PET optimization for improved assessment and accurate quantification of 90Y-microsphere biodistribution after radioembolization

被引:19
作者
Marti-Climent, Josep M. [1 ]
Prieto, Elena [1 ]
Elosua, Cesar [1 ]
Rodriguez-Fraile, Macarena [1 ]
Dominguez-Prado, Ines [1 ]
Vigil, Carmen [1 ]
Garcia-Velloso, Maria J. [1 ]
Arbizu, Javier [1 ]
Penuelas, Ivan [1 ]
Richter, Jose A. [1 ]
机构
[1] Univ Navarra Clin, Dept Nucl Med, Pamplona 31008, Spain
关键词
90-yttrium; radioembolization; SELECTIVE INTERNAL RADIOTHERAPY; TIME-OF-FLIGHT; POINT-SPREAD-FUNCTION; Y-90; MICROSPHERES; RADIATION-THERAPY; PAIR PRODUCTION; TOF PET; RECONSTRUCTION; PERFORMANCE; DOSIMETRY;
D O I
10.1118/1.4892383
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Y-90-microspheres are widely used for the radioembolization of metastatic liver cancer or hepatocellular carcinoma and there is a growing interest for imaging Y-90-microspheres with PET. The aim of this study is to evaluate the performance of a current generation PET/CT scanner for Y-90 imaging and to optimize the PET protocol to improve the assessment and the quantification of Y-90-microsphere biodistribution after radioembolization. Methods: Data were acquired on a Biograph mCT-TrueV scanner with time of flight (TOF) and point spread function (PSF) modeling. Spatial resolution was measured with a Y-90 point source. Sensitivity was evaluated using the NEMA 70 cm line source filled with Y-90. To evaluate the count rate performance, Y-90 vials with activity ranging from 3.64 to 0.035 GBq were measured in the center of the field of view (CFOV). The energy spectrum was evaluated. Image quality with different reconstructions was studied using the Jaszczak phantom containing six hollow spheres (diameters: 31.3, 28.1, 21.8, 16.1, 13.3, and 10.5 mm), filled with a 207 kBq/ml Y-90 concentration and a 5: 1 sphere-to-background ratio. Acquisition time was adjusted to simulate the quality of a realistic clinical PET acquisition of a patient treated with SIR-Spheres (R). The developed methodology was applied to ten patients after SIR-Spheres (R) treatment acquiring a 10 min per bed PET. Results: The energy spectrum showed the Y-90 bremsstrahlung radiation. The Y-90 transverse resolution, with filtered backprojection reconstruction, was 4.5 mm in the CFOV and degraded to 5.0mm at 10 cm off-axis. Y-90 absolute sensitivity was 0.40 kcps/MBq in the center of the field of view. Tendency of true and random rates as a function of the Y-90 activity could be accurately described using linear and quadratic models, respectively. Phantom studies demonstrated that, due to low count statistics in Y-90 PET acquisition, the optimal parameters for the standard OSEM+PSF reconstruction were only one iteration and a postreconstruction filter of 6 mm FWHM, for both TOF and non-TOF reconstructions. Moreover, when TOF is included, the signal to noise ratio increased and visibility achieved 100% by the experienced observers and 93.3% according to the Rose model of statistical detection. In 50% of patients, TOF allowed the visualization of Y-90 radioembolized lesions not seen without TOF, confirming phantom results. Liver activity was accurately quantified, with no significant differences between reconstructed and actual delivered activity to the whole-liver [mean relative difference (10.2 +/- 14.7)%]. Conclusions: Qualitative and quantitative Y-90 PET imaging improved with the introduction of TOF in a PET/CT scanner, thereby allowing the visualization of microsphere deposition in lesions not visible in non-TOF images. This technique accurately quantifies the total activity delivered to the liver during radioembolization with Y-90-microspheres and allows dose estimation. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:10
相关论文
共 34 条
[1]  
Baechler S, 2011, J NUCL MED, V52
[2]   90Y-PET for the assessment of microsphere biodistribution after selective internal radiotherapy [J].
Bagni, Oreste ;
D'Arienzo, Marco ;
Chiaramida, Paola ;
Chiacchiararelli, Laura ;
Cannas, Pina ;
D'Agostini, Antonio ;
Cianni, Roberto ;
Salvatori, Rita ;
Scopinaro, Francesco .
NUCLEAR MEDICINE COMMUNICATIONS, 2012, 33 (02) :198-204
[3]   Investigation of 18F-FDG 3D mode PET image quality versus acquisition time [J].
Brown, Colin ;
Dempsey, Mary-Frances ;
Gillen, Gerry ;
Elliott, Alex T. .
NUCLEAR MEDICINE COMMUNICATIONS, 2010, 31 (03) :254-259
[4]   The Rose model, revisited [J].
Burgess, AE .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1999, 16 (03) :633-646
[5]  
Carlier T., 2013, EJNMMI research, V3, P1
[6]  
Chiesa C, 2011, Q J NUCL MED MOL IM, V55, P168
[7]   Effect of randoms on signal-to-noise ratio in TOF PET [J].
Conti, Maurizio .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (03) :1188-1193
[8]   90Y PET-based dosimetry after selective internal radiotherapy treatments [J].
D'Arienzo, Marco ;
Chiaramida, Paola ;
Chiacchiararelli, Laura ;
Coniglio, Angela ;
Cianni, Roberto ;
Salvatori, Rita ;
Ruzza, Alberto ;
Scopinaro, Francesco ;
Bagni, Oreste .
NUCLEAR MEDICINE COMMUNICATIONS, 2012, 33 (06) :633-640
[9]   Quantitative Comparison of PET and Bremsstrahlung SPECT for Imaging the In Vivo Yttrium-90 Microsphere Distribution after Liver Radioembolization [J].
Elschot, Mattijs ;
Vermolen, Bart J. ;
Lam, Marnix G. E. H. ;
de Keizer, Bart ;
van den Bosch, Maurice A. A. J. ;
de Jong, Hugo W. A. M. .
PLOS ONE, 2013, 8 (02)
[10]   Internal Pair Production of 90Y Permits Hepatic Localization of Microspheres Using Routine PET: Proof of Concept [J].
Gates, Vanessa L. ;
Esmail, Abdulredha A. H. ;
Marshall, Karen ;
Spies, Stewart ;
Salem, Riad .
JOURNAL OF NUCLEAR MEDICINE, 2011, 52 (01) :72-76