Which feature influences on-eye power change of soft toric contact lenses: Design or corneal shape?

被引:7
作者
Doll, Tamsin [1 ]
Moore, Joshua [1 ,2 ]
Shihab, Ahmad H. [1 ,3 ]
Lopes, Bernardo T. [1 ,4 ]
Eliasy, Ashkan [1 ]
Maklad, Osama [1 ]
Wu, Richard [5 ,6 ]
White, Lynn [7 ]
Jones, Steve [1 ]
Elsheikh, Ahmed [1 ,8 ,9 ,10 ]
Abass, Ahmed [1 ]
机构
[1] Univ Liverpool, Sch Engn, Liverpool, Merseyside, England
[2] Univ Liverpool, Dept Math Sci, Liverpool, Merseyside, England
[3] Univ Hertfordshire, Sch Engn, Hatfield, Herts, England
[4] Univ Fed Sao Paulo, Dept Ophthalmol, Sao Paulo, Brazil
[5] Cent Taiwan Univ Sci & Technol, Dept Optometry, Taichung, Taiwan
[6] Pacific Univ, Coll Optometry, Forest Grove, OR USA
[7] Ultravis CLPL, Leighton Buzzard, England
[8] Beihang Univ, Sch Biol Sci & Biomed Engn, Beijing, Peoples R China
[9] Moorfields Eye Hosp NHS Fdn Trust, Natl Inst Hlth Res NIHR Biomed Res Ctr, London, England
[10] UCL Inst Ophthalmol, London, England
来源
PLOS ONE | 2020年 / 15卷 / 11期
关键词
BIOMECHANICAL PROPERTIES;
D O I
10.1371/journal.pone.0242243
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose This study investigates how both the peripheral zone design and corneal shape affect the behaviour of soft contact lenses on-eye. Methods In this study, soft contact lenses of varying nominal cylindrical powers and peripheral zone designs-a single-prism gravity-based stabilised lens (G1P), two-prism blink-based stabilised lens (B2P) and four-prism blink-based stabilised lens (B4P)-were generated as finite element models. The on-eye simulation results were analysed to identify the impact of each peripheral zone design (Each with different volume ratios) on the effective power change (EPC) when worn by a subject. Topographies of three eyes of varying average simulated anterior corneal curvature (flat, average & steep) were used in this study. Results The volume of the lens's peripheral zone as a ratio of the total lens volume (V-p) recorded very weak correlations with the effective power change (EPC) among the three investigated designs when they were fitted to the flat eye (R = -0.19, -0.15 & -0.22 respectively), moderate correlations with the average eye (R = 0.42, 0.43 & 0.43 respectively) and strong correlations with the steep eye (R = 0.91, 0.9 & 0.9 respectively). No significant differences were noticed among the three investigated designs and none of the cylindrical lenses designed with axis 90 degrees recorded EPC values outside the acceptance criteria range (ACR) of +/- 0.25 D. No significant differences in EPC were recorded among the three designs G1P, B2P and B4P (p>0.6) when they were designed with three axes at 90 degrees, 45 degrees and 0 degrees. Moving the toric lens axis away from 90 degrees dragged the EPC to the negative side and most of the investigated lenses with axes at 45 degrees and 0 degrees recorded EPCs outside the +/- 0.25D range. Conclusions In all cases, the shape of the cornea had a more dominant effect on EPC when compared to the peripheral zone design. Corneal shape influences the soft toric contact lens's on-eye power change more than the lens design.
引用
收藏
页数:23
相关论文
共 60 条
  • [1] Artefact-free topography based scleral-asymmetry
    Abass, Ahmed
    Lopes, Bernardo T.
    Eliasy, Ashkan
    Salomao, Marcella
    Wu, Richard
    Whites, Lynn
    Jones, Steve
    Clamps, John
    Ambrosio, Renato, Jr.
    Elsheikh, Ahmed
    [J]. PLOS ONE, 2019, 14 (07):
  • [2] Simulated optical performance of soft contact lenses on the eye
    Abass, Ahmed
    Stuart, Samantha
    Lopes, Bernardo T.
    Zhou, Dong
    Geraghty, Brendan
    Wu, Richard
    Jones, Steve
    Flux, Ilse
    Stortelder, Reinier
    Snepvangers, Arnoud
    Leca, Renato
    Elsheikh, Ahmed
    [J]. PLOS ONE, 2019, 14 (05):
  • [3] Three-dimensional non-parametric method for limbus detection
    Abass, Ahmed
    Lopes, Bernardo T.
    Eliasy, Ashkan
    Wu, Richard
    Jones, Steve
    Clamp, John
    Ambrosio, Renato, Jr.
    Elsheikh, Ahmed
    [J]. PLOS ONE, 2018, 13 (11):
  • [4] Comparison of central corneal thickness measurements by Pentacam, noncontact specular microscope, and ultrasound pachymetry in normal and post-LASIK eyes
    Al-Ageel, Saleh
    Al-Muammar, Abdulrahman M.
    [J]. SAUDI JOURNAL OF OPHTHALMOLOGY, 2009, 23 (3-4) : 181 - 187
  • [5] [Anonymous], 2018, OBSTET GYNECOL, V13, pe49
  • [6] Bennett AG, 1985, OPTICS CONTACT LENSE
  • [7] Corneal topography in keratoconus: state of the art
    Cavas-Martinez, F.
    De la Cruz Sanchez, E.
    Nieto Martinez, J.
    Fernandez Canavate, F. J.
    Fernandez-Pacheco, D. G.
    [J]. EYE AND VISION, 2016, 3
  • [8] Davis RL, 2005, CONTACT LENS SPECTRU
  • [9] Influence of Contact Lens Power Profile on Peripheral Refractive Error
    de la Jara, Percy Lazon
    Sankaridurg, Padmaja
    Ehrmann, Klaus
    Holden, Brien A.
    [J]. OPTOMETRY AND VISION SCIENCE, 2014, 91 (06) : 642 - 649
  • [10] Efron N, 2016, SOFT LENS DESIGN FIT, P86