Fracture Resistance of HeatTreated Nickel-Titanium Rotary Files After Usage and Autoclave Sterilization: An In Vitro Study

被引:3
作者
El Abed, Rashid [1 ]
Alshehhi, Aisha [1 ]
Kang, Yoo Jung [3 ]
Al Raeesi, Dana [4 ]
Khamis, Amar H. [2 ]
Jamal, Mohamed [1 ]
Kim, Hyeon-Cheol [3 ]
机构
[1] Mohammed Bin Rashid Univ Med & Hlth Sci, Hamdan Bin Mohammed Coll Dent Med, Endodont Dept, Dubai, U Arab Emirates
[2] Mohammed Bin Rashid Univ Med & Hlth Sci, Hamdan Bin Mohammed Coll Dent Med, Biostat Dept, Dubai, U Arab Emirates
[3] Pusan Natl Univ, Sch Dent, Dent Res Inst, Dept Conservat Dent, Yangsan, South Korea
[4] Emirates Hlth Serv Estab, Endodont Dept, Dubai, U Arab Emirates
关键词
Autoclave sterilization; cyclic fatigue; dynamic fatigue tests; fracture resistance; heat treatment; nickel-titanium; CYCLIC FATIGUE RESISTANCE; MECHANICAL-PROPERTIES; TRANSFORMATION BEHAVIOR; TORSIONAL PROPERTIES; ROTATIONAL SPEED; SHAPE-MEMORY; INSTRUMENTS; IMMERSION; STRESS;
D O I
10.1016/j.joen.2022.08.004
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: This study aimed to evaluate the dynamic cyclic fatigue resistance of heattreated nickel-titanium (NiTi) files after usage and autoclaving. Methods: Two heat-treated NiTi file systems with tip sizes of #25 (ie, ProTaper Gold F2 [Dentsply Sirona, Charlotte, NC] and TruNatomy Prime [Dentsply Sirona]) were selected. Files of each system were divided into 3 experimental subgroups (n 5 10). For subgroup 1, new files were tested to establish the baseline number of cycles to failure (NCF). For subgroup 2, files underwent canal instrumentation in a simulated J-shaped canal block with 35 degrees of curvature and a 16.5-mm length. The simulated canals were pre-enlarged using stainless steel files #10K and #15K. After instrumentation, the files were tested. For subgroup 3, files underwent canal instrumentation as in subgroup 2 followed by autoclave sterilization and then were tested for cyclic failure. The cyclic fatigue test was performed in dynamic pecking motion under body temperature. The time to fracture was recorded and calculated into NCF. Fracture fragment lengths were measured using a digital microcaliper. Statistical analysis was conducted using the independent t test and multifactorial analysis of variance with 95% significance. Scanning electron microscopy was used to examine the topographic characteristics after cyclic fatigue tests. Results: There were no significant differences between file systems in subgroup 1 using new files (P <.05). Although PG reduced the NCF after use, TN showed a significantly increased NCF (P <.05). However, after sterilization, TN showed a reduced NCF compared with the used one (P <.05). Conclusions: Cyclic loading by usage and heat treatment during sterilization may result in different cyclic fatigue resistance according to the geometry and alloy characteristics. (J Endod 2022;48:1428-1433.)
引用
收藏
页码:1428 / 1433
页数:6
相关论文
共 32 条
  • [1] Mechanical Properties of Glide Path Preparation Instruments with Different Pitch Lengths
    Al Raeesi, Dana
    Kwak, Sang Won
    Ha, Jung-Hong
    Sulaiman, Sameh
    El Abed, Rashid
    Kim, Hyeon-Cheol
    [J]. JOURNAL OF ENDODONTICS, 2018, 44 (05) : 864 - 868
  • [2] Ashok L., 2019, CONS DENT ENDOD J, V4, P34, DOI 10.5005/jpjournals-10048-0047
  • [3] Effect of Core Mass and Alloy on Cyclic Fatigue Resistance of Different Nickel-Titanium Endodontic Instruments in Matching Artificial Canals
    Buerklein, Sebastian
    Zupanc, Lennart
    Donnermeyer, David
    Tegtmeyer, Karsten
    Schaefer, Edgar
    [J]. MATERIALS, 2021, 14 (19)
  • [4] Bulem ÜK, 2013, J APPL ORAL SCI, V21, P505
  • [5] A Retrospective Study of Endodontic Treatment Outcome between Nickel-Titanium Rotary and Stainless Steel Hand Filing Techniques
    Cheung, Gary S. P.
    Liu, Christopher S. Y.
    [J]. JOURNAL OF ENDODONTICS, 2009, 35 (07) : 938 - 943
  • [6] Comparison of fracture rate, deformation rate, and efficiency between rotary endodontic instruments driven at 150 rpm and 350 rpm
    Daugherty, DW
    Gound, TG
    Comer, TL
    [J]. JOURNAL OF ENDODONTICS, 2001, 27 (02) : 93 - 95
  • [7] Laboratory comparison of the mechanical properties of TRUShape with several nickel-titanium rotary instruments
    Elnaghy, A. M.
    Elsaka, S. E.
    [J]. INTERNATIONAL ENDODONTIC JOURNAL, 2017, 50 (08) : 805 - 812
  • [8] Mechanical properties of ProTaper Gold nickel-titanium rotary instruments
    Elnaghy, A. M.
    Elsaka, S. E.
    [J]. INTERNATIONAL ENDODONTIC JOURNAL, 2016, 49 (11) : 1073 - 1078
  • [9] On the transformation behaviour, mechanical properties and biocompatibility of two NiTi-based shape memory alloys: NiTi42 and NiTi42Cu7
    Es-Souni, M
    Es-Souni, M
    Brandies, HF
    [J]. BIOMATERIALS, 2001, 22 (15) : 2153 - 2161
  • [10] Mechanical Properties of Various Heat-treated Nickel-titanium Rotary Instruments
    Goo, Hye-Jin
    Kwak, Sang Won
    Ha, Jung-Hong
    Pedulla, Eugenio
    Kim, Hyeon-Cheol
    [J]. JOURNAL OF ENDODONTICS, 2017, 43 (11) : 1872 - 1877