Comparison of sensitivity between gas chromatography-low-resolution mass spectrometry and gas chromatography-high-resolution mass spectrometry for determining metandienone metabolites in urine

被引:15
作者
Kokkonen, J
Leinonen, A
Tuominen, J
Seppälä, T
机构
[1] Univ Helsinki, Dept Chem, Analyt Chem Lab, FIN-00014 Helsinki, Finland
[2] VTT Chem Technol, VTT 02044, Finland
[3] United Labs, Doping Control Lab, FIN-00381 Helsinki, Finland
[4] Natl Publ Hlth Inst, FIN-00300 Helsinki, Finland
来源
JOURNAL OF CHROMATOGRAPHY B | 1999年 / 734卷 / 02期
关键词
metandienone; anabolic steroids;
D O I
10.1016/S0378-4347(99)00317-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In doping control laboratories the misuse of anabolic androgenic steroids is commonly investigated in urine by gas chromatography-low-resolution mass spectrometry with selected ion monitoring (GC-LRMS-SIM). By using high-resolution mass spectrometry (HRMS) detection sensitivity is improved due to reduction of biological background. In our study HRMS and LRMS methods were compared to each other. Two different sets were measured both with HRMS and LRMS. In the first set metandienone (I) metabolites 17 alpha-methyl-5 beta-androstan-3 alpha,17 beta-diol (II), 17-epimetandienone (III), 17 beta-methyl-5 beta-androst-1-ene-3 alpha,17 alpha-diol (IV) and 6 beta-hydroxymetandienone (V) were spiked in urine extract prepared by solid-phase extraction, hydrolysis with beta-glucuronidase from Escherichia coli and liquid-liquid extraction. In the second set the metabolites were first spiked in blank urine samples of four male persons before pretreatment. Concentration range of the spiked metabolites was 0.1-10 ng/ml in both sets. With HRMS (resolution of 5000) detection limits were 2-10 times lower than with LRMS. However, also with the HRMS method the biological background hampered detection and compounds from matrix were coeluted with some metabolites. For this reason the S/N values of the metabolites spiked had to be first compared to S/N values of coeluted matrix compounds to get any idea of detection limits. At trace concentrations selective isolation procedures should be implemented in order to confirm a positive result. The results suggest that metandienone misuse can be detected by HRMS for a prolonged period after stopping the intake of metandienone. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:179 / 189
页数:11
相关论文
共 4 条
[1]   PREPARATION OF TRIMETHYLSILYL, TRIETHYLSILYL AND TERT-BUTYLDIMETHYLSILYL ENOL ETHERS FROM KETOSTEROIDS FOR INVESTIGATIONS BY GAS-CHROMATOGRAPHY AND MASS-SPECTROMETRY [J].
DONIKE, M ;
ZIMMERMANN, J .
JOURNAL OF CHROMATOGRAPHY, 1980, 202 (03) :483-486
[2]  
Donike M., 1984, DTSCH Z SPORTMEDIZIN, V1, P14
[3]   METABOLISM OF ANABOLIC-STEROIDS IN MAN - SYNTHESIS AND USE OF REFERENCE SUBSTANCES FOR IDENTIFICATION OF ANABOLIC-STEROID METABOLITES [J].
SCHANZER, W ;
DONIKE, M .
ANALYTICA CHIMICA ACTA, 1993, 275 (1-2) :23-48
[4]   Long-term detection and identification of metandienone and stanozolol abuse in athletes by gas chromatography high-resolution mass spectrometry [J].
Schanzer, W ;
Delahaut, P ;
Geyer, H ;
MAchnik, M ;
Horning, S .
JOURNAL OF CHROMATOGRAPHY B, 1996, 687 (01) :93-108