Development of the transtibial prosthesis controlled pneumatically and electrically by microcomputer system

被引:0
作者
Shimada, Youichi [1 ]
Terayama, Yukio [1 ]
机构
[1] Kanazawa Inst Technol, Ishikari, Hokkaido 9218501, Japan
来源
Biomedical Sciences Instrumentation, Vol 42 | 2006年 / 42卷
关键词
transtibial prosthesis; microprocessor; Gait Analysis; Solenoid Valve; Air Cylinder; multiple sensors system;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This report represents the development of the prototype transtibial prosthesis to assist a smooth and comfortable walking for an unilateral amputee. This prosthesis is composed of two air cylinders, solenoid valves, portable and small air tank for compressed air storage, a multiple sensor system and a microprocessor. Two air cylinders are located around the rods to act as antagonistic and agonistic muscles. The system causes flexion and extension of the foot plate jointed at the ankle with compressed air, injected -or discharged via a solenoid or electromagnetic valves. The valves or solenoids are controlled with a microprocessor (Microchip Technology Inc., PIC 16F876), the microprocessor generates control signals to the interface circuits for valve opening and closing consistent with the foot position during the walking phase. The control patterns generated in the microprocessor are modified with feedback from the touch sensor, ankle joint angle sensor and the two dimensional acceleration sensor. The primary walking pattern for an individual amputee should be developed through the gait analysis with video.
引用
收藏
页码:452 / 457
页数:6
相关论文
共 4 条
[1]  
DEJNABADI H, IEEE T BIOMED ENG, V52, P1478
[2]  
GIANSANSANTI D, IEEE T BIOMED ENG, V52, P1271
[3]  
LORD M, 1995, BIOMEDICAL ENG HDB, P2055
[4]  
POPOVIC DJ, 1995, BIOMEDICAL ENG HDB, P2086