Toward practical application of paper-based microfluidics for medical diagnostics: state-of-the-art and challenges

被引:358
作者
Yamada, Kentaro [1 ]
Shibata, Hiroyuki [1 ]
Suzuki, Koji [1 ]
Citterio, Daniel [1 ]
机构
[1] Keio Univ, Dept Appl Chem, Kohoku Ku, 3-14-1 Hiyoshi, Yokohama, Kanagawa 2238522, Japan
基金
日本学术振兴会;
关键词
LAB-ON-PAPER; SIGNAL AMPLIFICATION STRATEGY; PROGRAMMING MULTISTEP ASSAYS; LATERAL FLOW IMMUNOASSAY; BLOOD-PLASMA SEPARATION; ANALYTICAL DEVICES LPAD; FLUIDIC TIME DELAYS; HUMAN SERUM-ALBUMIN; SOL-GEL MATERIALS; LOW-COST;
D O I
10.1039/c6lc01577h
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Microfluidic paper-based analytical devices (mu PADs) have emerged as a promising diagnostic platform a decade ago. In contrast to highly active academic developments, their entry into real-life applications is still very limited. This discrepancy is attributed to the gap between research developments and their practical utility, particularly in the aspects of operational simplicity, long-term stability of devices, and associated equipment. On the basis of these backgrounds, this review attempts to: 1) identify the reasons for success of paper-based devices already in the market, 2) describe the current status and remaining issues of mu PADs in terms of operational complexity, signal interpretation approaches, and storage stability, and 3) discuss the possibility of mass production based on established manufacturing technologies. Finally, the state-of-the-art in commercialisation of mu PADs is discussed, and the "upgrades" required from a laboratory-based prototype to an end user device are demonstrated on a specific example.
引用
收藏
页码:1206 / 1249
页数:44
相关论文
共 293 条
[1]   Inkjet-printed microfluidic multianalyte chemical sensing paper [J].
Abe, Koji ;
Suzuki, Koji ;
Citterio, Daniel .
ANALYTICAL CHEMISTRY, 2008, 80 (18) :6928-6934
[2]  
Alere, BINAXNOW INFL A B CA
[3]   Colorimetric Paper Bioassay for the Detection of Phenolic Compounds [J].
Alkasir, Ramiz S. J. ;
Ornatska, Maryna ;
Andreescu, Silvana .
ANALYTICAL CHEMISTRY, 2012, 84 (22) :9729-9737
[4]   Lateral-flow immunoassays for mycotoxins and phycotoxins: a review [J].
Anfossi, Laura ;
Baggiani, Claudio ;
Giovannoli, Cristina ;
D'Arco, Gilda ;
Giraudi, Gianfranco .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2013, 405 (2-3) :467-480
[5]  
[Anonymous], Laboratory Biosafety Manual
[6]  
[Anonymous], ERICSSON MOBILITY RE
[7]   Development of automated paper-based devices for sequential multistep sandwich enzyme-linked immunosorbent assays using inkjet printing [J].
Apilux, Amara ;
Ukita, Yoshiaki ;
Chikae, Miyuki ;
Chailapakul, Orawon ;
Takamura, Yuzuru .
LAB ON A CHIP, 2013, 13 (01) :126-135
[8]   STABILIZATION OF PROTEIN-STRUCTURE BY SUGARS [J].
ARAKAWA, T ;
TIMASHEFF, SN .
BIOCHEMISTRY, 1982, 21 (25) :6536-6544
[9]  
Armstrong P., IN VITRO DIAGNOSTIC
[10]   ENZYMES AND OTHER PROTEINS ENTRAPPED IN SOL-GEL MATERIALS [J].
AVNIR, D ;
BRAUN, S ;
LEV, O ;
OTTOLENGHI, M .
CHEMISTRY OF MATERIALS, 1994, 6 (10) :1605-1614