Temperature Evaluation During PMMA Screw Augmentation in Osteoporotic Bone-An In Vitro Study About the Risk of Thermal Necrosis in Human Femoral Heads

被引:63
作者
Boner, Vanessa [1 ]
Kuhn, Philipp [1 ]
Mendel, Thomas [1 ,2 ]
Gisep, Armando [1 ]
机构
[1] AO Dev Inst, Davos, Switzerland
[2] BG Kliniken Bergmannstrost, Klin Unfall & Wiederherstellungschirurg, Halle, Saale, Germany
关键词
osteoporosis; PMMA augmentation; exothermic reaction; thermal necrosis; JOINT REPLACEMENT; INTERNAL-FIXATION; CEMENT; POLYMERIZATION; FRACTURES; POLYMETHYLMETHACRYLATE; INTERFACE; NECK;
D O I
10.1002/jbm.b.31353
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The use of polymethylmethacrylate (PMMA) bone cement to augment hip screws reduces cut-out risk but is associated with an exothermic reaction. This in vitro investigation evaluated the risk of thermal necrosis when augmenting the implant purchase with PMMA. A pilot study analyzed the effects of different PMMA layer thicknesses on temperatures around an implant. The main study used either 3.0 or 6.0 cc PMMA for hip screw augmentation in human femoral heads. The risk of thermal necrosis was estimated according to critical values reported in literature. Highest temperatures were measured inside the PMMA with a significant drop of average maximum temperatures from the center of the PMMA to the PMMA/bone interface. Risk of thermal necrosis exists with PMMA layer thicknesses greater than 5.0 mm. In the main study, we found no risk of thermal necrosis at the PMMA/bone interface or in the surrounding bone, neither with 3.0 nor 6.0 cc PMMA. The results of the two studies were consistent regarding average peak temperatures related to associated cement layer thicknesses. The results of this in vitro study reduce objections concerning the risk of thermal necrosis when augmenting cancellous bone around hip screws with up to 6.0 cc PMMA. (C) 2009 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 90B: 842-848, 2009
引用
收藏
页码:842 / 848
页数:7
相关论文
共 27 条
[21]   BONE-CEMENT INTERFACE TEMPERATURE DURING TOTAL JOINT REPLACEMENT [J].
RECKLING, FW ;
DILLON, WL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1977, 59 (01) :80-82
[22]   THE EFFECT OF THICKNESS AND PRESSURE ON THE CURING OF PMMA BONE-CEMENT FOR THE TOTAL HIP-JOINT REPLACEMENT [J].
SIH, GC ;
CONNELLY, GM ;
BERMAN, AT .
JOURNAL OF BIOMECHANICS, 1980, 13 (04) :347-352
[23]  
Stanczyk M, 2004, J BIOMECH, V37, P1803, DOI [10.1016/j.biomech.2004.03.002, 10.1016/j.jbiomech.2004.03.002]
[24]   FINITE-ELEMENT TEMPERATURE ANALYSIS OF A TOTAL HIP-REPLACEMENT AND MEASUREMENT OF PMMA CURING TEMPERATURES [J].
SWENSON, LW ;
SCHURMAN, DJ ;
PIZIALI, RL .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1981, 15 (01) :83-96
[25]  
Toksvig-Larsen S, 1995, Eur Spine J, V4, P15, DOI 10.1007/BF00298412
[26]   Theoretical prediction and experimental determination of the effect of mold characteristics on temperature and monomer conversion fraction profiles during polymerization of a PMMA-based bone cement [J].
Vallo, CI .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2002, 63 (05) :627-642
[27]   Biomechanical evaluation of a new augmentation method for enhanced screw fixation in osteoporotic proximal femoral fractures [J].
von der Linden, P. ;
Gisep, A. ;
Boner, V. ;
Windolf, M. ;
Appelt, A. ;
Suhm, N. .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2006, 24 (12) :2230-2237