An ultra-sensitive and wideband magnetometer based on a superconducting quantum interference device

被引:83
作者
Storm, Jan-Hendrik [1 ]
Hoemmen, Peter [1 ]
Drung, Dietmar [1 ]
Koerber, Rainer [1 ]
机构
[1] Phys Tech Bundesanstalt, D-10587 Berlin, Germany
关键词
TO-NOISE RATIO; MRI; FIELDS;
D O I
10.1063/1.4976823
中图分类号
O59 [应用物理学];
学科分类号
摘要
The magnetic field noise in superconducting quantum interference devices (SQUIDs) used for biomagnetic research such as magnetoencephalography or ultra-low-field nuclear magnetic resonance is usually limited by instrumental dewar noise. We constructed a wideband, ultra-low noise system with a 45 mm diameter superconducting pick-up coil inductively coupled to a current sensor SQUID. Thermal noise in the liquid helium dewar is minimized by using aluminized polyester fabric as superinsulation and aluminum oxide strips as heat shields. With a magnetometer pickup coil in the center of the Berlin magnetically shielded room 2 (BMSR2), a noise level of around 150 aT Hz(-1/2) is achieved in the white noise regime between about 20 kHz and the system bandwidth of about 2.5 MHz. At lower frequencies, the resolution is limited by magnetic field noise arising from the walls of the shielded room. Modeling the BMSR2 as a closed cube with continuous l-metal walls, we can quantitatively reproduce its measured field noise. Published by AIP Publishing.
引用
收藏
页数:4
相关论文
共 20 条
[1]   SAMPLING THEORY FOR NEUROMAGNETIC DETECTOR ARRAYS [J].
AHONEN, AI ;
HAMALAINEN, MS ;
ILMONIEMI, RJ ;
KAJOLA, MJ ;
KNUUTILA, JET ;
SIMOLA, JT ;
VILKMAN, VA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1993, 40 (09) :859-869
[2]   Fetal magnetocardiography measurements with an array of microfabricated optically pumped magnetometers [J].
Alem, Orang ;
Sander, Tilmann H. ;
Mhaskar, Rahul ;
LeBlanc, John ;
Eswaran, Hari ;
Steinhoff, Uwe ;
Okada, Yoshio ;
Kitching, John ;
Trahms, Lutz ;
Knappe, Svenja .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (12) :4797-4811
[3]  
Bork Hahlbohm J. H. D., 2000, P 12 INT C BIOM, P970
[4]   IRREVERSIBILITY AND GENERALIZED NOISE [J].
CALLEN, HB ;
WELTON, TA .
PHYSICAL REVIEW, 1951, 83 (01) :34-40
[5]  
Clarke J., 2004, SQUID HDB, V1
[6]   SQUID-detected magnetic resonance imaging in microtesla fields [J].
Clarke, John ;
Hatridge, Michael ;
Moessle, Michael .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2007, 9 :389-413
[7]   Ultrahigh sensitivity magnetic field and magnetization measurements with an atomic magnetometer [J].
Dang, H. B. ;
Maloof, A. C. ;
Romalis, M. V. .
APPLIED PHYSICS LETTERS, 2010, 97 (15)
[8]   Highly sensitive and easy-to-use SQUID sensors [J].
Drung, D. ;
Assmann, C. ;
Beyer, J. ;
Kirste, A. ;
Peters, M. ;
Ruede, F. ;
Schurig, Th. .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2007, 17 (02) :699-704
[9]   Ultra-low-noise EEG/MEG systems enable bimodal non-invasive detection of spike-like human somatosensory evoked responses at 1 kHz [J].
Fedele, T. ;
Scheer, H. J. ;
Burghoff, M. ;
Curio, G. ;
Koerber, R. .
PHYSIOLOGICAL MEASUREMENT, 2015, 36 (02) :357-368
[10]   SIGNAL-TO-NOISE RATIO OF NUCLEAR MAGNETIC-RESONANCE EXPERIMENT [J].
HOULT, DI ;
RICHARDS, RE .
JOURNAL OF MAGNETIC RESONANCE, 1976, 24 (01) :71-85