Comparison of different contactless conductivity detectors for the determination of small inorganic ions by capillary electrophoresis

被引:31
作者
Kuban, Pavel
Evenhuis, Christopher J.
Macka, Mirek
Haddad, Paul R.
Hauser, Peter C.
机构
[1] Univ Basel, Dept Chem, CH-4004 Basel, Switzerland
[2] Univ Tasmania, Fac Sci Engn & Technol, Sch Chem, Australian Ctr Res Separat Sci, Hobart, Tas 7001, Australia
关键词
capillary electrophoresis; contactless conductivity detection; detectors; inorganic ions;
D O I
10.1002/elan.200603528
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The analytical performance parameters of four different contactless conductivity detectors for capillary electrophoresis were determined. These detectors were designed either with a miniature cell for versatility, with high voltage excitation for high sensitivity or with a battery power supply for field application. One of the units is commercially available. The plate numbers and reproducibility of peak areas (typically about 2%) were very similar, indicating that these parameters are generally limited by the separation and injection procedures rather than the detectors. The linear dynamic ranges are better than 2 orders of magnitude for all detectors and the correlation coefficients were also almost identical. Significant differences were found with regard to the detection limits. For the detector with a miniature cell, which lacks an in-cell amplifier, the detection limit was typically 1.5 mu M for the inorganic ions tested (K+, Ca2+, Na+, Mg2+, Li+, Cl-, NO3 and SO42-), while with the other 3 detectors this parameter was as low as about 0.1 mu M. The use of buffer solutions with relatively high background conductivity was found to lead to detection limits which were up to one order of magnitude higher. The extent of deterioration of this parameter with buffer conductivity was found to be related to the excitation voltage and was most pronounced for the high voltage detector.
引用
收藏
页码:1289 / 1296
页数:8
相关论文
共 25 条
  • [1] Baltussen E, 2002, ELECTROPHORESIS, V23, P2888, DOI 10.1002/1522-2683(200209)23:17<2888::AID-ELPS2888>3.0.CO
  • [2] 2-4
  • [3] Understanding capacitively coupled contactless conductivity detection in capillary and microchip electrophoresis. Part 1. Fundamentals
    Brito-Neto, JGA
    da Silva, JAF
    Blanes, L
    do Lago, CL
    [J]. ELECTROANALYSIS, 2005, 17 (13) : 1198 - 1206
  • [4] Understanding capacitively coupled contactless conductivity detection in capillary and microchip electrophoresis. Part 2. Peak shape, stray capacitance, noise, and actual electronics
    Brito-Neto, JGA
    da Silva, JAF
    Blanes, L
    do Lago, CL
    [J]. ELECTROANALYSIS, 2005, 17 (13) : 1207 - 1214
  • [5] da Silva JAF, 1998, ANAL CHEM, V70, P4339
  • [6] HIGH-FREQUENCY CONTACTLESS CONDUCTIVITY DETECTION IN ISOTACHOPHORESIS
    GAS, B
    DEMJANENKO, M
    VACIK, J
    [J]. JOURNAL OF CHROMATOGRAPHY, 1980, 192 (02): : 253 - 257
  • [7] GAS B, 1980, CHEM LISTY, V74, P652
  • [8] Conductivity detection for conventional and miniaturised capillary electrophoresis systems
    Guijt, RM
    Evenhuis, CJ
    Macka, M
    Haddad, PR
    [J]. ELECTROPHORESIS, 2004, 25 (23-24) : 4032 - 4057
  • [9] Thinly insulated wire cells -: A new device for sensitive contactless conductivity detection in flow analyses
    Hohercáková, Z
    Opekar, FE
    Stulík, K
    [J]. ELECTROANALYSIS, 2005, 17 (21) : 1924 - 1930
  • [10] Expanded electrical model of a contactless conductivity detector: Development and verification
    Johnston, SE
    Fadgen, KE
    Tolley, LT
    Jorgenson, JW
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2005, 1094 (1-2) : 148 - 157