Microextraction by Packed Molecularly Imprinted Polymer Combined Ultra-High-Performance Liquid Chromatography for the Determination of Levofloxacin in Human Plasma

被引:24
作者
Meng, Jia [1 ,2 ]
Wang, Xu [3 ]
机构
[1] Tianjin Med Univ, Hosp 2, Dept Kidney Dis, Tianjin 300211, Peoples R China
[2] Tianjin Med Univ, Hosp 2, Blood Purificat Ctr, Tianjin 300211, Peoples R China
[3] Tianjin Med Univ, Hosp 2, Tianjin Inst Urol, Dept Urol, Tianjin 300211, Peoples R China
关键词
DEEP EUTECTIC SOLVENT; SIMULTANEOUS QUANTIFICATION; LOCAL-ANESTHETICS; SORBENT MEPS; MOXIFLOXACIN; ANTIBIOTICS; PROSTATE; SAMPLES; PURIFICATION; GATIFLOXACIN;
D O I
10.1155/2019/4783432
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Fluoroquinolones are considered as gold standard for the prevention of bacterial infections. To improve assessment of antibacterial efficacy, a novel method for determination of levofloxacin was developed and validated. Deep eutectic solvents (DESs) as only green solvent were used as a porogen for preparation of water-compatible molecularly imprinted polymers (MIPs) with a pseudotemplate. The DESs-MIPs were characterized in detail, including scanning electron microscope, nitrogen sorption porosimetry, and Fourier transform-infrared spectra. Clearly, the maximum binding capacity of levofloxacin on DESs-MIPs in water and methanol was 0.216 and 0.077 mu mol g(-1), respectively. The DESs-MIPs as adsorbing materials were applied in microextraction by packed sorbent (MEPS), and the DESs-MIPs-MEPS conditions were optimized. The DESs-MIPs-MEPS coupled with ultra-high-performance liquid chromatography (UHPLC) was used to determine levofloxacin in human plasma. The method was found linear over 0.05-10 mu g mL(-1) with coefficient of correlation equal to 0.9988. The limit of detection and limit of quantification were 0.012 and 0.04 mu g mL(-1), respectively. At three spiked levels, the precision of proposed method was between 95.3% and 99.7% with intraday and interday relative standard deviations <= 8.9%. Finally, the developed method was used to examine levofloxacin from human plasma of 20 hospitalized patients after transrectal ultrasound-guided prostate biopsy, and the average concentration (+/- SD) of levofloxacin was 2.35 +/- 0.99 mu g mL(-1) in plasma.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Preparations of magnetic molecularly imprinted polymer for selective recognition and determination of 4-methylimidazole in soft beverage by high performance liquid chromatography
    Ye, Helin
    Chen, Xiaofen
    Feng, Zufei
    ADSORPTION SCIENCE & TECHNOLOGY, 2017, 35 (1-2) : 37 - 54
  • [22] Highly sensitive determination of cyromazine, melamine, and their metabolites in milk by molecularly imprinted solid-phase extraction combined with ultra-performance liquid chromatography
    Ge, Xusheng
    Wu, Xingqiang
    Wang, Junmin
    Liang, Shuxuan
    Sun, Hanwen
    JOURNAL OF DAIRY SCIENCE, 2015, 98 (04) : 2161 - 2171
  • [23] Molecularly imprinted solid-phase extraction for the selective determination of bromhexine in human serum and urine with high performance liquid chromatography
    Javanbakht, Mehran
    Namjumanesh, Mohammad Hadi
    Akbari-Adergani, Behrouz
    TALANTA, 2009, 80 (01) : 133 - 138
  • [24] Ultra Performance Liquid Chromatography PDA Method for Determination of Tigecycline in Human Plasma
    D'Avolio, Antonio
    Peila, Emanuela
    Simiele, Marco
    Pensi, Debora
    Baietto, Lorena
    Cusato, Jessica
    Cinnirella, Giacoma
    De Rosa, Francesco
    Di Perri, Giovanni
    THERAPEUTIC DRUG MONITORING, 2013, 35 (06) : 853 - 858
  • [25] Molecularly imprinted solid-phase extraction coupled with ultra high performance liquid chromatography and fluorescence detection for the determination of estrogens and their metabolites in wastewater
    Guedes-Alonso, Rayco
    Santana-Viera, Sergio
    Sosa-Ferrera, Zoraida
    Juan Santana-Rodriguez, Jose
    JOURNAL OF SEPARATION SCIENCE, 2015, 38 (22) : 3961 - 3968
  • [26] Determination of Ciprofloxacin, Enrofloxacin, and Marbofloxacin in Bovine Urine, Serum, and Milk by Microextraction by a Packed Sorbent Coupled to Ultra-High Performance Liquid Chromatography
    Aresta, Antonella
    Cotugno, Pietro
    Zambonin, Carlo
    ANALYTICAL LETTERS, 2019, 52 (05) : 790 - 802
  • [27] A novel molecularly imprinted polymer for simultaneous extraction and determination of sudan dyes by on-line solid phase extraction and high performance liquid chromatography
    Zhao, Chuande
    Zhao, Ting
    Liu, Xiaoyan
    Zhang, Haixia
    JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (45) : 6995 - 7002
  • [28] Determination of Chlorpyrifos in Rice Based on Magnetic Molecularly Imprinted Polymers Coupled with High-Performance Liquid Chromatography
    Ma, Guifu
    Chen, Ligang
    FOOD ANALYTICAL METHODS, 2014, 7 (02) : 377 - 388
  • [29] Molecularly imprinted solid-phase extraction for determination of tilmicosin in feed using high performance liquid chromatography
    Zheng, Yaqiu
    Liu, Yahong
    Guo, Hongbin
    He, Limin
    Fang, Binghu
    Zeng, Zhenling
    ANALYTICA CHIMICA ACTA, 2011, 690 (02) : 269 - 274
  • [30] Molecularly imprinted solid-phase extraction for the selective determination of valnemulin in feeds with high performance liquid chromatography
    Guo, Hongbin
    Liu, Kaiyong
    Liu, Yahong
    Fang, Binghu
    Liu, Min
    He, Limin
    Zeng, Zhenling
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2011, 879 (02): : 181 - 185