Vacancy ion-exclusion chromatography of carboxylic acids on a weakly acidic cation-exchange resin

被引:28
作者
Tanaka, K [1 ]
Ding, MY
Helaleh, MIH
Taoda, H
Takahashi, H
Hu, WZ
Hasebe, K
Haddad, PR
Fritz, JS
Sarzanini, C
机构
[1] Natl Inst Adv Ind Sci & Technol, Seto 4890884, Japan
[2] Chubu Univ, Grad Sch Engn, Seto 4890884, Japan
[3] Tsing Hua Univ, Beijing 100084, Peoples R China
[4] Hokkaido Univ, Sapporo, Hokkaido 0600810, Japan
[5] Univ Tasmania, Sch Chem, Australian Ctr Res Separat Sci, Hobart, Tas 7001, Australia
[6] Iowa State Univ, Ames, IA 50011 USA
[7] Univ Turin, I-10125 Turin, Italy
关键词
vacancy chromatography; ion-exclusion chromatography; carboxylic acids;
D O I
10.1016/S0021-9673(02)00393-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
in this preliminary study, a new approach to ion-exclusion chromatography is proposed to overcome the relatively poor conductivity detection response which occurs in ion-exclusion chromatography when acids are added to the eluent in order to improve peak shape. This approach, termed vacancy ion-exclusion chromatography, requires the sample to be used as eluent and a sample of water to be injected onto a weakly acidic cation-exchange column (TSKgel OApak-A). Vacancy peaks for each of the analytes appear at the retention times of these analytes, Highly sensitive conductivity detection is possible and sharp, well-shaped peaks are produced, leading to efficient separations. Retention times were found to be affected by the concentration of the analytes in the eluent, and also by the presence of an organic modifier such as methanol in the eluent. Detection limits for oxalic, formic, acetic, propionic, butyric and valeric acids were 0.1, 0.2, 0.3, 0.3, 0.4 and 0.5 muM, respectively, and linear ranges for some acids extended over two orders of magnitude. Precision values for retention times were 0.21% and for peak areas were <1.90%. The vacancy ion-exclusion chromatography method was found to give detection responses four to 10 times higher than conventional ion-exclusion chromatography using sulfuric acid eluent and two to five times higher than conventional ion-exclusion chromatography using benzoic acid eluent. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:209 / 214
页数:6
相关论文
共 17 条
[1]   Separation of perfluoroalkylsulfinic acids and perfluoroalkylsulfonic acids by ion-exclusion chromatography [J].
Abe, T ;
Baba, H ;
Itoh, E ;
Tanaka, K .
JOURNAL OF CHROMATOGRAPHY A, 2001, 920 (1-2) :173-180
[2]   Novel way of separating polyfluorocarboxylic acids by ion-exclusion chromatography [J].
Abe, T ;
Baba, H ;
Soloshonok, I ;
Tanaka, K .
JOURNAL OF CHROMATOGRAPHY A, 2000, 884 (1-2) :93-103
[3]   Simultaneous ion-exclusion chromatography and cation-exchange chromatography of anions and cations in environmental water samples on a weakly acidic cation-exchange resin by elution with pyridine-2,6-dicarboxylic acid [J].
Ding, MY ;
Tanaka, K ;
Hu, WZ ;
Hasebe, K ;
Haddad, PR .
ANALYST, 2001, 126 (05) :567-570
[4]   PRINCIPLES AND APPLICATIONS OF ION-EXCLUSION CHROMATOGRAPHY [J].
FRITZ, JS .
JOURNAL OF CHROMATOGRAPHY, 1991, 546 (1-2) :111-118
[5]   Simultaneous determination of anions and cations by ion-exclusion chromatography-cation-exchange chromatography with tartaric acid/18-crown-6 as eluent [J].
Kwon, SM ;
Lee, KP ;
Tanaka, K ;
Ohta, K .
JOURNAL OF CHROMATOGRAPHY A, 1999, 850 (1-2) :79-84
[6]   SIMULTANEOUS ION-EXCLUSION CHROMATOGRAPHY CATION-EXCHANGE CHROMATOGRAPHY WITH CONDUCTIMETRIC DETECTION OF ANIONS AND CATIONS IN ACID-RAIN WATERS [J].
TANAKA, K ;
OHTA, K ;
FRITZ, JS ;
MATSUSHITA, S ;
MIYANAGA, A .
JOURNAL OF CHROMATOGRAPHY A, 1994, 671 (1-2) :239-248
[7]   Acid-rain monitoring in East Asia with a portable-type ion-exclusion-cation-exchange chromatographic analyzer [J].
Tanaka, K ;
Ohta, K ;
Haddad, PR ;
Fritz, JS ;
Lee, KP ;
Hasebe, K ;
Ieuji, A ;
Miyanaga, A .
JOURNAL OF CHROMATOGRAPHY A, 1999, 850 (1-2) :311-317
[8]   Ion-exclusion chromatography of aliphatic carboxylic acids on a cation-exchange resin by elution with polyvinyl alcohol [J].
Tanaka, K ;
Ohta, K ;
Fritz, JS .
JOURNAL OF CHROMATOGRAPHY A, 1997, 770 (1-2) :211-218
[9]   ION-EXCLUSION CHROMATOGRAPHY WITH CONDUCTIMETRIC DETECTION OF ALIPHATIC CARBOXYLIC-ACIDS ON AN H+-FORM CATION-EXCHANGE RESIN COLUMN BY ELUTION WITH POLYOLS AND SUGARS [J].
TANAKA, K ;
OHTA, K ;
FRITZ, JS ;
LEE, YS ;
SHIM, SB .
JOURNAL OF CHROMATOGRAPHY A, 1995, 706 (1-2) :385-393
[10]   High-performance ion-exclusion/cation-exchange chromatography of anions and cations in acid rain waters on a weakly acidic cation-exchange resin [J].
Tanaka, K ;
Ohta, K ;
Haddad, PR ;
Fritz, JS ;
Miyanaga, A ;
Hu, WZ ;
Hasebe, K ;
Lee, KP ;
Sarzanini, C .
JOURNAL OF CHROMATOGRAPHY A, 2001, 920 (1-2) :239-245