Simple technique to increase the linearity and field-of-view in position sensitive photomultiplier tubes

被引:25
作者
Clancy, RL [1 ]
Thompson, CJ [1 ]
Robar, JL [1 ]
Bergman, AM [1 ]
机构
[1] MCGILL UNIV,MONTREAL NEUROL INST,MED PHYS UNIT,MONTREAL,PQ,CANADA
基金
加拿大自然科学与工程研究理事会;
关键词
D O I
10.1109/23.603698
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Crossed anode wire position sensitive photomultiplier tubes (PS-PMTs) detect the location of a light source and provide the X and Y axis coordinates of the event. These coordinates are typically generated using Anger logic, where a resistor chain divides the current flow into two signals for each coordinate (X+, X- & Y+, Y-). In the standard readout, identical resistor values are used across the entire resistor chain. While this arrangement provides a linear readout in the central portion of the photomultiplier face, the readout is non-linear and sometimes even double valued near the edges of the PS-PMT due to the truncation of the charge beyond the last anode wire. To counter this effect, we have increased the value of the resistance near the ends of each resistor chain in order to compensate for the charge lost beyond the anode wires. Measurements were made using a Hamamatsu R-3941 PS-PMT coupled to a pixellated EGO matrix of cut crystals with a 2 mm pitch in each direction. After changing the end resistors, the usable field-of-view increased by 39%. This simple modification should enhance the operation of PS-PMTs in applications such as positron emission mammography, and small animal PET imaging.
引用
收藏
页码:494 / 498
页数:5
相关论文
共 13 条
[1]   SCINTILLATION CAMERA [J].
ANGER, HO .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1958, 29 (01) :27-33
[2]   MULTICHANNEL READOUT OF CROSSED-WIRE ANODE PHOTOMULTIPLIERS [J].
BIRD, AJ ;
HE, Z ;
RAMSDEN, D .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 348 (2-3) :668-672
[3]  
CARRIER C, 1988, THESIS U SHERBROOKE
[4]   COLLECTION OF SCINTILLATION LIGHT FROM SMALL BGO CRYSTALS [J].
CHERRY, SR ;
SHAO, YP ;
TORNAI, MP ;
SIEGEL, S ;
RICCI, AR ;
PHELPS, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1058-1063
[5]  
CLANCY RL, 1995, IEEE NUCL SCI S C, P409
[6]   POSITION-SENSITIVE PHOTOMULTIPLIER TUBES FOR SCINTILLATION IMAGING [J].
KUME, H ;
MURAMATSU, S ;
IIDA, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :359-363
[7]  
ROBAR JL, 1996, THESIS MCGILL U
[8]  
ROBAR JL, UNPUB NUCL INST ME A
[9]   POSITRON EMISSION MAMMOGRAPHY (PEM) - A PROMISING TECHNIQUE FOR DETECTING BREAST-CANCER [J].
THOMPSON, CJ ;
MURTHY, K ;
PICARD, Y ;
WEINBERG, IN ;
MAKO, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1012-1017
[10]  
THOMPSON CJ, 1995, IEEE MED IM C, P1074