An algorithm to improve the estimation accuracy of a non-invasive method for cardiac output measurement based on prolonged expiration

被引:0
作者
Schena, E. [1 ]
Cecchini, S. [1 ]
Saccomandi, P. [1 ]
Leuzzi, M. [2 ]
Silvestri, S. [1 ]
机构
[1] Univ Campus Biomed Roma, Unit Measurements & Biomed Instrumentat, Ctr Integrated Res, Via Alvaro Portillo 21, I-00128 Rome, Italy
[2] Univ Campus Biomed Roma, \ Rome, Italy
来源
2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC) | 2013年
关键词
THERMODILUTION;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Cardiac output (CO) monitoring is important in the hemodynamic management of critically ill patients. In a previous study, a novel non-invasive technique for CO monitoring based on prolonged expiration was proposed. The novel method showed good agreement with thermodilution on stable mechanically ventilated patients; unstable patients were excluded. The aim of this study is to improve the outcome of the above mentioned method on hemodynamic unstable patients, requiring vasoactive medications, and showing marked cardiogenic oscillations on waveforms related to expired gases. This prospective study has been carried out on three cardiac surgery patients; eighteen CO measurements were performed on each patient, and these values were compared with data obtained by thermodilution. The designed and tested algorithm allowed to reach a good agreement between CO measured by our method and by thermodilution (e.g., the mean percentage differences were 4%, 11% and 3%). Even though further validation is necessary, the results are quite promising and the adopted solution appears to allow the suitability of the prolonged expiration method also on unstable patients.
引用
收藏
页码:1823 / 1826
页数:4
相关论文
共 15 条
[1]  
Cecchini S, 2011, IEEE ENG MED BIO, P1245, DOI 10.1109/IEMBS.2011.6090293
[2]  
CECCHINI S, 2012, P 34 ANN INT C IEEE, P2708
[3]  
Cecchini S., 2011, MEMEA 2011 IEEE INT, P143
[4]   Non-invasive Estimation of Cardiac Output in Mechanically Ventilated Patients: A Prolonged Expiration Method [J].
Cecchini, Stefano ;
Schena, Emiliano ;
Notaro, Maria ;
Carassiti, Massimiliano ;
Silvestri, Sergio .
ANNALS OF BIOMEDICAL ENGINEERING, 2012, 40 (08) :1777-1789
[5]   CARDIAC OSCILLATIONS IN EXPIRED GAS TENSIONS, AND REGIONAL PULMONARY BLOOD FLOW [J].
FOWLER, KT ;
READ, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1961, 16 (05) :863-&
[6]   A NEW METHOD FOR NON-INVASIVE BEDSIDE DETERMINATION OF PULMONARY BLOOD-FLOW [J].
GEDEON, A ;
FORSLUND, L ;
HEDENSTIERNA, G ;
ROMANO, E .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1980, 18 (04) :411-418
[7]   MODEL OF FLUCTUATING ALVEOLAR GAS-EXCHANGE DURING RESPIRATORY CYCLE [J].
HLASTALA, MP .
RESPIRATION PHYSIOLOGY, 1972, 15 (02) :214-&
[8]   Partial CO2 rebreathing cardiac output -: Operating principles of the NICO™ system [J].
Jaffe, MB .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (06) :387-401
[9]  
JCGM, 2008, GUID EXPR UNC MEAS, P100
[10]   Non-invasive cardiac output measurement using a fast mixing box to measure carbon dioxide elimination [J].
Killick, C. J. ;
Parkin, W. G. .