Comparative Evaluation of Stress Distribution in Experimentally Designed Nickel-titanium Rotary Files with Varying Cross Sections: A Finite Element Analysis

被引:15
|
作者
Ahamed, Shadir Bughari Basheer [1 ]
Vanajassun, Purushothaman Pranav [1 ]
Rajkumar, Kothandaraman [1 ]
Mahalaxmi, Sekar [1 ]
机构
[1] SRM Dent Coll, Dept Conservat Dent & Endodont, Madras 600089, Tamil Nadu, India
关键词
Bending stress; cross-sectional design; finite element analysis; nickel-titanium rotary; BENDING PROPERTIES; INSTRUMENTS; MODELS; RESISTANCE; FRACTURE;
D O I
10.1016/j.joen.2017.12.013
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: Single cross-sectional nickel-titanium (NiTi) rotary instruments during continuous rotations are subjected to constant and variable stresses depending on the canal anatomy. This study was intended to create 2 new experimental, theoretic single-file designs with combinations of triple U (TU), triangle (TR), and convex triangle (CT) cross sections and to compare their bending stresses in simulated root canals with a single cross-sectional instrument using finite element analysis. Methods: A 3-dimensional model of the simulated root canal with 45 curvature and NiTi files with 5 cross-sectional designs were created using Pro/ENGINEER Wildfire 4.0 software (PTC Inc, Needham, MA) and ANSYS software (version 17; ANSYS, Inc, Canonsburg, PA) for finite element analysis. The NiTi files of 3 groups had single cross-sectional shapes of CT, TR, and TU designs, and 2 experimental groups had a CT, TR, and TU (CTU) design and a TU, TR, and CT (UTC) design. The file was rotated in simulated root canals to analyze the bending stress, and the von Mises stress value for every file was recorded in MPa. Statistical analysis was performed using the Kruskal-Wallis test and the Bonferroni-adjusted Mann-Whitney test for multiple pair-wise comparison with a P value <.05 (95 A). Results: The maximum bending stress of the rotary file was observed in the apical third of the CT design, whereas comparatively less stress was recorded in the CTU design. The TU and TR designs showed a similar stress pattern at the curvature, whereas the UTC design showed greater stress in the apical and middle thirds of the file in curved canals. All the file designs showed a statistically significant difference. Conclusions: The CTU designed instruments showed the least bending stress on a 45 degrees angulated simulated root canal when compared with all the other tested designs.
引用
收藏
页码:654 / 658
页数:5
相关论文
共 50 条
  • [1] Effect of Cross-sectional Designs on Torsional Resistance of Endodontic Nickel-Titanium Files: A Finite Element Study
    Nanthaprathip, Natchaphon
    Morakul, Sarita
    Hiran-us, Sirawut
    Singhatanadgid, Pairod
    EUROPEAN JOURNAL OF DENTISTRY, 2024,
  • [2] Correlation between Experimental Cyclic Fatigue Resistance and Numerical Stress Analysis for Nickel-Titanium Rotary Files
    Lee, Min-Ho
    Versluis, Antheunis
    Kim, Byung-Min
    Lee, Chan-Joo
    Hur, Bock
    Kim, Hyeon-Cheol
    JOURNAL OF ENDODONTICS, 2011, 37 (08) : 1152 - 1157
  • [3] Prediction of Cyclic Fatigue Life of Nickel-Titanium Rotary Files by Virtual Modeling and Finite Elements Analysis
    Scattina, Alessandro
    Alovisi, Mario
    Paolino, Davide Salvatore
    Pasqualini, Damiano
    Scotti, Nicola
    Chiandussi, Giorgio
    Berutti, Elio
    JOURNAL OF ENDODONTICS, 2015, 41 (11) : 1867 - 1870
  • [4] Flexural Stiffness and Stresses in Nickel-Titanium Rotary Files for Various Pitch and Cross-sectional Geometries
    Versluis, Antheunis
    Kim, Hyeon-Cheol
    Lee, WooCheol
    Kim, Byung-Min
    Lee, Chan-Joo
    JOURNAL OF ENDODONTICS, 2012, 38 (10) : 1399 - 1403
  • [5] A Paradigm Shift for Torsional Stiffness of Nickel-Titanium Rotary Instruments: A Finite Element Analysis
    Zanza, Alessio
    Seracchiani, Marco
    Di Nardo, Dario
    Reda, Rodolfo
    Gambarini, Gianluca
    Testarelli, Luca
    JOURNAL OF ENDODONTICS, 2021, 47 (07) : 1149 - 1156
  • [6] Comparison of the Mechanical Behavior between Controlled Memory and Superelastic Nickel-Titanium Files via Finite Element Analysis
    Santos, Leandro de Arruda
    de Azevedo Bahia, Maria Guiomar
    de Las Casas, Estevam Barbosa
    Lopes Buono, Vicente Tadeu
    JOURNAL OF ENDODONTICS, 2013, 39 (11) : 1444 - 1447
  • [7] Effect of several nickel-titanium rotary systems on stress distribution in mandibular molars under occlusal forces: a finite element analysis
    Sonmez, Sinem
    Koc, Cemre
    ODONTOLOGY, 2024, : 666 - 674
  • [8] Finite element analysis of rotary nickel-titanium endodontic instruments: A critical review of the methodology
    Chien, Philip Y-H
    Walsh, Laurence J.
    Peters, Ove A.
    EUROPEAN JOURNAL OF ORAL SCIENCES, 2021, 129 (05)
  • [9] An In Vitro Comparison of Torsional Stress Properties of Three Different Rotary Nickel-Titanium Files with a Similar Cross-Sectional Design
    Wycoff, Reid C.
    Berzins, David W.
    JOURNAL OF ENDODONTICS, 2012, 38 (08) : 1118 - 1120
  • [10] Morphometric analysis of shank-to-flute ratio in rotary nickel-titanium files
    Biz, MT
    Figueiredo, JAP
    INTERNATIONAL ENDODONTIC JOURNAL, 2004, 37 (06) : 353 - 358