Use of a parallel path nebulizer for capillary-based microseparation techniques coupled with an inductively coupled plasma mass spectrometer for speciation measurements
被引:37
作者:
Yanes, EG
[1
]
Miller-Ihli, NJ
[1
]
机构:
[1] USDA ARS, Beltsville Human Nutr Res Ctr, Food Composit Lab, Beltsville, MD 20705 USA
关键词:
cobalamins;
vitamin B12;
microseparation;
ICP-MS;
Mira Mist CE nebulizer;
parallel path nebulizer;
CE;
micro-HPLC;
speciation;
D O I:
10.1016/j.sab.2004.03.005
中图分类号:
O433 [光谱学];
学科分类号:
0703 ;
070302 ;
摘要:
A low flow, parallel path Mira Mist CE nebulizer designed for capillary electrophoresis (CE) was evaluated as a function of make-up solution flow rate, composition, and concentration, as well as the nebulizer gas flow rate. This research was conducted in support of a project related to the separation and quantification of cobalamin (vitamin B-12) species using microseparation techniques combined with inductively coupled plasma mass spectrometry (ICP-MS) detection. As such, Co signals were monitored during the nebulizer characterization process. Transient effects in the ICP were studied to evaluate the suitability of using gradients for microseparations and the benefit of using methanol for the make-up solution was demonstrated. Co signal response changed significantly as a function of changing methanol concentrations of the make-up solution and maximum signal enhancement was seen at 20% methanol with a 15 mul/min flow rate. Evaluation of the effect of changing the nebulizer gas flow rates showed that argon flows from 0.8 to 1.2 l/min were equally effective. The Mira Mist CE parallel path nebulizer was then evaluated for interfacing capillary microseparation techniques including capillary electrophoresis (CE) and micro high performance liquid chromatography (muHPLC) to inductively coupled plasma mass spectrometry (ICP-MS). A mixture of four cobalamin species standards (cyanocobalamin, hydroxocobalamin, methylcobalamin, and 5' deoxyadenosylcobalamin) and the. corrinoid analogue cobinamide dicyanide were successfully separated using both CE-ICP-MS and muHPLC-ICP-MS using the parallel path nebulizer with a make-up solution containing 20% methanol with a flow rate of 15 mul/min. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:883 / 890
页数:8
相关论文