Technical Mistakes during the Acquisition of the Electrocardiogram

被引:58
作者
Garcia-Niebla, Javier [1 ]
Llontop-Garcia, Pablo [1 ]
Ignacio Valle-Racero, Juan [2 ]
Serra-Autonell, Guillem
Batchvarov, Velislav N. [3 ]
Bayes de Luna, Antonio [4 ]
机构
[1] Valle Golfo Hlth Ctr, Area Salud Hierro, Serv Sanitarios, Islas Canarias, Spain
[2] Hosp Univ Virgen Rocio, Serv Cardiol, Seville, Spain
[3] St Georges Univ London, Div Cardiac & Vasc Sci, London, England
[4] Hosp Santa Creu & Sant Pau, Inst Catala Ciencies Cardiovasc, Barcelona, Spain
关键词
electrocardiogram; artifacts; electrode misplacement; filters; leads; LIMB LEAD; 12-LEAD ELECTROCARDIOGRAM; MYOCARDIAL-INFARCTION; CLINICAL-CARDIOLOGY; SCIENTIFIC STATEMENT; CHEST ELECTRODE; LEFT ARM; LEFT LEG; P-WAVE; EASI;
D O I
10.1111/j.1542-474X.2009.00328.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In addition to knowledge of normal and pathological patterns, the correct interpretation of electrocardiographic (ECG) recordings requires the use of acquisition procedures according to approved standards. Most manuals on standard electrocardiography devote little attention to inadequate ECG recordings. In this article, we present the most frequent ECG patterns resulting from errors in limb and precordial lead placement, artifacts in 12-lead ECG as well as inadequate filter application; we also review alternative systems to the standard ECG, which may help minimize errors.
引用
收藏
页码:389 / 403
页数:15
相关论文
共 68 条