Validation of creatinine assays utilizing HPLC and IDMS traceable standards in sera of children

被引:38
作者
Schwartz, George J. [1 ]
Kwong, Tai [1 ]
Erway, Brian [1 ]
Warady, Bradley [2 ]
Sokoll, Lori [3 ]
Hellerstein, Stanley [2 ]
Dharnidharka, Vikas [4 ]
Furth, Susan [3 ]
Munoz, Alvaro [5 ]
机构
[1] Univ Rochester, Med Ctr, Rochester, NY 14642 USA
[2] Childrens Mercy Hosp, Kansas City, MO 64108 USA
[3] Johns Hopkins Med Inst, Baltimore, MD 21205 USA
[4] Univ Florida, Gainesville, FL USA
[5] Johns Hopkins Bloomberg Sch Publ Hlth, Baltimore, MD USA
关键词
High-performance liquid chromatography; Glomerular filtration rate; Enzymatic creatinine assay; Jaffe creatinine assay; Isotope dilution mass spectroscopy; Pediatric standards; PERFORMANCE LIQUID-CHROMATOGRAPHY; GLOMERULAR-FILTRATION-RATE; PLASMA CREATININE; KIDNEY-DISEASE; ADOLESCENTS;
D O I
10.1007/s00467-008-0957-0
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
The purpose of this study was to validate serum creatinine (SCr) concentrations assayed in the Central Biochemistry Laboratory of the National Institutes of Health (NIH)-funded Chronic Kidney Disease in Children (CKiD) study utilizing an enzymatic assay (Siemens Advia 2400) against a method traceable to reference isotope dilution mass spectroscopy (IDMS) developed by the National Institute of Standards and Technology (NIST). High-performance liquid chromatography (HPLC) measured SCr after external validation utilizing IDMS-based standard reference materials. Sera from the first 201 subjects enrolled in CKiD were analyzed and compared for creatinine concentration by enzymatic and HPLC methods. Fifty "normal" pediatric sera were subsequently analyzed. Finally, a "pediatric" reference standard was prepared and examined for accuracy and precision. Enzymatic SCr concentrations (median 1.4 mg/dl) of CKiD subjects were well correlated with HPLC (r=0.984) but were slightly higher (+ 7%; p<0.001). Agreement was poorer at lower SCr (median 0.4 mg/dl) when using samples from normal children and the "pediatric" reference standard. However, the Roche enzymatic assay was comparable with HPLC in accuracy and precision. Referring physicians should be aware of the accuracy and reproducibility of their laboratory's SCr assay. Our enzymatic assay agreed well with HPLC in CKiD subjects with elevated SCr. We suggest that NIST develop a pediatric SCr standard reference material for use by assay manufacturers to improve accuracy and precision of assays at the low SCr levels observed in most pediatric patients.
引用
收藏
页码:113 / 119
页数:7
相关论文
共 10 条
  • [1] AMBROSE RT, 1983, CLIN CHEM, V29, P256
  • [2] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [3] National Kidney Foundation's Kidney Disease Outcomes Quality Initiative clinical practice guidelines for chronic kidney disease in children and adolescents: Evaluation, classification, and stratification
    Hogg, RJ
    Furth, S
    Lemley, KV
    Portman, R
    Schwartz, GJ
    Coresh, J
    Balk, E
    Lau, J
    Levin, A
    Kausz, AT
    Eknoyan, G
    Levey, AS
    [J]. PEDIATRICS, 2003, 111 (06) : 1416 - 1421
  • [4] CREATININE .12. COMPARISON OF ASSAYS OF LOW SERUM CREATININE LEVELS USING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND 2 PICRATE METHODS
    HUANG, YC
    CHIOU, WL
    [J]. JOURNAL OF PHARMACEUTICAL SCIENCES, 1983, 72 (07) : 836 - 837
  • [5] Plasma creatinine determination in mice and rats: An enzymatic method compares favorably with a high-performance liquid chromatography assay
    Keppler, A.
    Gretz, N.
    Schmidt, R.
    Kloetzer, H-M
    Groene, H-J
    Lelongt, B.
    Meyer, M.
    Sadick, M.
    Pill, J.
    [J]. KIDNEY INTERNATIONAL, 2007, 71 (01) : 74 - 78
  • [6] A more accurate method to estimate glomerular filtration rate from serum creatinine: A new prediction equation
    Levey, AS
    Bosch, JP
    Lewis, JB
    Greene, T
    Rogers, N
    Roth, D
    [J]. ANNALS OF INTERNAL MEDICINE, 1999, 130 (06) : 461 - +
  • [7] Why commutability matters
    Miller, WG
    Myers, GL
    Rej, R
    [J]. CLINICAL CHEMISTRY, 2006, 52 (04) : 553 - 554
  • [8] Recommendations for improving serum creatinine measurement: A report from the laboratory working group of the National Kidney Disease Education Program
    Myers, GL
    Miller, WG
    Coresh, J
    Fleming, J
    Greenberg, N
    Greene, T
    Hostetter, T
    Levey, AS
    Panteghini, M
    Welch, M
    Eckfeldt, JH
    [J]. CLINICAL CHEMISTRY, 2006, 52 (01) : 5 - 18
  • [9] PLASMA CREATININE AND UREA CONCENTRATION IN CHILDREN - NORMAL VALUES FOR AGE AND SEX
    SCHWARTZ, GJ
    HAYCOCK, GB
    CHIR, B
    SPITZER, A
    [J]. JOURNAL OF PEDIATRICS, 1976, 88 (05) : 828 - 830
  • [10] SCHWARTZ GJ, 1987, PEDIATR CLIN N AM, V34, P571