Wideband 2-D Array Design Optimization With Fabrication Constraints for 3-D US Imaging

被引:37
|
作者
Roux, Emmanuel [1 ,2 ,3 ]
Ramalli, Alessandro [1 ]
Liebgott, Herve [2 ,3 ]
Cachard, Christian [2 ,3 ]
Robini, Marc C. [2 ,3 ]
Tortoli, Piero [1 ]
机构
[1] Univ Florence, Dept Informat Engn, I-50139 Florence, Italy
[2] Univ Lyon, INSA Lyon, CREATIS Lab, F-69007 Lyon, France
[3] Univ Claude Bernard Lyon 1, F-69100 Villeurbanne, France
关键词
COLUMN-ADDRESSED ARRAYS; 3D ULTRASOUND; SYSTEM; TRANSDUCER; ALGORITHM; ELEMENTS; PROBE;
D O I
10.1109/TUFFC.2016.2614776
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Ultrasound (US) 2-D arrays are of increasing interest due to their electronic steering capability to investigate 3-D regions without requiring any probe movement. These arrays are typically populated by thousands of elements that, ideally, should be individually driven by the companion scanner. Since this is not convenient, the so-called microbeamforming methods, yielding a prebeamforming stage performed in the probe handle by suitable custom integrated circuits, have so far been implemented in a few commercial high-end scanners. A possible approach to implement relatively cheap and efficient 3-D US imaging systems is using 2-D sparse arrays in which a limited number of elements can be coupled to an equal number of independent transmit/receive channels. In order to obtain US beams with adequate characteristics all over the investigated volume, the layout of such arrays must be carefully designed. This paper provides guidelines to design, by using simulated annealing optimization, 2-D sparse arrays capable of fitting specific applications or fabrication/implementation constraints. In particular, an original energy function based on multidepth 3-D analysis of the beam pattern is also exploited. A tutorial example is given, addressed to find the Ne elements that should be activated in a 2-D fully populated array to yield efficient acoustic radiating performance over the entire volume. The proposed method is applied to a 32x32 array centered at 3 MHz to select the 128, 192, and 256 elements that provide the best acoustic performance. It is shown that the 256-element optimized array yields sidelobe levels even lower (by 5.7 dB) than that of the reference 716-element circular and (by 10.3 dB) than that of the reference 1024-element array.
引用
收藏
页码:108 / 125
页数:18
相关论文
共 50 条
  • [1] 3-D elasticity imaging with a 2-D array
    Fisher, T.
    Hall, T.
    Panda, S.
    Jiang, J.
    Resnick, I.
    Barnes, S.
    MEDICAL PHYSICS, 2007, 34 (06) : 2366 - 2366
  • [2] Sparse 2-D Arrays for 3-D phased array imaging - Design methods
    Austeng, A
    Holm, S
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2002, 49 (08) : 1073 - 1086
  • [3] ADS-Based Array Design for 2-D and 3-D Ultrasound Imaging
    Oliveri, Giacomo
    Massa, Andrea
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2010, 57 (07) : 1568 - 1582
  • [4] Theoretical assessment of a crossed electrode 2-D array for 3-D imaging
    Morton, CE
    Lockwood, GR
    2003 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2003, : 968 - 971
  • [5] 3-D waveguides created by 2-D fabrication
    Bains, S
    LASER FOCUS WORLD, 2004, 40 (08): : 17 - +
  • [6] Wideband 2-D sparse array optimization combined with multiline reception for real-time 3-D medical ultrasound
    Sciallero, Claudia
    Trucco, Andrea
    ULTRASONICS, 2021, 111
  • [7] 3-D Super-Resolution Ultrasound Imaging With a 2-D Sparse Array
    Harput, Sevan
    Tortoli, Piero
    Eckersley, Robert J.
    Dunsby, Chris
    Tang, Meng-Xing
    Christensen-Jeffries, Kirsten
    Ramalli, Alessandro
    Brown, Jemma
    Zhu, Jiaqi
    Zhang, Ge
    Leow, Chee Hau
    Toulemonde, Matthieu
    Boni, Enrico
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2020, 67 (02) : 269 - 277
  • [8] Sparse 2-D arrays for 3-D phased array imaging - Experimental validation
    Austeng, A
    Holm, S
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2002, 49 (08) : 1087 - 1093
  • [9] Minimally Redundant 2-D Array Designs for 3-D Medical Ultrasound Imaging
    Karaman, Mustafa
    Wygant, Ira O.
    Oralkan, Oemer
    Khuri-Yakub, Butrus T.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2009, 28 (07) : 1051 - 1061
  • [10] Topology gradient optimization in 2-D and 3-D for the design of microwave components
    Khalil, Hassan
    Assadihaghi, Atousa
    Bila, Stephane
    Baillargeat, Dominique
    Aubourg, M.
    Verdeyme, Serge
    Puech, Jerome
    Lapierre, Luc
    MICROWAVE AND OPTICAL TECHNOLOGY LETTERS, 2008, 50 (10) : 2739 - 2743