Feasibility of spinal stereotactic body radiotherapy in Elekta Unity® MR-Linac

被引:0
作者
Han, Eun Young [1 ]
Aima, Manik [1 ]
Hughes, Neil [2 ]
Briere, Tina M. [1 ]
Yeboa, Debra N. [2 ]
Castillo, Pam [2 ]
Wang, Jihong [1 ]
Yang, Jinzhong [1 ]
Vedam, Sastry [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Unit 94,1515 Holcomb Blvd, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Radiat Oncol, Houston, TX 77030 USA
来源
JOURNAL OF RADIOSURGERY AND SBRT | 2020年 / 7卷 / 02期
关键词
Spine SBRT; Unity MR-Linac; TrueBeam; MAGNETIC-FIELD; MONTE-CARLO; REFERENCE DOSIMETRY; ACCELERATOR; OPTIMIZATION; METASTASES; SCANNER; IMPACT;
D O I
暂无
中图分类号
R61 [外科手术学];
学科分类号
摘要
The Elekta Unity MR-Linac (MRL) is expected to benefit spine stereotactic body radiotherapy (SBRT) due to the improved soft tissue contrast available with onboard MR imaging. However, the irradiation geometry and beam configuration of the MRL deviates from the conventional linear accelerator (Linac). The purpose of the study was to investigate the feasibility of spine SBRT on the MRL. Treatment plans were generated for lumbar and thoracic spines. Target and spinal cord doses were measured with two cylindrical ion chambers inserted into an anthropomorphic spine phantom. Our study indicated that the Monaco treatment planning system (TPS) could generate clinical treatment plans for the MRL that were of comparable quality to the RayStation TPS with a conventional Linac. For both Linacs the planned dose within the gross tumor volume agreed with measurements within +/- 3%. For the spinal cord, while the measured doses from the TrueBeam were 1.8% higher for the lumbar spine plan and 6.9% higher for thoracic spine plan, the measured doses from MRL were 0.6% lower for the lumbar spine plan and 3.9% higher for the thoracic spine plan. In conclusion, the feasibility of spine SBRT in Elekta Unity MRL has been demonstrated, however, more effort is needed for such as optimizing the online plan adaptation method.
引用
收藏
页码:127 / 134
页数:8
相关论文
共 26 条
  • [1] An online replanning method using warm start optimization and aperture morphing for flattening-filter-free beams
    Ahunbay, Ergun E.
    Ates, O.
    Li, X. A.
    [J]. MEDICAL PHYSICS, 2016, 43 (08) : 4575 - 4584
  • [2] AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
    Almond, PR
    Biggs, PJ
    Coursey, BM
    Hanson, WF
    Huq, MS
    Nath, R
    Rogers, DWO
    [J]. MEDICAL PHYSICS, 1999, 26 (09) : 1847 - 1870
  • [3] Treating locally advanced lung cancer with a 1.5 T MR-Linac - Effects of the magnetic field and irradiation geometry on conventionally fractionated and isotoxic dose-escalated radiotherapy
    Bainbridge, Hannah E.
    Menten, Martin J.
    Fast, Martin F.
    Nill, Simeon
    Oelfke, Uwe
    McDonald, Fiona
    [J]. RADIOTHERAPY AND ONCOLOGY, 2017, 125 (02) : 280 - 285
  • [4] Outcomes for Spine Stereotactic Body Radiation Therapy and an Analysis of Predictors of Local Recurrence
    Bishop, Andrew J.
    Tao, Randa
    Rebueno, Neal C.
    Christensen, Eva N.
    Allen, Pamela K.
    Wang, Xin A.
    Amini, Behrang
    Tannir, Nizar M.
    Tatsui, Claudio E.
    Rhines, Laurence D.
    Li, Jing
    Chang, Eric L.
    Brown, Paul D.
    Ghia, Amol J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 92 (05): : 1016 - 1026
  • [5] Phase I/II study of stereotactic body radiotherapy for spinal metastasis and its pattern of failure
    Chang, Eric L.
    Shiu, Almon S.
    Mendel, Ehud
    Mathews, Leni A.
    Mahajan, Anita
    Allen, Pamela K.
    Weinberg, Jeffrey S.
    Brown, Barry W.
    Wang, Xin Shelly
    Woo, Shiao Y.
    Cleeland, Charles
    Maor, Moshe H.
    Rhines, Laurence D.
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2007, 7 (02) : 151 - 160
  • [6] Technical Note: Dose effects of 1.5 T transverse magnetic field on tissue interfaces in MRI-guided radiotherapy
    Chen, Xinfeng
    Prior, Phil
    Chen, Guang-Pei
    Schultz, Christopher J.
    Li, X. Allen
    [J]. MEDICAL PHYSICS, 2016, 43 (08) : 4797 - 4802
  • [7] Comparison of treatment plan quality among MRI-based IMRT with a linac, MRI-based IMRT with tri-Co-60 sources, and VMAT for spine SABR
    Choi, Chang Heon
    Kim, Jin Ho
    Kim, Jung-in
    Park, Jong Min
    [J]. PLOS ONE, 2019, 14 (07):
  • [8] International Spine Radiosurgery Consortium Consensus Guidelines for Target Volume Definition in Spinal Stereotactic Radiosurgery
    Cox, Brett W.
    Spratt, Daniel E.
    Lovelock, Michael
    Bilsky, Mark H.
    Lis, Eric
    Ryu, Samuel
    Sheehan, Jason
    Gerszten, Peter C.
    Chang, Eric
    Gibbs, Iris
    Soltys, Scott
    Sahgal, Arjun
    Deasy, Joe
    Flickinger, John
    Quader, Mubina
    Mindea, Stefan
    Yamada, Yoshiya
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 83 (05): : E597 - E605
  • [9] GHANEM AI, 2018, CUREUS, V10, DOI DOI 10.7759/CUREUS.2412.PUBMED
  • [10] GPUMCD: A new GPU-oriented Monte Carlo dose calculation platform
    Hissoiny, Sami
    Ozell, Benoit
    Bouchard, Hugo
    Despres, Philippe
    [J]. MEDICAL PHYSICS, 2011, 38 (02) : 754 - 764