DETERMINATION OF ACYCLOVIR IN RABBIT PLASMA BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHIC (HPLC) TECHNIQUE

被引:3
|
作者
Malik, Nadia Shamshad [1 ]
Ahmads, Mahmood [2 ]
Minhas, Muhammad Usman [3 ]
Khalid, Qandeel [3 ]
机构
[1] Capital Univ Sci & Technol, Dept Pharm, Islamabad, Pakistan
[2] Univ Cent Punjab, Fac Pharm, Lahore, Pakistan
[3] Islamia Univ Bahawalpur, Fac Pharm & Alternat Med, Bahawalpur, Pakistan
来源
ACTA POLONIAE PHARMACEUTICA | 2019年 / 76卷 / 03期
关键词
reversed-phase high-performance liquid chromatography; acyclovir; plasma; SOLID-PHASE EXTRACTION; GANCICLOVIR; ASSAY; VALIDATION; METABOLITE; COLUMN; SERUM;
D O I
10.32383/appdr/100388
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
A rapid, sensitive and simple reversed-phase high-performance liquid chromatographic (HPLC) method has been developed and validated for the determination of acyclovir (ACV) in rabbit plasma. BDS C18 column was used to conduct analysis using ammonium dihydrogen phosphate buffer (50 mM) and methanol as mobile phase (98 : 2), with pH adjusted to 2.5 using orthophosphoric acid. The flow rate was kept at 1 mL/min. Selective precipitation of plasma proteins was done by adding 5% perchloric acid. Precipitated plasma proteins were separated by centrifugation. ACV moves in a supernatant, which was snapped and passed through a syringe filtration assembly. Direct injection of supernatant was given into a BDS C18 column and ACV was detected at 256 nm. The limit of detection for ACV in plasma was estimated as 15 ng/mL whereas the limit of quantitation was calculated as 25 ng/mL. Moreover, the developed method has been found to be selective and linear into concentration range of 25-2000 ng/mL. The present method could be successfully applied to samples from bioavailability and bioequivalence studies.
引用
收藏
页码:421 / 429
页数:9
相关论文
共 50 条
  • [1] A new validated ultra performance liquid chromatographic method for determination of acyclovir
    Hasan, S.
    Chander, P.
    Ali, J.
    Baboota, S.
    Ali, M.
    DRUG TESTING AND ANALYSIS, 2011, 3 (03) : 187 - 190
  • [2] High performance liquid chromatographic determination of fluvoxamine and paroxetine in plasma
    Skibinski, R
    Misztal, G
    Olajossy, M
    CHEMIA ANALITYCZNA, 2000, 45 (06): : 815 - 823
  • [3] Technique validation by liquid chromatography for the determination of acyclovir in plasma
    Fernández, M
    Sepúlveda, J
    Aránguiz, T
    von Plessing, C
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 791 (1-2): : 357 - 363
  • [4] Determination of acyclovir in horse plasma and body fluids by high-performance liquid chromatography combined with fluorescence detection and heated electrospray ionization tandem mass spectrometry
    Maes, A.
    Garre, B.
    Desmet, N.
    van der Meulen, K.
    Nauwynck, H.
    De Backer, P.
    Croubels, S.
    BIOMEDICAL CHROMATOGRAPHY, 2009, 23 (02) : 132 - 140
  • [5] Robust and Sensitive High-Performance Liquid Chromatographic-UV Detection Technique for the Determination of Tigecycline in Rabbit Plasma
    Zorpas, Kostas M.
    Valsami, Georgia N.
    Vryonis, Evangelos V.
    Skoutelis, Athanasios T.
    Archontaki, Helen A.
    JOURNAL OF AOAC INTERNATIONAL, 2011, 94 (03) : 847 - 856
  • [6] Determination of acyclovir in dog plasma by HPLC using a column switching technique
    K. Balogh Nemes
    B. Dalmadi Kiss
    I. Klebovich
    Chromatographia, 2000, 51 : S211 - S216
  • [7] Determination of acyclovir in dog plasma by HPLC using a column switching technique
    Nemes, KB
    Kiss, BD
    Klebovich, I
    CHROMATOGRAPHIA, 2000, 51 (1) : S211 - S216
  • [8] Simple high-performance liquid chromatographic method for the determination of acyclovir in human plasma using fluorescence detection
    Peh, KK
    Yuen, KH
    JOURNAL OF CHROMATOGRAPHY B, 1997, 693 (01): : 241 - 244
  • [9] A simple high-performance liquid chromatographic method for the determination of acyclovir in human plasma and application to a pharmacokinetic study
    Yu, Liyan
    Xiang, Bingren
    Zhan, Ying
    ARZNEIMITTEL-FORSCHUNG-DRUG RESEARCH, 2008, 58 (04): : 199 - 202
  • [10] Validated high performance liquid chromatography (HPLC) method for the determination of sumatriptan in rabbit plasma: Application to pharmacokinetic study
    Sheshala, Ravi
    Khan, Nurzalina
    Darwis, Yusrida
    AFRICAN JOURNAL OF PHARMACY AND PHARMACOLOGY, 2012, 6 (02): : 98 - 107