Outcome of the pivotal study of the Aptus endovascular abdominal aortic aneurysms repair system

被引:30
|
作者
Mehta, Manish [1 ]
Henretta, John [2 ]
Glickman, Marc [3 ]
Deaton, David [4 ]
Naslund, Thomas C. [5 ]
Gray, Bruce [6 ]
McCann, Richard [7 ]
Jordan, William [8 ]
Fairman, Ronald [9 ]
机构
[1] Albany Med Coll, Dept Vasc & Endovasc Surg, Albany, NY 12208 USA
[2] Mission Hlth, Dept Vasc Surg, Asheville, NC USA
[3] Sentara Heart Hosp, Dept Vasc Surg, Norfolk, VA USA
[4] Georgetown Univ, Dept Vasc Surg, Washington, DC USA
[5] Vanderbilt Univ, Dept Vasc Surg, Nashville, TN 37235 USA
[6] Greenville Hosp Syst, Dept Vasc Surg, Greenville, SC USA
[7] Duke Univ, Dept Vasc Surg, Durham, NC 27706 USA
[8] Univ Alabama Birmingham, Dept Vasc Surg, Birmingham, AL 35487 USA
[9] Univ Penn, Dept Vasc Surg & Endovasc Therapy, Philadelphia, PA USA
关键词
MULTICENTER TRIAL; LIFELINE REGISTRY; CLINICAL-TRIAL; STENT GRAFT; COMPLICATIONS; NECKS;
D O I
10.1016/j.jvs.2014.02.017
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: Endovascular treatment of abdominal aortic aneurysm (AAA) is associated with benefits over open surgery, yet limitations remain with current endovascular devices. This study was performed to assess outcomes of AAA repair with the Aptus endograft and EndoAnchors (Aptus Endosystems, Sunnyvale, Calif). Methods: This prospective, multicenter, single-arm investigational device exemption trial was conducted at 25 sites in the United States. A total of 155 patients were enrolled in the trial (mean age, 73 +/- 8 years; male, 93.5%; mean AAA diameter, 53.6 +/- 7.9 mm). The Aptus endograft is a two-component system: a multilumen, modular endograft with two docking limbs (Aptus Endograft System) and the Heli-FX Aortic Securement System comprising an electronically controlled applier (Heli-FX Applier) with helical EndoAnchors provided in a cassette and a deflectable sheath (Heli-FX Guide) designed for delivery of the applier to the target location for EndoAnchor implantation. The main eligibility criteria included proximal neck length of >= 12 mm, diameter of 19 to 29 mm, and infrarenal angulation of <= 60 degrees. The primary safety end point was freedom from major adverse events at 30 days, and the primary effectiveness end point was successful aneurysm treatment at 12 months. Thrombus-related events (TRE) were defined as limb occlusion or thromboembolism from the endograft. Subjects were observed for a median of 4.2 years, with imaging end points analyzed by a core laboratory and adverse events adjudicated by a clinical events committee. Results: Among 155 enrolled subjects, 153 (98.7%) underwent successful implantation of the Aptus endograft and a median of five EndoAnchors; two subjects were converted to open surgical repair during the initial procedure. Overall, the primary safety and effectiveness end points were met in 98.1% and 97.4% of the subjects, respectively. Aneurysm-related mortality was 0.6%, with one postdischarge cardiac death 18 days after implantation. There were noAAAruptures. There were no fractures of stents or EndoAnchors. There was one type I endoleak and one type III endoleak. Stent graft migration was noted in five subjects, none associated with sac enlargement, type I endoleak, or EndoAnchor dislocation from the endograft. AAA sac shrinkage of >= 5 mm at 1, 2, and 3 years was observed in 60.3%, 72.9%, and 81.7%, respectively. Sixty-one subjects (39.4%) experienced 113 TRE, associated with 80 reinterventions (in 58 subjects) unassociated with limb loss or death. A root cause analysis of TRE identified small, out-of-specification docking limbs with graft infolding and high local shear, resulting in thrombus formation within the endograft with subsequent distal embolization in some cases. Conclusions: Early results of the Aptus endograft trial met its safety and effectiveness end points; however, a high rate of TRE was observed because of manufacturing discrepancies. The findings confirm a low rate of type I and type III endoleaks, migration, and non-TRE reintervention with a high rate of aneurysm sac regression during midterm follow-up.
引用
收藏
页码:275 / 285
页数:11
相关论文
共 50 条
  • [41] Our early and mid-term results for endovascular repair of abdominal aortic aneurysms
    Karakisi, Sedat Ozan
    Ergene, Saban
    Hemsinli, Dogus
    Kucuker, Seref Alp
    TURK GOGUS KALP DAMAR CERRAHISI DERGISI-TURKISH JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2019, 27 (01): : 29 - 34
  • [42] Comparison of midterm results of endovascular aneurysm repair for ruptured and elective abdominal aortic aneurysms
    Oliveira-Pinto, Jose
    Soares-Ferreira, Rita
    Oliveira, Nelson F. G.
    Bastos Goncalves, Frederico M.
    Hoeks, Sanne
    Van Rijn, Marie Josee
    Ten Raa, Sander
    Mansilha, Armando
    Verhagen, Hence J. M.
    JOURNAL OF VASCULAR SURGERY, 2020, 71 (05) : 1554 - +
  • [43] Patient-specific in silico endovascular repair of abdominal aortic aneurysms: application and validation
    André Hemmler
    Brigitta Lutz
    Günay Kalender
    Christian Reeps
    Michael W. Gee
    Biomechanics and Modeling in Mechanobiology, 2019, 18 : 983 - 1004
  • [44] Contemporary Results of Endovascular Repair of Abdominal Aortic Aneurysms: Effect of Anatomical Fixation on Outcomes
    Carpenter, Jeffrey P.
    Garcia, Mark J.
    Harlin, Stuart A.
    Jordan, William D., Jr.
    Jung, Matthew T.
    Krajcer, Zvonimir
    Rodriguez-Lopez, Julio A.
    JOURNAL OF ENDOVASCULAR THERAPY, 2010, 17 (02) : 153 - 162
  • [45] Reintervention Rate after Open Surgery and Endovascular Repair for Nonruptured Abdominal Aortic Aneurysms
    Hwang, Deokbi
    Park, Sujin
    Kim, Hyung-Kee
    Lee, Jong-Min
    Huh, Seung
    ANNALS OF VASCULAR SURGERY, 2017, 43 : 134 - 143
  • [46] Endovascular stent graft repair of abdominal aortic aneurysms: Current status and future directions
    Sun, Zhonghua
    WORLD JOURNAL OF RADIOLOGY, 2009, 1 (01): : 63 - 71
  • [47] Endovascular Aneurysm Repair of Abdominal Aortic Aneurysms: Standards, Technical Options and Advanced Indications
    Ketelsen, D.
    Thomas, C.
    Schmehl, J.
    Koenig, C. W.
    Syha, R.
    Rittig, K.
    Balletshofer, B.
    Claussen, C. D.
    Brechtel, K.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2014, 186 (04): : 337 - 347
  • [48] Translumbar embolization of type 2 endoleaks after endovascular repair of abdominal aortic aneurysms
    Baum, RA
    Cope, C
    Fairman, RM
    Carpenter, JP
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 12 (01) : 111 - 116
  • [49] Optimal Technique for Imaging Iliac Segments During Endovascular Repair of Abdominal Aortic Aneurysms
    Oshin, Olufemi A.
    Bown, Matthew J.
    McWilliams, Richard G.
    Brennan, John A.
    Vallabhaneni, S. Rao
    JOURNAL OF ENDOVASCULAR THERAPY, 2009, 16 (03) : 373 - 379
  • [50] Pathophysiology of proximal perigraft endoleak following endovascular repair of abdominal aortic aneurysms: a study using a flow model
    Albertini, JN
    Macierewicz, JA
    Yusuf, SW
    Wenham, PW
    Hopkinson, BR
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2001, 22 (01) : 53 - 56