Bias and Variability in Morphology Features of Lung Lesions Across CT Imaging Conditions

被引:1
|
作者
Hoye, Jocelyn [1 ,2 ,3 ,4 ]
Solomon, Justin B. [1 ,2 ,3 ,4 ]
Sauer, Thomas [1 ,2 ,3 ,4 ]
Robins, Marthony [1 ,2 ,3 ,4 ]
Samei, Ehsan [1 ,2 ,3 ,4 ]
机构
[1] Duke Univ, Med Ctr, Carl E Ravin Adv Imaging Labs, Durham, NC 27705 USA
[2] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27705 USA
[3] Duke Univ, Med Ctr, Med Phys Grad Program, Durham, NC 27705 USA
[4] Duke Univ, Med Ctr, Durham, NC 27705 USA
来源
MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING | 2018年 / 10573卷
关键词
Morphology; Lung Lesions; Quantitative Imaging; Computed Tomography; Imaging Conditions; COMPUTED-TOMOGRAPHY;
D O I
10.1117/12.2293545
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
CT imaging method can influence radiomic features. The purpose of this study was to characterize the intra-protocol and inter-protocol variability and bias of quantitative morphology features of lung lesions across a range of CT imaging conditions. A total of 15 lung lesions were simulated (five in each of three spiculation classes: low, medium, and high). For each lesion, a series of simulated CT images representing different imaging conditions were synthesized by applying 3D blur and adding correlated noise based on the measured noise and resolution properties of five commercial multi-slice CT systems, representing three dose levels (CTDIvol of 1.90, 3.75, 7.50 mGy), three slice thicknesses (0.625, 1.25, 2.5 mm), and 33 clinical reconstruction kernels. Five repeated image volumes were synthesized for each lesion and imaging condition. A series of 21 shape-based morphology features were extracted from both "ground truth" (i.e., pre-blur without noise) and "image rendered" lesions (i.e., post-blur and with noise). For each morphology feature, the intra-protocol and inter-protocol variability was characterized by calculating the average coefficient of variation (COV) across repeats and imaging conditions, respectively (average was across all lesions). The bias was quantified by comparing the percent relative error in the morphology metric between the imaged lesions and ground truth lesions. The average intra-protocol COV metric ranged from 0.2% to 3%. The average inter-protocol COV ranged from 3% to 106% with most features being around 30%. Percent relative error was most biased at 73% for Ellipsoid Volume and least biased at -0.27% for Flatness. Results of the study indicate that different reconstructions can lead to significant bias and variability in the measurements of morphological features.
引用
收藏
页数:11
相关论文
共 35 条
  • [1] Accuracy and Variability of Texture-based Radiomics Features of Lung Lesions across CT Imaging Conditions
    Zheng, Yuese
    Solomon, Justin
    Choudhury, Kingshuk
    Marin, Daniele
    Samei, Ehsan
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [2] Systematic analysis of bias and variability of morphologic features for lung lesions in computed tomography
    Hoye, Jocelyn
    Solomon, Justin
    Sauer, Thomas J.
    Robins, Marthony
    Samei, Ehsan
    JOURNAL OF MEDICAL IMAGING, 2019, 6 (01)
  • [3] Accuracy and variability of radiomics in photon-counting CT: texture features and lung lesion morphology
    Rajagopal, Jayasai R.
    Hoye, Jocelyn
    Robins, Marthony
    Jones, Elizabeth C.
    Samei, Ehsan
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948
  • [4] Quantification of Minimum Detectable Difference in Radiomics Features Across Lesions and CT Imaging Conditions
    Hoye, Jocelyn
    Solomon, Justin B.
    Sauer, Thomas J.
    Samei, Ehsan
    ACADEMIC RADIOLOGY, 2021, 28 (11) : 1570 - 1581
  • [5] Quantifying Truth-based change in Radiomics Features between CT Imaging Conditions
    Hoye, Jocelyn
    Solomon, Justin B.
    Samei, Ehsan
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948
  • [6] CT imaging of cavitary lung lesions
    Mama, N.
    Dhifallah, M.
    Ben Aicha, S.
    Kadri, K.
    Arifa, N.
    Hasni, I.
    Tlili-Graiess, K.
    FEUILLETS DE RADIOLOGIE, 2014, 54 (02) : 69 - 83
  • [7] Small-cell lung carcinoma: CT imaging features
    Ridene, I.
    Ben Miled-M'rad, K.
    Zidi, A.
    Hantous-Zannad, S.
    Baccouche, I.
    JOURNAL DE RADIOLOGIE, 2011, 92 (01): : 3 - 7
  • [8] Interobserver Variability in the Assessment of CT Imaging Features of Traumatic Brain Injury
    Chun, Kimberly A.
    Manley, Geoffrey T.
    Stiver, Shirley I.
    Aiken, Ashley H.
    Phan, Nicholas
    Wang, Vincent
    Meeker, Michele
    Cheng, Su-Chun
    Gean, A. D.
    Wintermark, Max
    JOURNAL OF NEUROTRAUMA, 2010, 27 (02) : 325 - 330
  • [9] Analysis of the Influence of CT Imaging Conditions on the CT Image Features of Pulmonary Nodule Phantoms
    Li, Jing
    Zhong, Daidi
    Wang, Shuo
    Tao, Chunjing
    Wang, Hao
    IEEE ACCESS, 2025, 13 : 35101 - 35112
  • [10] Quantitative Imaging of Regional Aerosol Deposition, Lung Ventilation and Morphology by Synchrotron Radiation CT
    Porra, L.
    Degrugilliers, L.
    Broche, L.
    Albu, G.
    Strengell, S.
    Suhonen, H.
    Fodor, G. H.
    Petak, F.
    Suortti, P.
    Habre, W.
    Sovijarvi, A. R. A.
    Bayat, S.
    SCIENTIFIC REPORTS, 2018, 8