Iliosacral screw insertion using CT-3D-fluoroscopy matching navigation

被引:60
作者
Takao, Masaki [1 ]
Nishii, Takashi [2 ]
Sakai, Takashi [1 ]
Yoshikawa, Hideki [1 ]
Sugano, Nobuhiko [2 ]
机构
[1] Osaka Univ, Grad Sch Med, Dept Orthopaed Surg, Suita, Osaka 5650871, Japan
[2] Osaka Univ, Grad Sch Med, Dept Orthopaed Med Engn, Suita, Osaka 5650871, Japan
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2014年 / 45卷 / 06期
基金
日本学术振兴会;
关键词
Iliosacral screw; Navigation; 3D fluoroscopy; CT-based plan; SACROILIAC SCREWS; PLACEMENT; FLUOROSCOPY; FIXATION; SYSTEM; FRACTURES; ACCURACY; REGIONS;
D O I
10.1016/j.injury.2014.01.015
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Percutaneous iliosacral screw insertion requires substantial experience and detailed anatomical knowledge to find the proper entry point and trajectory even with the use of a navigation system. Our hypothesis was that three-dimensional (3D) fluoroscopic navigation combined with a preoperative computed tomography (CT)-based plan could enable surgeons to perform safe and reliable iliosacral screw insertion. The purpose of the current study is two-fold: (1) to demonstrate the navigation accuracy for sacral fractures and sacroiliac dislocations on widely displaced cadaveric pelves; and (2) to report the technical and clinical aspects of percutaneous iliosacral screw insertion using the CT-3D-fluoroscopy matching navigation system. Methods: We simulated three types of posterior pelvic ring disruptions with vertical displacements of 0, 1, 2 and 3 cm using cadaveric pelvic rings. A total of six fiducial markers were fixed to the anterior surface of the sacrum. Target registration error over the sacrum was assessed with the fluoroscopic imaging centre on the second sacral vertebral body. Six patients with pelvic ring fractures underwent percutaneous iliosacral screw placement using the CT-3D-fluoroscopy matching navigation. Three pelvic ring fractures were classified as type B2 and three were classified as type C1 according to the AO-OTA classification. Iliosacral screws for the S1 and S2 vertebra were inserted. Results: The mean target registration error over the sacrum was 1.2 mm (0.5-1.9 mm) in the experimental study. Fracture type and amount of vertical displacement did not affect the target registration error. All 12 screws were positioned correctly in the clinical series. There were no postoperative complications including nerve palsy. The mean deviation between the planned and the inserted screw position was 2.5 mm at the screw entry point, 1.8 mm at the area around the nerve root tunnels and 2.2 mm at the tip of the screw. Conclusion: The CT-3D-fluoroscopy matching navigation system was accurate and robust regardless of pelvic ring fracture type and fragment displacement. Percutaneous iliosacral screw insertion with the navigation system is clinically feasible. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:988 / 994
页数:7
相关论文
共 31 条
[1]   Evaluation of 2D and 3D navigation for iliosacral screw fixation [J].
Behrendt, Daniel ;
Muetze, Maria ;
Steinke, Hanno ;
Koestler, Martin ;
Josten, Christoph ;
Boehme, Jorg .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2012, 7 (02) :249-255
[2]   Computer-assisted screw insertion into the first sacral vertebra using a three-dimensional image intensifier: results of a controlled experimental investigation [J].
Briem, D. ;
Linhart, W. ;
Lehmann, W. ;
Begemann, P. G. ;
Adam, G. ;
Schumacher, U. ;
Cullinane, D. M. ;
Rueger, J. M. ;
Windolf, J. .
EUROPEAN SPINE JOURNAL, 2006, 15 (06) :757-763
[3]   Standard multiplanar fluoroscopy versus a fluoroscopically based navigation system for the percutaneous insertion of iliosacral screws - A cadaver model [J].
Collinge, C ;
Coons, D ;
Tornetta, P ;
Aschenbrenner, J .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2005, 19 (04) :254-258
[4]   Quantification of the Upper and Second Sacral Segment Safe Zones in Normal and Dysmorphic Sacra [J].
Gardner, Michael J. ;
Morshed, Saam ;
Nork, Sean E. ;
Ricci, William M. ;
Routt, Milton L. Chip, Jr. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2010, 24 (10) :622-629
[5]  
Gautier E, 2001, CLIN ORTHOP RELAT R, P310
[6]   Screw Placement for Acetabular Fractures: Which Navigation Modality (2-Dimensional vs. 3-Dimensional) Should Be Used? An Experimental Study [J].
Gras, Florian ;
Marintschev, Ivan ;
Klos, Kajetan ;
Mueckley, Thomas ;
Hofmann, Gunther O. ;
Kahler, David M. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2012, 26 (08) :466-473
[7]   2D-fluoroscopic navigated percutaneous screw fixation of pelvic ring injuries - a case series [J].
Gras, Florian ;
Marintschev, Ivan ;
Wilharm, Arne ;
Klos, Kajetan ;
Mueckley, Thomas ;
Hofmann, Gunther O. .
BMC MUSCULOSKELETAL DISORDERS, 2010, 11
[8]   Computer-Assisted Surgery and Intraoperative Three-Dimensional Imaging for Screw Placement in Different Pelvic Regions [J].
Grossterlinden, Lars ;
Nuechtern, Jakob ;
Begemann, Philipp G. C. ;
Fuhrhop, Ina ;
Petersen, Jan P. ;
Ruecker, Andreas ;
Rupprecht, Martin ;
Lehmann, Wolfgang ;
Schumacher, Udo ;
Rueger, Johannes M. ;
Briem, Daniel .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2011, 71 (04) :926-932
[9]   Factors influencing the accuracy of iliosacral screw placement in trauma patients [J].
Grossterlinden, Lars ;
Rueger, Johannes ;
Catala-Lehnen, Philip ;
Rupprecht, Martin ;
Lehmann, Wolfgang ;
Ruecker, Andreas ;
Briem, Daniel .
INTERNATIONAL ORTHOPAEDICS, 2011, 35 (09) :1391-1396
[10]  
Hinsche AF, 2002, CLIN ORTHOP RELAT R, P135