RF shielding and eddy currents in NMR probes

被引:4
作者
Conradi, Mark S. [1 ]
Zens, Albert P. [2 ]
机构
[1] ABQMR Inc, 2301 Yale Blvd SE,Suite C2, Albuquerque, NM 87106 USA
[2] JEOL, 10 Dearborn Rd, Peabody, MA 01960 USA
关键词
Shielding; Skin depth; NMR probes; Eddy currents; rf heating; Infrared shielding; Gradients; CAVITY;
D O I
10.1016/j.jmr.2019.06.011
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Shielding of the NMR sample by a thin coating of metal between the sample and the rf coil arises in some geometries. The shielding may be used for rf heating of the sample tube or for a high infrared reflectivity coating in cryoprobe applications. An important application is for a shield that prevents noise from entering the rf probe circuit while allowing pulsed magnetic field gradients or field modulation to pass. We show by simple, approximate derivations that the criterion for shielding is not whether the coating exceeds the classical electromagnetic skin depth delta at the operating frequency (as is often stated), but whether the geometric mean between the thickness and an appropriate radius r exceeds delta. Thus, because r is typically much larger than delta, conducting layers substantially thinner than delta may still be good shields. Measurements are performed at high audio frequencies to confirm the calculations, using geometries relevant to rf saddle coils and to rf solenoids. Measurements of the slowing of the edges of a pulsed field gradient are also in accord with the calculations. (C) 2019 Elsevier Inc. All rights reserved.
引用
收藏
页码:180 / 184
页数:5
相关论文
共 15 条
  • [2] X-BAND CAVITY FOR LOW-TEMPERATURE ELECTRON-SPIN RESONANCE EXPERIMENTS
    CASINI, G
    GRAZIANI, P
    LINARI, R
    TRONCONI, A
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1976, 47 (03) : 318 - 320
  • [3] X BAND ESR CAVITY FOR STUDIES OF PARAMAGNETIC GASES
    EVENSON, KM
    BURCH, DS
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1966, 37 (02) : 236 - &
  • [4] Griffiths D.J., 2005, Introduction to Electrodynamics
  • [5] High pressure magnetic resonance imaging with metallic vessels
    Han, Hui
    Ouellette, Matthew
    MacMillan, Bryce
    Goora, Frederic
    MacGregor, Rodney
    Green, Derrick
    Balcom, Bruce J.
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2011, 213 (01) : 90 - 97
  • [6] PROPERTIES OF A THICK-WALLED CONDUCTING ENCLOSURE IN LOW-FREQUENCY MAGNETIC SHIELDING
    HEINONEN, P
    TUOMOLA, M
    LEKKALA, J
    MALMIVUO, J
    [J]. JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1980, 13 (05): : 569 - 570
  • [7] Jackson J.D., 2001, Classical Electrodynmaics, VThird
  • [8] HIGH-TEMPERATURE NMR USING INDUCTIVE HEATING
    MARESCH, GG
    KENDRICK, RD
    YANNONI, CS
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (01) : 77 - 80
  • [9] Payne A., SHIELDING MAGNETIC F
  • [10] Poole CP., 1967, ELECT SPIN RESONANCE