Fluid Embeddable Coupled Coil Sensor for Wireless pH Monitoring in a Bioreactor

被引:26
作者
Bhadra, Sharmistha [1 ]
Blunt, Warren [2 ]
Dynowski, Chris [1 ]
McDonald, Michael [3 ]
Thomson, Douglas J. [1 ]
Freund, Michael S. [3 ]
Cicek, Nazim [2 ]
Bridges, Greg E. [1 ]
机构
[1] Univ Manitoba, Dept Elect & Comp Engn, Winnipeg, MB R3T 5V6, Canada
[2] Univ Manitoba, Dept Biosyst Engn, Winnipeg, MB R3T 2N2, Canada
[3] Univ Manitoba, Dept Chem, Winnipeg, MB R3T 2N2, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Bioreactor; coupled coil; in-fluid; passive; pH sensor; Saccharomyces cerevisiae; wireless; Yarrowia lipolytica; PASSIVE SENSOR; ELECTRODE; SYSTEM;
D O I
10.1109/TIM.2013.2292279
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A passive embeddable coupled coil sensor for remote bioprocess pH monitoring is described. The sensor is sterilizable, able to operate in a fluid medium, and small enough to fit inside a small bioreactor or test tube. It consists of a planar spiral inductor connected parallel to a varactor forming an LC resonant circuit. A pH combination electrode made of an iridium/iridium oxide sensing electrode and a silver/silver chloride reference electrode is connected parallel to the varactor. A potential difference change across the electrodes due to pH variation of the medium changes the voltage-dependent capacitance and shifts the resonant frequency of the sensor. An interrogator coil is inductively coupled to the sensor coil and remotely tracks the resonant frequency of the sensor. For in-fluid monitoring, the sensor is encapsulated in a manner that reduces the influence of the permittivity and conductivity of the medium. The sensor, calibrated over 2-12 pH range, exhibits a rapid response with 2.477-MHz/pH sensitivity. The sensor was used for remote pH monitoring of Yarrowia lipolytica and Saccharomyces cerevisiae fermentation in a shake flask over 63 and 25 h, respectively. The experiments demonstrate that the medium pH can be monitored repeatedly with an accuracy of 0.08 pH.
引用
收藏
页码:1337 / 1346
页数:10
相关论文
共 37 条
  • [1] [Anonymous], 2012, NONINVASIVE PH SENSO
  • [2] [Anonymous], 2012, SENSOLUX
  • [3] [Anonymous], 2012, AMBR SYSTEM OVERVIEW
  • [4] [Anonymous], 2004, ELECTROCHEM SOC INTE
  • [5] A Wireless Passive pH Sensor Based on pH Electrode Potential Measurement
    Bhadra, S.
    Bridges, G. E.
    Thomson, D. J.
    Freund, M. S.
    [J]. 2010 IEEE SENSORS, 2010, : 927 - 930
  • [6] Bhadra S, 2013, IEEE IMTC P, P276
  • [7] A Wireless Passive Sensor for Temperature Compensated Remote pH Monitoring
    Bhadra, Sharmistha
    Tan, Damaris S. Y.
    Thomson, Douglas J.
    Freund, Michael S.
    Bridges, Greg E.
    [J]. IEEE SENSORS JOURNAL, 2013, 13 (06) : 2428 - 2436
  • [8] Electrode Potential-Based Coupled Coil Sensor for Remote pH Monitoring
    Bhadra, Sharmistha
    Bridges, Greg E.
    Thomson, Douglas J.
    Freund, Michael S.
    [J]. IEEE SENSORS JOURNAL, 2011, 11 (11) : 2813 - 2819
  • [9] Introduction to advantages and problems of shaken cultures
    Büchs, J
    [J]. BIOCHEMICAL ENGINEERING JOURNAL, 2001, 7 (02) : 91 - 98
  • [10] In situ measurement of Cl- concentrations and pH at the reinforcing steel/concrete interface by combination sensors
    Du, Rong-Gui
    Hu, Rong-Gang
    Huang, Ruo-Shuang
    Lin, Chang-Jian
    [J]. ANALYTICAL CHEMISTRY, 2006, 78 (09) : 3179 - 3185