Improved lung edema monitoring with coronary vein pacing leads: a simulation study

被引:16
作者
Belalcazar, A [1 ]
Patterson, RP [1 ]
机构
[1] Univ Minnesota, Inst Biomed Engn, Bakken Med Instrumentat & Devide Lab MIND, Minneapolis, MN 55455 USA
关键词
impedance; pacemakers; cardiac resynchronization therapy; congestive heart failure; edema;
D O I
10.1088/0967-3334/25/2/007
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This computer simulation study compared the ability of left ventricular coronary vein (LV) pacemaker leads against right ventricular (RV) and right atrial (RA) leads to monitor lung edema using electrical impedance measurements. MRI images were used to construct electrical models of the thorax. Four lead configurations were tested with increases of pulmonary edema, intravascular fluids and heart dilation. The impedance changes observed at end systole with severe lung edema were 8.5%, 11.2%, 12.3% and 26.8% for the RA, RV, RV coil and LV configurations, respectively. Sensitivities in ohms per litre of lung fluid were 19.15, 19.15, 25.07 and 52.11 for the same configurations. The impedance changes for intravascular fluid overload with constant lung status were 1%, 1.3%, 9.2% and 6.4% while the sensitivities were 2, 2, 17 and 11 ohms per litre of intravascular fluid, respectively. Regional analysis of the thoracic sources of impedance revealed a high sensitivity near pacing electrodes and generator, and a low sensitivity to the right lung and all pulmonary vessels. Simulations showed that LV leads have a threefold advantage in sensitivity when monitoring lung edema in comparison to conventional RV leads. To monitor vascular and lung fluids independently, combined impedance configurations may be used. Regional sensitivities must be taken into account for proper clinical interpretation of impedance changes.
引用
收藏
页码:475 / 487
页数:13
相关论文
共 20 条
[1]  
BELALACAZAR A, 2003, PACING CLIN ELECTROP, V26, P933
[2]   Measurement of minute ventilation with different DDDR pacemaker electrode configurations [J].
Bonnet, JL ;
Ritter, P ;
Pioger, G .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (01) :4-10
[3]  
de Jongh AL, 1999, PACE, V22, P152, DOI 10.1111/j.1540-8159.1999.tb00323.x
[4]   Rate responsive pacing using transthoracic impedance minute ventilation sensors: A multicenter study on calibration stability [J].
Duru, F ;
Cho, Y ;
Wilkoff, BL ;
Cole, CR ;
Adler, S ;
Jensen, DN ;
Strobel, U ;
Radicke, D ;
Candinas, R .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (12) :1679-1684
[5]  
Duru F, 2000, PACE, V23, P1767
[6]   Impact of a comprehensive heart failure management program on hospital readmission and functional status of patients with advanced heart failure [J].
Fonarow, GC ;
Stevenson, LW ;
Walden, JA ;
Livingston, NA ;
Steimle, AE ;
Hamilton, MA ;
Moriguchi, J ;
Tillisch, JH ;
Woo, MA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1997, 30 (03) :725-732
[7]   SPECIFIC RESISTANCE OF BIOLOGICAL MATERIAL-A COMPENDUM OF DATA FOR BIOMEDICAL ENGINEER AND PHYSIOLOGIST [J].
GEDDES, LA ;
BAKER, LE .
MEDICAL & BIOLOGICAL ENGINEERING, 1967, 5 (03) :271-&
[9]   ACC/AHA guidelines for the evaluation and management of chronic heart failure in the adult: Executive summary - A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee to revise the 1995 Guidelines for the Evaluation and Management of Heart Failure) [J].
Hunt, SA ;
Baker, DW ;
Chin, MH ;
Cinquegrani, MP ;
Feldman, AM ;
Francis, GS ;
Ganiats, TG ;
Goldstein, S ;
Gregoratos, G ;
Jessup, ML ;
Noble, RJ ;
Packer, M ;
Silver, MA ;
Stevenson, LW ;
Gibbons, RJ ;
Antman, EM ;
Alpert, JS ;
Faxon, DP ;
Fuster, V ;
Gregoratos, G ;
Jacobs, AK ;
Hiratzka, LF ;
Russell, RO ;
Smith, SC .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 38 (07) :2101-2113
[10]   Sensitivity distributions of impedance cardiography using band and spot electrodes analyzed by a three-dimensional computer model [J].
Kauppinen, PK ;
Hyttinen, JA ;
Malmivuo, JA .
ANNALS OF BIOMEDICAL ENGINEERING, 1998, 26 (04) :694-702