Femoral access-related complications during percutaneous transcatheter aortic valve implantation comparing single versus double Prostar XL device closure

被引:8
|
作者
Saleh, Nawsad [1 ]
De Palma, Rodney [1 ]
Settergren, Magnus [1 ]
Ruck, Andreas [1 ]
机构
[1] Karolinska Univ Hosp, Karolinska Inst, Dept Cardiol, Solna, Sweden
关键词
TAVI; vascular closure device; VARC; ENDOVASCULAR ANEURYSM REPAIR; END-POINT DEFINITIONS; SUTURE; SITES;
D O I
10.1002/ccd.25966
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
ObjectivesTo evaluate the efficacy and safety of a double Prostar XL suture-based closure technique compared to a conventional single Prostar XL technique in elective transcatheter aortic valve implantation (TAVI) via the common femoral artery. BackgroundTAVI is recommended as a treatment for symptomatic severe aortic stenosis for those who are at high or prohibitive risk of surgical valve replacement. Vascular complications remain the most frequent category of procedural complication. The most efficacious and safest percutaneous suture-based closure technique is unknown. MethodsProspective observational study of Prostar XL device closures used in 126 consecutive patients between 2012 and 2014. Single Prostar XL closure was used in 63 patients and double Prostar XL closure in a further 63 patients. Outcomes from the groups were compared. All patients were treated transfemorally through an 18Fr sheath. Technical success was defined as hemostasis not requiring interventional or surgical repair during hospital admission. Bleeding and vascular complications were defined using the second consensus of the valvular academic research consortium (VARC-2) criteria. ResultsThe cohort was aged 83+/-6 and 48% were female with a logistic Euroscore of 24+/-11.6. Technical success was 86% and 98% respectively (P=0.017) with systematic single and double Prostar XL closure. Composite VARC-2 vascular and bleeding complications occurred more frequently in the single Prostar XL group compared to the double Prostar XL group (10 [16%] v 3 [5%] P<0.04, and 17 [27%] v 6 [10%] P<0.004). ConclusionA systematic double Prostar XL closure technique for large caliber arterial access sites during TAVI is feasible, safe and associated with fewer technical failures, fewer vascular complications, and less bleeding compared with single Prostar XL. (c) 2015 Wiley Periodicals, Inc.
引用
收藏
页码:1255 / 1261
页数:7
相关论文
共 35 条
  • [31] Totally Percutaneous Transfemoral Transcatheter Aortic Valve Replacement Despite Failure to Deploy a Vascular Closure Device: A Single-Centre Case Series
    Eitan, Amnon
    Shiran, Avinoam
    Sliman, Hussein
    Zafrir, Barak
    Jaffe, Ronen
    HEART LUNG AND CIRCULATION, 2022, 31 (03) : 390 - 394
  • [32] SAT-TAVI (single antiplatelet therapy for TAVI) study: A pilot randomized study comparing double to single antiplatelet therapy for transcatheter aortic valve implantation
    Stabile, Eugenio
    Pucciarelli, Armando
    Cota, Linda
    Sorropago, Giovanni
    Tesorio, Tullio
    Salemme, Luigi
    Popusoi, Grigore
    Ambrosini, Vittorio
    Cioppa, Angelo
    Agrusta, Marco
    Catapano, Donato
    Moscariello, Cesare
    Trimarco, Bruno
    Esposito, Giovanni
    Rubino, Paolo
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2014, 174 (03) : 624 - 627
  • [33] Simple, Effective and Safe Vascular Access Site Closure with the Double-ProGlide Preclose Technique in 162 Patients Receiving Transfemoral Transcatheter Aortic Valve Implantation
    Griese, Daniel P.
    Reents, Wilko
    Diegeler, Anno
    Kerber, Sebastian
    Babin-Ebell, Joerg
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2013, 82 (05) : E734 - E741
  • [34] Percutaneous Versus Surgical Cutdown Access for Transfemoral Transcatheter Aortic Valve Implantation: A Systematic Review and Meta-Analysis Focusing on Propensity-Score Matched Studies
    Riaz, Sania
    Shiva, Pavan Kumara Kasam
    Krishnamurthi, Jaya Surya Manimekalai
    Shah, Roopshri Sunilkumar
    Cherukuri, Anjani Mahesh Kumar
    Bhatia, Pranav
    Arul, Subiksha
    Multani, Monika
    Singh, Adishwar
    Suyambu, Jenisha
    Asif, Kainat
    Al-Tawil, Mohammed
    HEART LUNG AND CIRCULATION, 2024, 33 (10) : 1393 - 1403
  • [35] Comparing anticoagulation therapy alone versus anticoagulation plus single antiplatelet drug therapy after transcatheter aortic valve implantation in patients with an indication for anticoagulation: a systematic review and meta-analysis
    Zhu, Yong
    Meng, Shuai
    Chen, Maolin
    Jia, Ruofei
    Nan, Jing
    Li, Hong
    Zhu, Huagang
    Li, Long
    Jin, Zening
    CARDIOVASCULAR DRUGS AND THERAPY, 2021, 35 (05) : 995 - 1002