Coupling of capillary electrophoresis with electrospray ionization multiplexing ion mobility spectrometry

被引:4
作者
Guo, Jianheng [1 ,2 ]
Li, Guozhu [3 ]
Zhang, Hanghang [3 ]
Jia, Xu [3 ]
Meng, Qingyan [3 ]
Liu, Wenjie [1 ,3 ]
机构
[1] Xiangtan Univ, Coll Chem Engn, Xiangtan, Hunan, Peoples R China
[2] Southwest MinZu Univ, Coll Pharm, Chengdu, Sichuan, Peoples R China
[3] Xinjiang Prod & Construct Corps Key Lab Protect &, Alar, Xinjiang, Peoples R China
基金
中国国家自然科学基金;
关键词
Capillary electrophoresis; Electrospray ionization; Ion mobility spectrometry;
D O I
10.1002/elps.201800462
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In this work, ion mobility spectrometry (IMS) function as a detector and another dimension of separation was coupled with CE to achieve two-dimensional separation. To improve the performance of hyphenated CE-IMS instrument, electrospray ionization correlation ion mobility spectrometry is evaluated and compared with traditional signal averaging data acquisition method using tetraalkylammonium bromide compounds. The effect of various parameters on the separation including sample introduction, sheath fluid of CE and drift gas, data acquisition method of IMS were investigated. The experimental result shows that the optimal conditions are as follows: hydrodynamic sample injection method, the electrophoresis voltage is 10 kilo volts, 5 mmol/L ammonium acetate buffer solution containing 80% acetonitrile as both the background electrolyte and the electrospray ionization sheath fluid, the ESI liquid flow rate is 4.5 mu L/min, the drift voltage is 10.5 kilo volts, the drift gas temperature is 383 K and the drift gas flow rate is 300 mL/min. Under the above conditions, the mixture standards of seven tetraalkylammoniums can be completely separated within 10 min both by CE and IMS. The linear range was 5-250 mu g/mL, with LOD of 0.152, 0.204, 0.277, 0.382, 0.466, 0.623 and 0.892 mu g/mL, respectively. Compared with traditional capillary electrophoresis detection methods, the developed CE-ESI-IMS method not only provide two sets of qualitative parameters including electrophoresis migration time and ion drift time, ion mobility spectrometer can also provide an additional dimension of separation and could apply to the detection ultra-violet transparent compounds or none fluorescent compounds.
引用
收藏
页码:1639 / 1647
页数:9
相关论文
共 50 条
  • [21] Rapid analysis of underivatized fatty acids by electrospray-ionization-ion mobility spectrometry
    Tufariello, Jessica A.
    Grows, Kristopher
    Davis, Eric J.
    Harden, Charles S.
    Siems, William W.
    Hill, Herbert H., Jr.
    INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2015, 18 (1-2) : 95 - 104
  • [22] Effect of Multi Emitter Electrospray Array Ionization Source on Performance of Ion Mobility Spectrometry
    Liu Wen-Jie
    Li Guo-Zhu
    Luo Bi
    Meng Qin-Yan
    Lu Ya-Ling
    Ying Du-Ling
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2015, 43 (05) : 788 - 793
  • [23] Detection of inorganic ions from water by electrospray ionization-ion mobility spectrometry
    Dion, HM
    Ackerman, LK
    Hill, HH
    TALANTA, 2002, 57 (06) : 1161 - 1171
  • [24] DETERMINATION OF AMINO-ACIDS BY ONLINE CAPILLARY ELECTROPHORESIS-ELECTROSPRAY IONIZATION MASS-SPECTROMETRY
    LU, W
    YANG, G
    COLE, RB
    ELECTROPHORESIS, 1995, 16 (04) : 487 - 492
  • [25] Capillary electrophoresis electrospray ionization-mass spectrometry for peptidomics-based processing site determination
    Tsuchiya, Takashi
    Nakayama, Aya
    Kawamura, Takeshi
    Sasaki, Kazuki
    BIOCHEMICAL AND BIOPHYSICAL RESEARCH COMMUNICATIONS, 2020, 533 (04) : 872 - 878
  • [26] Capillary electrophoresis - electrospray spray ionization-mass spectrometry for the characterization of natural organic matter: An evaluation with free flow electrophoresis-off-line flow injection electrospray ionization-mass spectrometry
    Schmitt-Kopplin, P
    Kettrup, A
    ELECTROPHORESIS, 2003, 24 (17) : 3057 - 3066
  • [27] Isolation and characterization of β-cyclodextrin sulfates by preparative gradient polyacrylamide gel electrophoresis, capillary electrophoresis and electrospray ionization-mass spectrometry
    Hileman, RE
    Siegel, MM
    Tabei, K
    Balagurunathan, K
    Linhardt, RJ
    ELECTROPHORESIS, 1998, 19 (15) : 2677 - 2681
  • [28] Evaluation of volatile eluents and electrolytes for high-performance liquid chromatography-electrospray ionization mass spectrometry and capillary electrophoresis-electrospray ionization mass spectrometry of proteins - II. Capillary electrophoresis
    Huber, CG
    Premstaller, A
    Kleindienst, G
    JOURNAL OF CHROMATOGRAPHY A, 1999, 849 (01) : 175 - 189
  • [29] Rapid separation of phosphopeptides by microchip electrophoresis-electrospray ionization mass spectrometry
    Ollikainen, Elisa
    Bonabi, Ashkan
    Nordman, Nina
    Jokinen, Ville
    Kotiaho, Tapio
    Kostiainen, Risto
    Sikanen, Tiina
    JOURNAL OF CHROMATOGRAPHY A, 2016, 1440 : 249 - 254
  • [30] Fourier transform electrospray ion mobility spectrometry
    Chen, YH
    Siems, WF
    Hill, HH
    ANALYTICA CHIMICA ACTA, 1996, 334 (1-2) : 75 - 84