COMBINED VECTOR VELOCITYAND SPECTRAL DOPPLER IMAGING FOR IMPROVED IMAGING OF COMPLEX BLOOD FLOWIN THE CAROTIDARTERIES

被引:38
作者
Ekroll, Ingvild Kinn [1 ,2 ]
Dahl, Torbjorn [3 ]
Torp, Hans [1 ,2 ]
Lovstakken, Lasse [1 ,2 ]
机构
[1] Norwegian Univ Sci & Technol NTNU, Med Imaging Lab, Trondheim, Norway
[2] Norwegian Univ Sci & Technol NTNU, Dept Circulat & Med Imaging, Trondheim, Norway
[3] St Olavs Univ Hosp, Trondheim, Norway
关键词
Blood flow imaging; Carotid artery stenosis; Vector Doppler imaging; Pulsed wave Doppler calibration; Plane wave imaging; MAXIMUM VELOCITY; CAROTID-ARTERY; ULTRASOUND; FREQUENCY;
D O I
10.1016/j.ultrasmedbio.2014.01.021
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Color flow imaging and pulsed wave (PW) Doppler are important diagnostic tools in the examination of patients with carotid artery disease. However, measurement of the true peak systolic velocity is dependent on sample volume placement and the operator's ability to provide an educated guess of the flow direction. Using plane wave transmissions and a duplex imaging scheme, we present an all-in-one modality that provides both vector velocity and spectral Doppler imaging from one acquisition, in addition to separate B-mode images of sufficient quality. The vector Doppler information was used to provide automatically calibrated (angle-corrected) PW Doppler spectra at every image point. It was demonstrated that the combined information can be used to generate spatial maps of the peak systolic velocity, highlighting regions of high velocity and the extent of the stenotic region, which could be used to automate work flow as well as improve the accuracy of measurement of true peak systolic velocity. The modality was tested in a small group (N = 12) of patients with carotid artery disease. PW Doppler, vector velocity and B-mode images could successfully be obtained from a single recording for all patients with a body mass index ranging from 21 to 31 and a carotid depth ranging from 16 to 28 mm. (E-mail: ingvild. k. ekroll@ntnu. no or ingvild.k.ekroll@ntnu. no) (C) 2014 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:1629 / 1640
页数:12
相关论文
共 33 条
[1]   CLASSIFICATION AND NATURAL-HISTORY OF CLINICALLY IDENTIFIABLE SUBTYPES OF CEREBRAL INFARCTION [J].
BAMFORD, J ;
SANDERCOCK, P ;
DENNIS, M ;
BURN, J ;
WARLOW, C .
LANCET, 1991, 337 (8756) :1521-1526
[2]   Ultrafast Compound Doppler Imaging: Providing Full Blood Flow Characterization [J].
Bercoff, Jeremy ;
Montaldo, Gabriel ;
Loupas, Thanasis ;
Savery, David ;
Meziere, Fabien ;
Fink, Mathias ;
Tanter, Mickael .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2011, 58 (01) :134-147
[3]   Speckle tracking for multi-dimensional flow estimation [J].
Bohs, LN ;
Geiman, BJ ;
Anderson, ME ;
Gebhart, SC ;
Trahey, GE .
ULTRASONICS, 2000, 38 (1-8) :369-375
[4]   HIGH-RESOLUTION FREQUENCY-WAVENUMBER SPECTRUM ANALYSIS [J].
CAPON, J .
PROCEEDINGS OF THE IEEE, 1969, 57 (08) :1408-&
[5]   Cross-beam vector Doppler ultrasound for angle-independent velocity measurements [J].
Dunmire, B ;
Beach, KW ;
Labs, KH ;
Plett, M ;
Strandness, DE .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 (08) :1213-1235
[6]  
Ekroll IK, 2012, IEEE T ULTRASON FERR, V59, P1182, DOI [10.1109/TUFFC.2012.2308, http://dx.doi.org/10.1109/TUFFC.2012.2308]
[7]   Simultaneous Quantification of Flow and Tissue Velocities Based on Multi-Angle Plane Wave Imaging [J].
Ekroll, Ingvild Kinn ;
Swillens, Abigail ;
Segers, Patrick ;
Dahl, Torbjorn ;
Torp, Hans ;
Lovstakken, Lasse .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2013, 60 (04) :727-738
[8]   Guidelines for the Primary Prevention of Stroke A Guideline for Healthcare Professionals From the American Heart Association/American Stroke Association [J].
Goldstein, Larry B. ;
Bushnell, Chair Cheryl D. ;
Adams, Robert J. ;
Appel, Lawrence J. ;
Braun, Lynne T. ;
Chaturvedi, Seemant ;
Creager, Mark A. ;
Culebras, Antonio ;
Eckel, Robert H. ;
Hart, Robert G. ;
Hinchey, Judith A. ;
Howard, Virginia J. ;
Jauch, Edward C. ;
Levine, Steven R. ;
Meschia, James F. ;
Moore, Wesley S. ;
Nixon, J. V. ;
Pearson, Thomas A. .
STROKE, 2011, 42 (02) :517-584
[9]   Adaptive Spectral Doppler Estimation [J].
Gran, Fredrik ;
Jakobsson, Andreas ;
Jensen, Jorgen Arendt .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2009, 56 (04) :700-714
[10]   In-vivo Examples of Flow Patterns With The Fast Vector Velocity Ultrasound Method [J].
Hansen, K. L. ;
Udesen, J. ;
Gran, F. ;
Jensen, J. A. ;
Nielsen, M. Bachmann .
ULTRASCHALL IN DER MEDIZIN, 2009, 30 (05) :471-477