Polyurethane versus silicone port a cath: What's going on at removal?

被引:12
作者
Alzahrani, Khalid [1 ]
Lejeune, Julien [2 ]
Lakhal, Walid [1 ]
Morel, Baptiste [3 ]
Cook, Ann-Rose [1 ]
Braik, Karim [1 ]
Lardy, Hubert [1 ]
Binet, Aurelien [1 ]
机构
[1] Hosp Univ Ctr Tours, Gatien Clocheville Hosp, Pediat Surg Unit, F-37000 Tours, France
[2] Hosp Univ Ctr Tours, Gatien Clocheville Hosp, Pediat Oncohematol Unit, F-37000 Tours, France
[3] Hosp Univ Ctr Tours, Gatien Clocheville Hosp, Pediat Radiol Unit, F-37000 Tours, France
关键词
Vascular access devices; Oncology; Pediatric surgery; VENOUS ACCESS; COMPLICATIONS; CHILDREN; DEVICES; STUCK; RARE;
D O I
10.1016/j.jpedsurg.2017.06.025
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Background: Port-a-cath (PAC) is an essential device in the management of the patients of chronic illness, but despite theirs benefits there arc many complications either at the time of insertion or at time of removal. Our aim of this study is to evaluate the fracture rate of the catheter at removal time in comparison with catheter type either polyurethane or silicone. Methods: A retrospective monocentric study of all PACs which were removed at our university pediatric hospital between 1 January 2006 and 31 December 2016. Two groups were compared: polyurethane group and silicone group. Results: Total of 216 central lines were removed, the mean age at the time of extraction was 9.7 +/- 4.9 years and the mean time for both catheter was 2.7 +/- 1.6 years, fracture occurred in 11 catheter of the polyurethane group (n = 119), with no fracture of silicone group (n = 86), in the polyurethane group, the risk of catheter fracture is significantly related to the duration of the PAC in place. Conclusion: We found that the polyurethane-based catheters are more vulnerable for rupture and retained fragment in the blood vessels, especially if left in place for long time, for this reason we have switched to silicone-based catheter and all catheters should be remove within duration maximal of 2 years. Type of study: Prognosis study. (C) 2017 Elsevier Inc. All rights reserved.
引用
收藏
页码:1417 / 1419
页数:3
相关论文
共 13 条
[1]   Calcified catheter "cast": a rare complication of indwelling central venous catheters in infants [J].
Anderson, MAB ;
Poenaru, D ;
Kamal, I .
PEDIATRIC SURGERY INTERNATIONAL, 1998, 13 (08) :610-612
[2]   Rare but serious complications of central line insertion [J].
Askegard-Giesmann, Johanna R. ;
Caniano, Donna A. ;
Kenney, Brian D. .
SEMINARS IN PEDIATRIC SURGERY, 2009, 18 (02) :73-83
[3]   Risk of thrombosis and infections of central venous catheters and totally implanted access ports in patients treated for cancer [J].
Beckers, M. M. J. ;
Ruven, H. J. T. ;
Seldenrijk, C. A. ;
Prins, M. H. ;
Biesma, D. H. .
THROMBOSIS RESEARCH, 2010, 125 (04) :318-321
[4]   Removal of Retained Port-A Catheter in Central Lines in a Pediatric Population [J].
Chen, Po-Chuan ;
Chen, Chau-Jing .
TZU CHI MEDICAL JOURNAL, 2007, 19 (04) :245-248
[5]   Complications associated with an implantable vascular access device [J].
Dillon, Patrick A. ;
Foglia, Robert P. .
JOURNAL OF PEDIATRIC SURGERY, 2006, 41 (09) :1582-1587
[6]   Jugular vein catheter placement: Histologic features and development of catheter-related (fibrin) sheaths in a swine model [J].
Forauer, Andrew R. ;
Theoharis, Constantine G. A. ;
Dasika, Narasimham L. .
RADIOLOGY, 2006, 240 (02) :427-434
[7]   A new technique to remove a "stuck" totally implantable venous access catheter [J].
Huang, Shu-Chien ;
Tsai, Ming-Shian ;
Lai, Hong-Shiee .
JOURNAL OF PEDIATRIC SURGERY, 2009, 44 (07) :1465-1467
[8]   CONTINUOUS VENOUS ACCESS IN CHILDREN FOR LONG-TERM CHEMOTHERAPY BY MEANS OF AN IMPLANTABLE SYSTEM [J].
KRUL, EJ ;
VANLEEUWEN, EF ;
VOS, A ;
VOUTE, PA .
JOURNAL OF PEDIATRIC SURGERY, 1986, 21 (08) :689-690
[9]   Follow-up of 452 totally implantable vascular devices in cystic fibrosis patients [J].
Munck, A ;
Malbezin, S ;
Bloch, J ;
Gerardin, M ;
Lebourgeois, M ;
Derelle, J ;
Bremont, F ;
Sermet, I ;
Munck, MR ;
Navarro, J .
EUROPEAN RESPIRATORY JOURNAL, 2004, 23 (03) :430-434
[10]   Totally implantable venous access port systems and risk factors for complications: A one-year prospective study in a cancer centre [J].
Narducci, F. ;
Jean-Laurent, M. ;
Boulanger, L. ;
El Bedoui, S. ;
Mallet, Y. ;
Houpeau, J. L. ;
Hamdani, A. ;
Penel, N. ;
Fournier, C. .
EJSO, 2011, 37 (10) :913-918