Spontaneous disinsertion of a haptic from a 3-piece acrylic foldable intraocular lens after surgery

被引:7
作者
Fujishima, K
Yoshitomi, F
Oshika, T
机构
[1] Univ Tsukuba, Inst Clin Med, Dept Ophthalmol, Tsukuba, Ibaraki 3058575, Japan
[2] Yoshitomi Eye Ctr, Fujishima Eye Clin, Fukuoka, Japan
关键词
D O I
10.1016/S0886-3350(02)01219-1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
A 49-year-old man with cataracts had uneventful phacoemulsification, continuous curvilinear capsulorhexis, and implantation of an MA30BA acrylic foldable intraocular lens (IOL) in the bag. One week later, 1 haptic spontaneously dropped off and the optic dislocated into the anterior chamber. The IOL was surgically removed and replaced with a new acrylic foldable IOL. Examination of the explanted IOL revealed that the loop hole was not created in the middle of the side face, leading to in vivo detachment of the haptic postoperatively.
引用
收藏
页码:1296 / 1298
页数:3
相关论文
共 17 条
[1]   Damage of foldable intraocular lenses by incorrect folder forceps [J].
Baldeschi, L ;
Rizzo, S ;
Nardi, M .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1997, 124 (02) :245-247
[2]   Traumatic fracture of flexible anterior chamber intraocular lenses [J].
Capoferri, C ;
Vacchi, S ;
Tafi, A ;
Voglini, IM .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1997, 23 (09) :1418-1420
[3]  
Carlson KH, 1995, OPHTHALMIC SURG LAS, V26, P572
[4]   BROKEN SUPERIOR HAPTIC [J].
COHEN, KL .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1989, 15 (02) :233-233
[5]   SPONTANEOUS INVIVO FRACTURE OF AN ANTERIOR-CHAMBER INTRAOCULAR-LENS LOOP [J].
CRAIG, MT ;
MAMALIS, N ;
CRANDALL, AS ;
MCDONALD, DK .
ARCHIVES OF OPHTHALMOLOGY, 1989, 107 (05) :642-643
[7]   Cracked cartridges during foldable intraocular lens implantation [J].
Habib, NE ;
Singh, J ;
Adams, AD ;
Bartholomew, RS .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 (05) :630-632
[8]  
Heremans L, 1998, Bull Soc Belge Ophtalmol, V270, P13
[9]   BROKEN INTRAOCULAR-LENS DURING CATARACT-SURGERY [J].
KIRKPATRICK, JNP ;
COOK, SD .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1992, 76 (08) :509-509
[10]   Cracked acrylic intraocular lens requiring explantation [J].
Lee, GA ;
DalPra, ML .
AUSTRALIAN AND NEW ZEALAND JOURNAL OF OPHTHALMOLOGY, 1997, 25 (01) :71-73