Determination of acid phosphatase activity in a double injection flow system with electrochemical detection

被引:8
|
作者
Castanón-Fernández, J [1 ]
Fernández-Abedul, MT [1 ]
Costa-García, A [1 ]
机构
[1] Univ Oviedo, Dept Quim Fis & Analit, Asturias, Spain
关键词
acid phosphatase; double-injection; stopped-flow; FIA-ECD;
D O I
10.1016/S0003-2670(99)00772-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An enzyme assay for the measurement of total and prostatic acid phosphatase activity by flow injection analysis with electrochemical detection has been developed. The substrate (alpha-naphthyl phosphate) and enzyme are simultaneously injected to the system by means of a double injection valve. The alpha-naphtol produced after stopping the flow is amperometrically quantified. The calibration curve is linear from 0.01 to 1.3 u/l with a limit of detection of 0.006 u/l. Results were compared with that obtained spectrophotometrically. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:225 / 232
页数:8
相关论文