scratching of a foldable acrylic intraocular lens [J].
Vrabec, MP ;
Syverud, JC ;
Burgess, CJ .
ARCHIVES OF OPHTHALMOLOGY, 1996, 114 (06) :777-777