Wrist-worn optical and chest strap heart rate comparison in a heterogeneous sample of healthy individuals and in coronary artery disease patients

被引:38
作者
Sartor, Francesco [1 ]
Gelissen, Jos [1 ]
Van Dinther, Ralph [2 ]
Roovers, David [3 ]
Papini, Gabriele B. [1 ,4 ]
Coppola, Giuseppe [3 ]
机构
[1] Philips Res, Dept Personal Hlth, High Tech Campus 34,P005,POB WB61, NL-5656 AE Eindhoven, Netherlands
[2] Philips Res, Standardizat Res & Robust Sensing, Eindhoven, Netherlands
[3] Philips Patient Care & Monitoring Solut, Connected Sensing, Eindhoven, Netherlands
[4] Eindhoven Univ Technol, Dept Elect Engn, Eindhoven, Netherlands
关键词
Heart rate; PPG; Evaluation; Accuracy; RATE MONITORS; ACCURACY; PHOTOPLETHYSMOGRAPHY;
D O I
10.1186/s13102-018-0098-0
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
Background: The need for unobtrusive HR (heart rate) monitoring has led to the development of a new generation of strapless HR monitors. The aim of this study was to determine whether such an unobtrusive, wrist-worn optical HR monitor (OHRM) could be equivalent and therefore a valid alternative to a traditional chest strap during a broad range of activities in a heterogeneous healthy population and coronary artery disease (CAD) patients. Methods: One hundred ninety-nine healthy volunteers, 84 males and 115 females, including 35 overweight-obese subjects, 53 pregnant women, and 20 CAD patients were tested in the present study. Second-by-second HR measured by the OHRM was concurrently evaluated against an ECG-based chest strap monitor during a broad range of activities (i.e., walking, running, cycling, gym, household, and sedentary activities). Results: Data coverage, percentage of time the OHRM provides a HR not larger than 10 bpm from the reference, went from a minimum of 92% of the time in the least periodic activity (i.e., gym), to 95% during the most intense activity (i.e., running), and to a maximum of 98% for sedentary activities. The limits of agreement of the difference between the OHRM and the chest strap HR were within the range of +/- 15 bpm. The OHRM showed a concordance correlation coefficient of 0.98. Overall, the mean absolute error was not larger than 3 bpm, which can be considered clinically acceptable for a number of applications. A similar performance was found for CAD (94.2% coverage, 2.4 bpm error), but the small sample size does not allow any quantitative comparison. Conclusion: Heart rate measured by OHRM at the wrist and ECG-based HR measured via a traditional chest strap are acceptably close in a broad range of activities in a heterogeneous, healthy population, and showed initial promising results also in CAD patients.
引用
收藏
页数:10
相关论文
共 33 条
[1]  
Abboud F.M., 1984, Handbook of Physiology: Peripheral Circulation and Organ Blood Flow, V3, P675
[2]   Heart rate monitoring - Applications and limitations [J].
Achten, J ;
Jeukendrup, AE .
SPORTS MEDICINE, 2003, 33 (07) :517-538
[3]   Photoplethysmography and its application in clinical physiological measurement [J].
Allen, John .
PHYSIOLOGICAL MEASUREMENT, 2007, 28 (03) :R1-R39
[4]  
Andre David, 2007, J Diabetes Sci Technol, V1, P760
[5]  
ANSI/AAMI, 2002, CARD MON HEART RAT M, VEC13
[6]   The Accuracy of Heart Rate Monitoring by Some Wrist-Worn Activity Trackers [J].
Cadmus-Bertram, Lisa ;
Gangnon, Ronald ;
Wirkus, Emily J. ;
Thraen-Borowski, Keith M. ;
Gorzelitz-Liebhauser, Jessica .
ANNALS OF INTERNAL MEDICINE, 2017, 166 (08) :610-+
[7]   Accuracy of smartphone apps for heart rate measurement [J].
Coppetti, Thomas ;
Brauchlin, Andreas ;
Mueggler, Simon ;
Attinger-Toller, Adrian ;
Templin, Christian ;
Schoenrath, Felix ;
Hellermann, Jens ;
Luescher, Thomas F. ;
Biaggi, Patric ;
Wyss, Christophe A. .
EUROPEAN JOURNAL OF PREVENTIVE CARDIOLOGY, 2017, 24 (12) :1287-1293
[8]  
CUMMING GR, 1977, CAN MED ASSOC J, V117, P1138
[9]  
Delgado-Gonzalo R, 2015, IEEE ENG MED BIO, P430, DOI 10.1109/EMBC.2015.7318391
[10]  
Fick A, 1870, PHYSIOMED SOC WURZUN, V2, P16