Investigation on Passive and Autonomous Mode Operation of Floating Gate Dosimeters

被引:9
作者
Brucoli, M. [1 ,2 ]
Cesari, J. [3 ]
Danzeca, S. [1 ]
Brugger, M. [1 ]
Masi, A. [1 ]
Pineda, A. [3 ]
Dusseau, L. [2 ]
Wrobel, F. [2 ]
机构
[1] European Org Nucl Res CERN, CH-1211 Geneva 23, Switzerland
[2] Univ Montpellier, CNRS, IES UMR 5214, F-34097 Montpellier, France
[3] IC Malaga, Alaro 07340, Spain
关键词
Dosimeter; floating gate dosimeter (FGDOS); ionizing radiation sensors; passive dosimeter; radiation monitoring; MOSFET DOSIMETER; SPACE;
D O I
10.1109/TNS.2019.2895366
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In the context of achieving an efficient radiation monitoring system, while also aiming to increase the flexibility of the measurement system, the integrated digital and analog circuitries of the floating gate dosimeter (FGDOS) have been upgraded. The capability to control autonomously the recharge process and the easy access to the data allows exploiting the FGDOS in both the autonomous and passive mode. The former provides the cumulated dose measurement by a sensor coupled to a digital interface. The latter implies the reading of the dose only after the irradiation time, performing the dose detection by positioning the sensor without any supply. In this paper, the functionality and characterization of the new features of the FGDOS are demonstrated. The resolution, the dose range, and the error introduced on the measurements are discussed in detail by analyzing the results of irradiation experiments. The use of a filtering technique in different radiation environments is discussed, as well as the capability to retain the information after multiple readings and the power consumption for each configuration.
引用
收藏
页码:1620 / 1627
页数:8
相关论文
共 15 条
  • [1] Total Ionizing Dose Characterization of a Prototype Floating Gate MOSFET Dosimeter for Space Applications
    Alvarez, Maite
    Hernando, Carlos
    Cesari, Joan
    Pineda, Alvaro
    Garcia-Moreno, Eugeni
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (06) : 4281 - 4288
  • [2] Floating Gate Dosimeter Suitability for Accelerator-Like Environments
    Brucoli, M.
    Danzeca, S.
    Brugger, M.
    Masi, A.
    Pineda, A.
    Cesari, J.
    Dusseau, L.
    Wrobel, F.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2017, 64 (08) : 2054 - 2060
  • [3] Brucoli M., 2018, RADECS C GOTH SWED S
  • [4] Brucoli M., 2017, RADECS C GEN SWITZ O
  • [5] Characterization and Modeling of a Floating Gate Dosimeter with Gamma and Protons at Various Energies
    Danzeca, S.
    Cesari, J.
    Brugger, M.
    Dusseau, L.
    Masi, A.
    Pineda, A.
    Spiezia, G.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2014, 61 (06) : 3451 - 3457
  • [6] Floating Gate sensor for in-vivo dosimetry in radiation therapies. Design and first characterization.
    Faigon, A.
    Martinez Vazquez, I.
    Carbonetto, S.
    Garcia Inza, M.
    [J]. VIII INTERNATIONAL CONGRESS OF ENGINEERING PHYSICS, 2017, 792
  • [7] Radiation Sensor Compatible With Standard CMOS Technology
    Garcia Moreno, E.
    Picos, R.
    Isern, E.
    Roca, M.
    Bota, S.
    Suenaga, K.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (05) : 2910 - 2915
  • [8] Floating Gate CMOS Dosimeter With Frequency Output
    Garcia-Moreno, E.
    Isern, E.
    Roca, M.
    Picos, R.
    Font, J.
    Cesari, J.
    Pineda, A.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (02) : 373 - 378
  • [9] ANALYSIS OF THE EFFECTS OF CONSTANT-CURRENT FOWLER-NORDHEIM-TUNNELING INJECTION WITH CHARGE TRAPPING INSIDE THE POTENTIAL BARRIER
    LOPEZVILLANUEVA, JA
    JIMENEZTEJADA, JA
    CARTUJO, P
    BAUSELLS, J
    CARCELLER, JE
    [J]. JOURNAL OF APPLIED PHYSICS, 1991, 70 (07) : 3712 - 3720
  • [10] CHARM: A Mixed Field Facility at CERN for Radiation Tests in Ground, Atmospheric, Space and Accelerator Representative Environments
    Mekki, J.
    Brugger, M.
    Alia, R. G.
    Thornton, A.
    Mota, N. C. Dos Santos
    Danzeca, S.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2016, 63 (04) : 2106 - 2114