Crystal surface and reflector optimization for the SiPM-based dual-ended readout TOF-DOI PET detector

被引:34
作者
Kang, Han Gyu [1 ]
Yamaya, Taiga [1 ]
Han, Young Been [2 ]
Song, Seong Hyun [2 ]
Ko, Guen Bae [3 ,4 ]
Lee, Jae Sung [3 ,4 ,5 ]
Hong, Seong Jong [2 ,6 ]
机构
[1] Natl Inst Quantum & Radiol Sci & Technol QST, Natl Inst Radiol Sci NIRS, Inage Ku, 4-9-1 Anagawa, Chiba, Japan
[2] Eulji Univ, Grad Sch, Dept Senior Healthcare, Daejeon, South Korea
[3] Brightonix Imaging Inc, Seoul, South Korea
[4] Seoul Natl Univ, Inst Radiat Med, Med Res Ctr, Seoul, South Korea
[5] Seoul Natl Univ, Dept Nucl Med, Coll Med, Seoul, South Korea
[6] Eulji Univ, Dept Radiol Sci, Gyeonggi Do, South Korea
基金
新加坡国家研究基金会;
关键词
silicon photomultiplier (SiPM); time-of-flight (TOF); depth-of-interaction (DOI); dual-ended readout; scintillation detector; TIME-OF-FLIGHT; DEPTH; PERFORMANCE; RESOLUTION; DESIGN; BLOCK;
D O I
10.1088/2057-1976/abc45a
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Silicon photomultipliers (SiPMs) are now widely used for positron emission tomography (PET) applications because of their high gain and low noise characteristics. The PET image quality has been improved with the advancement of time-of-flight (TOF) and depth-of-interaction (DOI) measurement techniques. For brain-dedicated PET systems, both TOF and DOI information are beneficial for enhancing the reconstructed PET image quality. In a previous study, we proposed SiPM-based dual-ended readout PET detectors that used a mean time method to achieve coincidence timing resolution (CTR) of 349 ps and DOI resolution of 2.9 mm. However, the coincidence timing resolution (CTR) was worse than 300 ps since the crystal surface and the reflector type were not optimized. This study aimed at investigating the optimal crystal surface treatment and the reflector material to achieve a sub-200 ps CTR and sub-3 mm DOI resolution with a dual-ended readout PET detector using an LYSO crystal (2.9 x 2.9 x 20 mm(3)). The scintillation light inside the LYSO crystal was read out by two SiPMs using the dual-ended readout method. The CTR and DOI resolution were measured with two different crystal surfaces (polished and saw-cut) and three different reflector material scenarios of ESR without grease (i.e., air coupling), ESR with optical grease and Teflon. We digitized the timing and energy signals by using a V775N TDC module (35 ps bit(-1)) and V965 QDC module, respectively. The combination of the saw-cut LYSO crystal and the ESR with air coupling resulted in the best CTR (188 32 ps) and DOI resolution (2.9 0.2 mm) with the dual-ended readout configuration. We concluded the dual-ended readout method in combination with the saw-cut crystal and the ESR reflector with air coupling can provide a sub-200 ps CTR and sub-3.0 mm DOI resolution simultaneously.
引用
收藏
页数:12
相关论文
共 42 条
[1]   Design and Detector Performance of the PET Component of the TRIMAGE PET/MR/EEG Scanner [J].
Belcari, Nicola ;
Bisogni, Maria Giuseppina ;
Camarlinghi, Niccolo ;
Carra, Pietro ;
Cerello, Piergiorgio ;
Morrocchi, Matteo ;
Patera, Alessandra ;
Sportelli, Giancarlo ;
Del Guerra, Alberto .
IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2019, 3 (03) :292-301
[2]   A combined time-of-flight and depth-of-interaction detector for total-body positron emission tomography [J].
Berg, Eric ;
Roncali, Emilie ;
Kapusta, Maciej ;
Du, Junwei ;
Cherry, Simon R. .
MEDICAL PHYSICS, 2016, 43 (02) :939-950
[3]   Optimizing light transport in scintillation crystals for time-of-flight PET: an experimental and optical Monte Carlo simulation study [J].
Berg, Eric ;
Roncali, Emilie ;
Cherry, Simon R. .
BIOMEDICAL OPTICS EXPRESS, 2015, 6 (06) :2220-2230
[4]   Physical Performance of the new hybrid PET/CT Discovery-690 [J].
Bettinardi, V. ;
Presotto, L. ;
Rapisarda, E. ;
Picchio, M. ;
Gianolli, L. ;
Gilardi, M. C. .
MEDICAL PHYSICS, 2011, 38 (10) :5394-5411
[5]   Sub-3 mm, near-200 ps TOF/DOI-PET imaging with monolithic scintillator detectors in a 70 cm diameter tomographic setup [J].
Borghi, Giacomo ;
Tabacchini, Valerio ;
Bakker, Rene ;
Schaart, Dennis R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (15)
[6]   DESIGN OF A HIGH-RESOLUTION POSITRON EMISSION TOMOGRAPH USING SOLID-STATE SCINTILLATION DETECTORS [J].
CARRIER, C ;
MARTEL, C ;
SCHMITT, D ;
LECOMTE, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :685-690
[7]   Time-over-threshold for pulse shape discrimination in a time-of-flight phoswich PET detector [J].
Chang, Chen-Ming ;
Cates, Joshua W. ;
Levin, Craig S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (01) :258-271
[8]   SiPM-based dual-ended-readout DOI-TOF PET module based on mean-time method [J].
Han, Y. B. ;
Kang, H. G. ;
Song, S. H. ;
Ko, G. B. ;
Lee, J. S. ;
Hong, S. J. .
JOURNAL OF INSTRUMENTATION, 2019, 14 (02)
[9]   Characterization of a 64 channel PET detector using photodiodes for crystal identification [J].
Huber, JS ;
Moses, WW ;
Derenzo, SE ;
Ho, MH ;
Andreaco, MS ;
Paulus, MJ ;
Nutt, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1197-1201
[10]   Continuous depth-of-interaction measurement in a single-layer pixelated crystal array using a single-ended readout [J].
Ito, Mikiko ;
Lee, Min Sun ;
Lee, Jae Sung .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (05) :1269-1282