On the evolution and relative merits of hard X-ray phase-contrast imaging methods

被引:114
作者
Wilkins, S. W. [1 ]
Nesterets, Ya. I. [1 ]
Gureyev, T. E. [1 ]
Mayo, S. C. [1 ]
Pogany, A. [1 ]
Stevenson, A. W. [1 ]
机构
[1] CSIRO Mat Sci & Engn, Clayton, Vic 3169, Australia
来源
PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES | 2014年 / 372卷 / 2010期
关键词
phase contrast; imaging; X-ray imaging; SIGNAL-TO-NOISE; SYNCHROTRON-RADIATION; INTERFEROMETER; RESOLUTION; RETRIEVAL; IMAGES; TOMOGRAPHY; ABSORPTION; OBJECT; WAVE;
D O I
10.1098/rsta.2013.0021
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
This review provides a brief overview, albeit from a somewhat personal perspective, of the evolution and key features of various hard X-ray phase-contrast imaging (PCI) methods of current interest in connection with translation to a wide range of imaging applications. Although such methods have already found wide-ranging applications using synchrotron sources, application to dynamic studies in a laboratory/clinical context, for example for in vivo imaging, has been slow due to the current limitations in the brilliance of compact laboratory sources and the availability of suitable high-performance X-ray detectors. On the theoretical side, promising new PCI methods are evolving which can record both components of the phase gradient in a single exposure and which can accept a relatively large spectral bandpass. In order to help to identify the most promising paths forward, we make some suggestions as to how the various PCI methods might be compared for performance with a particular view to identifying those which are the most efficient, given the fact that source performance is currently a key limiting factor on the improved performance and applicability of PCI systems, especially in the context of dynamic sample studies. The rapid ongoing development of both suitable improved sources and detectors gives strong encouragement to the view that hard X-ray PCI methods are poised for improved performance and an even wider range of applications in the near future.
引用
收藏
页数:19
相关论文
共 76 条
[1]  
ANDO M, 1972, 6TH P INT C XRAY OPT, P63
[2]   Oriental magic mirrors and the Laplacian image [J].
Berry, MV .
EUROPEAN JOURNAL OF PHYSICS, 2006, 27 (01) :109-118
[3]   AN X-RAY INTERFEROMETER WITH LONG SEPARATED INTERFERING BEAM PATHS (E) [J].
BONSE, U ;
HART, M .
APPLIED PHYSICS LETTERS, 1965, 7 (04) :99-&
[4]   AN X-RAY INTERFEROMETER [J].
BONSE, U ;
HART, M .
APPLIED PHYSICS LETTERS, 1965, 6 (08) :155-&
[5]  
Bragg William., 1933, The Universe of Light
[6]   X-ray phase-contrast imaging: from pre-clinical applications towards clinics [J].
Bravin, Alberto ;
Coan, Paola ;
Suortti, Pekka .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (01) :R1-R35
[7]   Diffraction enhanced x-ray imaging [J].
Chapman, D ;
Thomlinson, W ;
Johnston, RE ;
Washburn, D ;
Pisano, E ;
Gmur, N ;
Zhong, Z ;
Menk, R ;
Arfelli, F ;
Sayers, D .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (11) :2015-2025
[8]  
Clauser J. F, 1998, United States patent, Patent No. [US 5,812,629, 5812629]
[9]   NEW THEORETICAL AND EXPERIMENTAL RESULTS IN FRESNEL OPTICS WITH APPLICATIONS TO MATTER-WAVE AND X-RAY INTERFEROMETRY [J].
CLAUSER, JF ;
REINSCH, MW .
APPLIED PHYSICS B-PHOTOPHYSICS AND LASER CHEMISTRY, 1992, 54 (05) :380-395
[10]   Holotomography: Quantitative phase tomography with micrometer resolution using hard synchrotron radiation x rays [J].
Cloetens, P ;
Ludwig, W ;
Baruchel, J ;
Van Dyck, D ;
Van Landuyt, J ;
Guigay, JP ;
Schlenker, M .
APPLIED PHYSICS LETTERS, 1999, 75 (19) :2912-2914