Presidential guest lecture - Tribology of alternative beatings

被引:112
作者
Fisher, John [1 ]
Jin, Zhongmin [1 ]
Tipper, Joanne [1 ]
Stone, Martin [1 ]
Ingham, Eileen [1 ]
机构
[1] Univ Leeds, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
关键词
D O I
10.1097/01.blo.0000238871.07604.49
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The tribological performance and biological activity of the wear debris produced has been compared for highly crosslinked polyethylene, ceramic-on-ceramic, metal-on-metal, and modified metal bearings in a series of in vitro studies from a single laboratory. The functional lifetime demand of young and active patients is 10-fold greater than the estimated functional lifetime of traditional polyethylene. There is considerable interest in using larger diameter heads in these high demand patients. Highly cross-linked polyethylene show a four-fold reduction in functional biological activity. Ceramic-on-ceramic bearings have the lowest wear rates and least reactive wear debris. The functional biological activity is 20-fold lower than with highly cross-linked polyethylene. Hence, ceramic-on-ceramic bearings address the tribological lifetime demand of highly active patients. Metal-on-metal bearings have substantially lower wear rates than highly cross-linked polyethylene and wear decreases with head diameter. Bedding in wear is also lower with reduced radial clearance. Differential hardness ceramic-on-metal bearings and the application of ceramic-like coatings reduce metal wear and ion levels.
引用
收藏
页码:25 / 34
页数:10
相关论文
共 62 条
[1]   A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates [J].
Barbour, PSM ;
Stone, MH ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H6) :455-467
[2]   A hip joint simulator study using new and physiologically scratched femoral heads with ultra-high molecular weight polyethylene acetabular cups [J].
Barbour, PSM ;
Stone, MH ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H6) :569-576
[3]   Wear and surface cracking in early retrieved highly cross-linked polyethylene acetabular liners [J].
Bradford, L ;
Baker, DA ;
Graham, J ;
Chawan, A ;
Ries, MD ;
Pruitt, LA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (06) :1271-1282
[4]  
Chan FW, 1996, CLIN ORTHOP RELAT R, P96
[5]   Biomechanics of large femoral heads - What they do and don't do [J].
Crowninshield, RD ;
Maloney, WJ ;
Wentz, DH ;
Humphrey, SM ;
Blanchard, CR .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (429) :102-107
[6]   Polyethylene cross-linking by two different methods reduces acetabular liner wear in a hip joint wear simulator [J].
D'Lima, DD ;
Hermida, JC ;
Chen, PC ;
Colwell, CW .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2003, 21 (05) :761-766
[7]   Highly cross-linked polyethylene in cemented THA -: Randomized study of 61 hips [J].
Digas, G ;
Kärrholm, J ;
Thanner, J ;
Malchau, H ;
Herberts, P .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (417) :126-138
[8]  
Doorn PF, 1998, J BIOMED MATER RES, V42, P103, DOI 10.1002/(SICI)1097-4636(199810)42:1<103::AID-JBM13>3.3.CO
[9]  
2-D
[10]   A hip joint simulator study of the performance of metal-on-metal joints - Part II: Design [J].
Dowson, D ;
Hardaker, C ;
Flett, M ;
Isaac, GH .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :124-130