Outflow resistance of the Baerveldt glaucoma drainage implant and modifications for early postoperative intraocular pressure control

被引:14
作者
Breckenridge, RR [1 ]
Bartholomew, LR [1 ]
Crosson, CE [1 ]
Kent, AR [1 ]
机构
[1] Med Univ S Carolina, Dept Ophthalmol, Hewitt Lab, Ola B Williams Glaucoma Ctr, Charleston, SC 29425 USA
关键词
D O I
10.1097/01.ijg.0000131759.48295.5c
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To determine outflow resistance of the Baerveldt glaucoma implant using different tube configurations. Methods: Outflow resistance of 6 tube configurations (C1-C6) of Baerveldt implants was measured under conditions of constant pressure perfusion. Pressures ranged from 2 to 55 mm Hg. Venting slits were created using a 7-0 Vicryl, spatulated suture-needle. Seton tubes were occluded by threading a retrograde suture similar to1.5 cm into the lumen. Results: At pressures between 2 and 55 mm Hg, mean outflow resistance of the normally configured scion (ie, open tube; C1) was 0.41 (+/-0.6) mm Hg/muL/min. Resistance was unchanged (mean 0.41 (+/-0.4) mm Hg/muL/min) by the addition of 4 venting slits (C2) to the seton tube. Occlusion of the open scion tube with a 3-0 Supramid(R) suture (0) significantly increased (P < 0.001) mean outflow resistance to 14.99 (+/-0.6) mm Hg/muL/min. Occlusion of the tube with a 4-0 Supramid(R) suture (C4) significantly increased (P < 0.001) mean outflow resistance to 1.09 (+/-0.5) min Hg/muL/min. In implants where tubes were occluded with a 3-0 Supramid(R) suture, the addition of venting slits (C5) significantly decreased (P = 0.038) mean outflow resistance to 8.98 (+/-0.4) min Hg/muL/min. In tubes occluded with a 4-0 Supramid(R) suture, the addition of venting slits (C6) decreased mean outflow resistance to 0.98 (+/-0.6) min Hg/muL/min. Conclusions: Although these results cannot be directly correlated to the clinical setting, they do show that outflow resistance can be modified at the time of surgery by changing tube configuration of the Baerveldt glaucoma implant. Configuration C5 (3-0 Supramid(R) with venting slits) closely approximates the outflow rate in the normal intraocular pressure range.
引用
收藏
页码:396 / 399
页数:4
相关论文
共 10 条
  • [1] EGBERT PR, 1989, OPHTHALMIC SURG LAS, V20, P53
  • [2] Eisenberg DL, 1999, OPHTHALMIC SURG LAS, V30, P662
  • [3] Feldman RM, 1997, J GLAUCOMA, V6, P10
  • [4] Pathophysiology of and prophylaxis against late Ahmed Glaucoma Valve occlusion[J]. Hill, RA;Pirouzian, A;Liaw, LH. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2000(05)
  • [5] Effectiveness of nonocclusive ligature and fenestration used in Baerveldt aqueous shunts for early postoperative intraocular pressure control[J]. Kansal, S;Moster, MR;Kim, D;Schmidt, CM;Wilson, RP;Katz, LJ. JOURNAL OF GLAUCOMA, 2002(01)
  • [6] IN-VITRO AND IN-VIVO FLOW CHARACTERISTICS OF GLAUCOMA DRAINAGE IMPLANTS[J]. PRATA, JA;MERMOUD, A;LABREE, L;MINCKLER, DS. OPHTHALMOLOGY, 1995(06)
  • [7] PREVENTION OF EARLY HYPOTONY ASSOCIATED WITH MOLTENO IMPLANTS BY A NEW OCCLUDING STENT TECHNIQUE[J]. SHERWOOD, MB;SMITH, MF. OPHTHALMOLOGY, 1993(01)
  • [8] COMPARISON OF THE BAERVELDT GLAUCOMA IMPLANT WITH THE DOUBLE-PLATE MOLTENO DRAINAGE IMPLANT[J]. SMITH, MF;DOYLE, JW;SHERWOOD, MB. ARCHIVES OF OPHTHALMOLOGY, 1995(04)
  • [9] Follow-up of the original cohort with the Ahmed glaucoma valve implant[J]. Topouzis, F;Coleman, AL;Choplin, N;Bethlem, MM;Hill, R;Yu, F;Panek, WC;Wilson, MR. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1999(02)
  • [10] THE LONG-TUBE SINGLE PLATE MOLTENO IMPLANT FOR THE TREATMENT OF RECALCITRANT GLAUCOMA[J]. TRAVERSO, CE;TOMEY, KF;ALKAFF, A. INTERNATIONAL OPHTHALMOLOGY, 1989(1-2)