Adaptive Kinematic Control of a Robotic Venipuncture Device Based on Stereo Vision, Ultrasound, and Force Guidance

被引:47
作者
Balter, Max L. [1 ]
Chen, Alvin I. [1 ]
Maguire, Timothy J. [2 ]
Yarmush, Martin L. [1 ,3 ]
机构
[1] Rutgers State Univ, Dept Biomed Engn, Piscataway, NJ 08854 USA
[2] VascuLogic LLC, Piscataway, NJ 08854 USA
[3] Massachusetts Gen Hosp, Boston, MA 02108 USA
基金
美国国家卫生研究院; 美国国家科学基金会;
关键词
Medical robots; motion control; robot vision systems; SYSTEMS; DESIGN;
D O I
10.1109/TIE.2016.2557306
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Robotic systems have slowly entered the realm of modern medicine; however, outside the operating room, medical robotics has yet to be translated to more routine interventions such as blood sampling or intravenous fluid delivery. In this paper, we present a medical robot that safely and rapidly cannulates peripheral blood vessels-a procedure commonly known as venipuncture. The device uses near-infrared and ultrasound imaging to scan and select suitable injection sites, and a 9-DOF robot to insert the needle into the center of the vessel based on image and force guidance. We first present the system design and visual-servoing scheme of the latest generation robot, and then evaluate the performance of the device through workspace simulations and free-space positioning tests. Finally, we perform a series of motion tracking experiments using stereo vision, ultrasound, and force sensing to guide the position and orientation of the needle tip. Positioning experiments indicate sub-millimeter accuracy and repeatability over the operating workspace of the system, while tracking studies demonstrate real-time needle servoing in response to moving targets. Finally, robotic phantom cannulations demonstrate the use of multiple system states to confirm that the needle has reached the center of the vessel.
引用
收藏
页码:1626 / 1635
页数:10
相关论文
共 30 条
  • [21] Perry TS, 2013, IEEE SPECTRUM, V50, P23, DOI 10.1109/MSPEC.2013.6565554
  • [22] Detection of needle puncture to blood vessel using puncture force measurement
    Saito, H
    Togowa, T
    [J]. MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2005, 43 (02) : 240 - 244
  • [23] Ultrasonographically Guided Peripheral Intravenous Cannulation in Emergency Department Patients With Difficult Intravenous Access: A Randomized Trial
    Stein, John
    George, Brian
    River, Gerin
    Hebig, Anke
    McDermott, Daniel
    [J]. ANNALS OF EMERGENCY MEDICINE, 2009, 54 (01) : 33 - 40
  • [24] Stoianovici D, 1998, LECT NOTES COMPUT SC, V1496, P404, DOI 10.1007/BFb0056225
  • [25] The VeinViewer Vascular Imaging System Worsens First-Attempt Cannulation Rate for Experienced Nurses in Infants and Children with Anticipated Difficult Intravenous Access
    Szmuk, Peter
    Steiner, Jeffrey
    Pop, Radu B.
    Farrow-Gillespie, Alan
    Mascha, Edward J.
    Sessler, Daniel I.
    [J]. ANESTHESIA AND ANALGESIA, 2013, 116 (05) : 1087 - 1092
  • [26] Implementation of Bilateral Control System Based on Acceleration Control Using FPGA for Multi-DOF Haptic Endoscopic Surgery Robot
    Tanaka, Hiroyuki
    Ohnishi, Kouhei
    Nishi, Hiroaki
    Kawai, Toshikazu
    Morikawa, Yasuhide
    Ozawa, Soji
    Furukawa, Toshiharu
    [J]. IEEE TRANSACTIONS ON INDUSTRIAL ELECTRONICS, 2009, 56 (03) : 618 - 627
  • [27] Medical robotics in computer-integrated surgery
    Taylor, RH
    Stoianovici, D
    [J]. IEEE TRANSACTIONS ON ROBOTICS AND AUTOMATION, 2003, 19 (05): : 765 - 781
  • [28] Walsh G., 2008, J. Assoc. Vasc. Access, V13, P198, DOI DOI 10.2309/JAVA.13-4-7
  • [29] Object tracking: A survey
    Yilmaz, Alper
    Javed, Omar
    Shah, Mubarak
    [J]. ACM COMPUTING SURVEYS, 2006, 38 (04)
  • [30] Mechatronic design of a new robot for force control in minimally invasive surgery
    Zemiti, Nabil
    Morel, Guillaume
    Ortmaier, Tobias
    Bonnet, Nicolas
    [J]. IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2007, 12 (02) : 143 - 153