Fixation of a Trabecular Metal Knee Arthroplasty Component A Prospective Randomized Study

被引:83
|
作者
Dunbar, M. J. [1 ]
Wilson, D. A. J. [1 ]
Hennigar, A. W. [1 ]
Amirault, J. D. [1 ]
Gross, M. [1 ]
Reardon, G. P. [1 ]
机构
[1] QEII Hlth Sci Ctr, Halifax, NS B3H 3A7, Canada
来源
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME | 2009年 / 91A卷 / 07期
关键词
CEMENTED TIBIAL COMPONENTS; ROENTGEN STEREOPHOTOGRAMMETRIC ANALYSIS; COMPARING EARLY FIXATION; POROUS TANTALUM; FOLLOW-UP; MIGRATION; RADIOSTEREOMETRY; PROSTHESES; JOINT; RSA;
D O I
10.2106/JBJS.H.00282
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Uncemented tibial components of total knee replacements have lower survival rates than cemented components. Radiostereometric analysis is a highly accurate, effective tool for investigating new implant designs. The purpose of this study was to compare an uncemented Trabecular Metal tibial component with a conventional cemented stemmed tibial component of the same design. Methods: Seventy subjects undergoing total knee replacement were randomized to receive either the Trabecular Metal or the cemented tibial component. Radiostereometric analysis of micromotion of the tibial components was performed postoperatively at six, twelve, and twenty-four months, and the maximum total point motion of the implant and three-dimensional translations and rotations were recorded. Results: Follow-up was complete for twenty-eight subjects in the Trabecular Metal group and twenty-one subjects in the cemented group. A subset of the Trabecular Metal components migrated extensively in the postoperative period, but all stabilized by one year and the proportion considered to be at risk for early aseptic loosening was 0.0 (95% confidence interval, 0.0 to 0.12) in the group as a whole. Four cemented components were considered to be at risk for early aseptic loosening (proportion at risk, 0.19; 95% confidence interval, 0.08 to 0.4). Conclusions: This study suggests that the Trabecular Metal component may be an effective alternative to the standard cemented tibial component.
引用
收藏
页码:1578 / 1586
页数:9
相关论文
共 50 条
  • [1] Preliminary Results of an Uncemented Trabecular Metal Tibial Component in Total Knee Arthroplasty
    Helm, Anthony T.
    Kerin, Cronan
    Ghalayini, Simon R. A.
    McLauchlan, George J.
    JOURNAL OF ARTHROPLASTY, 2009, 24 (06): : 941 - 944
  • [2] Does metal backing improve fixation of tibial component in unicondylar knee arthroplasty?: A randomized radiostereometric analysis
    Hyldahl, HC
    Regnér, L
    Carlsson, L
    Kärrholm, J
    Weidenhielm, L
    JOURNAL OF ARTHROPLASTY, 2001, 16 (02): : 174 - 179
  • [3] TIBIAL COMPONENT FIXATION IN KNEE ARTHROPLASTY
    RYD, L
    LINDSTRAND, A
    ROSENQUIST, R
    SELVIK, G
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1986, (213) : 141 - 149
  • [4] Trabecular Metal in Total Knee Arthroplasty Associated with Higher Knee Scores: A Randomized Controlled Trial
    Fernandez-Fairen, Mariano
    Hernandez-Vaquero, Daniel
    Murcia, Antonio
    Torres, Ana
    Llopis, Rafael
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (11) : 3543 - 3553
  • [5] Epiphyseal fixation in revision total knee arthroplasty: a comparison between trabecular metal and titanium augments
    Franceschini, Vincenzo
    Cavallo, Giuseppe
    Lamberti, Alfredo
    Pastore, Francesco
    Montenegro, Luca
    Baldini, Andrea
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2024, 144 (11) : 4667 - 4675
  • [6] Total Knee Arthroplasty with an Uncemented Trabecular Metal Tibial Component A Registry-Based Analysis
    Niemelainen, Mika
    Skytta, Eerik T.
    Remes, Ville
    Makela, Keijo
    Eskelinen, Antti
    JOURNAL OF ARTHROPLASTY, 2014, 29 (01): : 57 - 60
  • [7] Minimum 6 year results of an uncemented trabecular metal tibial component in total knee arthroplasty
    Ghalayini, S. R. A.
    Helm, A. T.
    McLauchlan, G. J.
    KNEE, 2012, 19 (06): : 872 - 874
  • [8] Deep-dish congruent tibial component use in total knee arthroplasty - A randomized prospective study
    Laskin, RS
    Maruyama, Y
    Villaneuva, M
    Bourne, R
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2000, (380) : 36 - 44
  • [9] Trabecular metal augments in revision total knee arthroplasty
    Berli, M. C.
    Sporer, S. M.
    Paprosky, W. G.
    SWISS MEDICAL WEEKLY, 2009, 139 (23-24) : 7S - 7S
  • [10] USE OF TANTALUM TRABECULAR METAL IN TOTAL KNEE ARTHROPLASTY
    Camera, A.
    Cattaneo, G.
    Tedino, R.
    MINERVA ORTOPEDICA E TRAUMATOLOGICA, 2010, 61 (03) : 159 - 164