Conditions to minimize soft single biomolecule imaging with atomic force microscopy

被引:12
作者
Godon, Christian [1 ]
Teulon, Jean-Marie [2 ,3 ,4 ]
Odorico, Michael [5 ]
Basset, Christian [6 ]
Meillan, Matthieu [7 ,8 ]
Vellutini, Luc [7 ,8 ]
Chen, Shu-wen W. [9 ]
Pellequer, Jean-Luc [2 ,3 ,4 ]
机构
[1] CEA, BIAM, F-13108 St Paul Les Durance, France
[2] Univ Grenoble Alpes, IBS, F-38044 Grenoble, France
[3] CNRS, IBS, F-38044 Grenoble, France
[4] CEA, IBS, F-38044 Grenoble, France
[5] CEA CNRS UM2 ENSCM, ICSM UMR5257, F-30207 Bagnols Sur Ceze, France
[6] CEA, BIG, F-38054 Grenoble, France
[7] Univ Bordeaux, ISM, UMR 5255, F-33400 Talence, France
[8] CNRS, ISM, UMR 5255, F-33400 Talence, France
[9] 478 Rue Cyprien Jullin, F-38470 Vinay, France
关键词
AFM topography; Single molecule deformation; Mica; Self-assembled monolayers; Peakforce tapping; Tobacco mosaic virus; TOBACCO-MOSAIC-VIRUS; RAY FIBER DIFFRACTION; HIGH-RESOLUTION; AFM TOPOGRAPHS; SURFACE; RECONSTRUCTION; PROTEINS; MOLECULE; MICA; MACROMOLECULES;
D O I
10.1016/j.jsb.2016.12.011
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
A recurrent interrogation when imaging soft biomolecules using atomic force microscopy (AFM) is the putative deformation of molecules leading to a bias in recording true topographical surfaces. Deformation of biomolecules comes from three sources: sample instability, adsorption to the imaging substrate, and crushing under tip pressure. To disentangle these causes, we measured the maximum height of a well-known biomolecule, the tobacco mosaic virus (TMV), under eight different experimental conditions positing that the maximum height value is a specific indicator of sample deformations. Six basic AFM experimental factors were tested: imaging in air (AIR) versus in liquid (LIQ), imaging with flat minerals (MICA) versus flat organic surfaces (self-assembled monolayers, SAM), and imaging forces with oscillating tapping mode (TAP) versus PeakForce tapping (PFT). The results show that the most critical parameter in accurately measuring the height of TMV in air is the substrate. In a liquid environment, regardless of the substrate, the most critical parameter is the imaging mode. Most importantly, the expected TMV height values were obtained with both imaging with the PeakForce tapping mode either in liquid or in air at the condition of using self-assembled monolayers as substrate. This study unambiguously explains previous poor results of imaging biomolecules on mica in air and suggests alternative methodologies for depositing soft biomolecules on well organized self-assembled monolayers. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:322 / 329
页数:8
相关论文
共 60 条
[1]   Pretransition and progressive softening of bovine carbonic anhydrase II as probed by single molecule atomic force microscopy [J].
Afrin, R ;
Alam, MT ;
Ikai, A .
PROTEIN SCIENCE, 2005, 14 (06) :1447-1457
[2]   ATTRACTIVE-MODE IMAGING OF BIOLOGICAL-MATERIALS WITH DYNAMIC FORCE MICROSCOPY [J].
ANSELMETTI, D ;
LUTHI, R ;
MEYER, E ;
RICHMOND, T ;
DREIER, M ;
FROMMER, JE ;
GUNTHERODT, HJ .
NANOTECHNOLOGY, 1994, 5 (02) :87-94
[3]  
AZIMZADEH A, 1992, Journal of Molecular Recognition, V5, P9, DOI 10.1002/jmr.300050103
[4]   TM-AFM threshold analysis of macromolecular orientation: A study of the orientation of IgG and IgE on mica surfaces [J].
Bergkvist, M ;
Carlsson, J ;
Karlsson, T ;
Oscarsson, S .
JOURNAL OF COLLOID AND INTERFACE SCIENCE, 1998, 206 (02) :475-481
[5]   ATOMIC FORCE MICROSCOPE [J].
BINNIG, G ;
QUATE, CF ;
GERBER, C .
PHYSICAL REVIEW LETTERS, 1986, 56 (09) :930-933
[6]   SURFACE STUDIES BY SCANNING TUNNELING MICROSCOPY [J].
BINNING, G ;
ROHRER, H ;
GERBER, C ;
WEIBEL, E .
PHYSICAL REVIEW LETTERS, 1982, 49 (01) :57-61
[7]   Factor Va alternative conformation reconstruction using atomic force microscopy [J].
Chaves, Rui C. ;
Dahmane, Selma ;
Odorico, Michael ;
Nicolaes, Gerry A. F. ;
Pellequer, Jean-Luc .
THROMBOSIS AND HAEMOSTASIS, 2014, 112 (06) :1167-1173
[8]   DockAFM: benchmarking protein structures by docking under AFM topographs [J].
Chaves, Rui C. ;
Pellequer, Jean-Luc .
BIOINFORMATICS, 2013, 29 (24) :3230-3231
[9]   Atomic force microscope, molecular imaging, and analysis [J].
Chen, Shu-wen W. ;
Teulon, Jean-Marie ;
Godon, Christian ;
Pellequer, Jean-Luc .
JOURNAL OF MOLECULAR RECOGNITION, 2016, 29 (01) :51-55
[10]   Nanoscale structural features determined by AFM for single virus particles [J].
Chen, Shu-Wen W. ;
Odorico, Michael ;
Meillan, Matthieu ;
Vellutini, Luc ;
Teulon, Jean-Marie ;
Parot, Pierre ;
Bennetau, Bernard ;
Pellequer, Jean-Luc .
NANOSCALE, 2013, 5 (22) :10877-10886