Radiotracer Dose Reduction in Integrated PET/MR: Implications from National Electrical Manufacturers Association Phantom Studies

被引:43
作者
Oehmigen, Mark [1 ,2 ,3 ]
Ziegler, Susanne [1 ]
Jakoby, Bjoern W. [4 ,5 ]
Georgi, Jens-Christoph [4 ]
Paulus, Daniel H. [1 ]
Quick, Harald H. [1 ,2 ,3 ]
机构
[1] Univ Erlangen Nurnberg, Inst Med Phys, D-91054 Erlangen, Germany
[2] Univ Duisburg Essen, Erwin L Hahn Inst Magnet Resonance Imaging, D-45141 Essen, Germany
[3] Univ Hosp Essen, Essen, Germany
[4] Siemens Healthcare Sect, Erlangen, Germany
[5] Univ Surrey, Guildford GU2 5XH, Surrey, England
关键词
radiotracer dose reduction; NEMA image quality measurements; integrated PET/MR hybrid imaging; PET/MR phantom measurements; WHOLE-BODY PET/MRI; ATTENUATION CORRECTION; CLINICAL-EXPERIENCE; SCANNER; IMAGE;
D O I
10.2967/jnumed.114.139147
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
With the replacement of ionizing CT by MR imaging, integrated PET/MR in selected clinical applications may reduce the overall patient radiation dose when compared with PET/CT. Further potential for radiotracer dose reduction, while maintaining PET image quality (IQ) in integrated PET/MR, may be achieved by increasing the PET acquisition duration to match the longer time needed for MR data acquisition. To systematically verify this hypothesis under controlled conditions, this dose-reduction study was performed using a standardized phantom following the National Electrical Manufacturers Association (NEMA) IQ protocol. Methods: All measurements were performed on an integrated PET/MR whole-body hybrid system. The NEMA IQ phantom was filled with water and a total activity of 50.35 MBq of F-18-FDG. The sphere-to-background activity ratio was 8:1. Multiple PET data blocks of 20-min acquisition time were acquired in list-mode format and were started periodically at multiples of the F-18-FDG half-lives. Different sinograms (2, 4, 8, and 16 min in duration) were reconstructed. Attenuation correction of the filled NEMA phantom was performed using a CT-based attenuation map template. The attenuation-corrected PET images were then quantitatively evaluated following the NEMA IQ protocol, investigating contrast recovery, background variability, and signal-to-noise ratio. Image groups with half the activity and twice the acquisition time were evaluated. For better statistics, the experiment was repeated 3 times. Results: Contrast recovery, background variability, and signal-to-noise ratio remained almost constant over 3 half-life periods when the decreasing radiotracer activity (100%, 50%, 25%, and 12.5%) was compensated by increasing acquisition time (2, 4, 8, and 16 min). The variation of contrast recovery over 3 half-life periods was small (-6% to +7%), with a mean variation of 2%, compared with the reference setting (100%, 2 min). The signal-to-noise ratio of the hot spheres showed only minor variations over 3 half-life periods (5%). Image readers could not distinguish subjective IQ between the different PET acquisition setups. Conclusion: An approach to reduce the injected radiotracer activity in integrated PET/MR imaging, while maintaining PET 10, was presented and verified under idealized experimental conditions. This experiment may serve as a basis for further clinical PET/MR studies using reduced radiotracer dose as compared with conventional PET/CT studies.
引用
收藏
页码:1361 / 1367
页数:7
相关论文
共 14 条
[1]   FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0 [J].
Boellaard, Ronald ;
O'Doherty, Mike J. ;
Weber, Wolfgang A. ;
Mottaghy, Felix M. ;
Lonsdale, Markus N. ;
Stroobants, Sigrid G. ;
Oyen, Wim J. G. ;
Kotzerke, Joerg ;
Hoekstra, Otto S. ;
Pruim, Jan ;
Marsden, Paul K. ;
Tatsch, Klaus ;
Hoekstra, Corneline J. ;
Visser, Eric P. ;
Arends, Bertjan ;
Verzijlbergen, Fred J. ;
Zijlstra, Josee M. ;
Comans, Emile F. I. ;
Lammertsma, Adriaan A. ;
Paans, Anne M. ;
Willemsen, Antoon T. ;
Beyer, Thomas ;
Bockisch, Andreas ;
Schaefer-Prokop, Cornelia ;
Delbeke, Dominique ;
Baum, Richard P. ;
Chiti, Arturo ;
Krause, Bernd J. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 (01) :181-200
[2]   Evaluation of the attenuation properties of MR equipment for its use in a whole-body PET/MR scanner [J].
Delso, G. ;
Martinez-Moeller, A. ;
Bundschuh, R. A. ;
Ladebeck, R. ;
Candidus, Y. ;
Faul, D. ;
Ziegler, S. I. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (15) :4361-4374
[3]  
Delso G, 2011, J NUCL MED, V52, P1914, DOI 10.2967/jnumed.111.092726
[4]   First Clinical Experience with Integrated Whole-Body PET/MR: Comparison to PET/CT in Patients with Oncologic Diagnoses [J].
Drzezga, Alexander ;
Souvatzoglou, Michael ;
Eiber, Matthias ;
Beer, Ambros J. ;
Fuerst, Sebastian ;
Martinez-Moeller, Axel ;
Nekolla, Stephan G. ;
Ziegler, Sibylle ;
Ganter, Carl ;
Rummeny, Ernst J. ;
Schwaiger, Markus .
JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (06) :845-855
[5]   Optimizing Positron Emission Tomography Image Acquisition Protocols in Integrated Positron Emission Tomography/Magnetic Resonance Imaging [J].
Hartung-Knemeyer, Verena ;
Beiderwellen, Karsten J. ;
Buchbender, Christian ;
Kuehl, Hilmar ;
Lauenstein, Thomas C. ;
Bockisch, Andreas ;
Poeppel, Thorsten D. .
INVESTIGATIVE RADIOLOGY, 2013, 48 (05) :290-294
[6]   Workflow and Scan Protocol Considerations for Integrated Whole-Body PET/MRI in Oncology [J].
Martinez-Moeller, Axel ;
Eiber, Matthias ;
Nekolla, Stephan G. ;
Souvatzoglou, Michael ;
Drzezga, Alexander ;
Ziegler, Sibylle ;
Rummeny, Ernst J. ;
Schwaiger, Markus ;
Beer, Ambros J. .
JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (09) :1415-1426
[7]   Tissue Classification as a Potential Approach for Attenuation Correction in Whole-Body PET/MRI: Evaluation with PET/CT Data [J].
Martinez-Moeller, Axel ;
Souvatzoglou, Michael ;
Delso, Gaspar ;
Bundschuh, Ralph A. ;
Chefd'hotel, Christophe ;
Ziegler, Sibylle I. ;
Navab, Nassir ;
Schwaiger, Markus ;
Nekolla, Stephan G. .
JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (04) :520-526
[8]  
National Electrical Manufacturers Association, 2007, PERF MEAS POS EM TOM, P26
[9]   Towards improved hardware component attenuation correction in PET/MR hybrid imaging [J].
Paulus, D. H. ;
Tellmann, L. ;
Quick, H. H. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (22) :8021-8040
[10]  
Pichler BJ, 2006, J NUCL MED, V47, P639