Image distortion correction for MRI in low field permanent magnet systems with strong B0 inhomogeneity and gradient field nonlinearities

被引:27
|
作者
Koolstra, Kirsten [1 ]
O'Reilly, Thomas [2 ]
Bornert, Peter [2 ,3 ]
Webb, Andrew [2 ]
机构
[1] Leiden Univ, Div Image Proc, Radiol, Med Ctr, Albinusdreef 2, NL-2333 ZA Leiden, Netherlands
[2] Leiden Univ, Med Ctr, CJ Gorter Ctr High Field MRI, Radiol, Albinusdreef 2, NL-2333 ZA Leiden, Netherlands
[3] Philips Res, Rontgenstr 24-26, D-22335 Hamburg, Germany
基金
欧洲研究理事会;
关键词
Distortion correction; B-0; mapping; Conjugate phase reconstruction; Model-based reconstruction; Low field MRI;
D O I
10.1007/s10334-021-00907-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective To correct for image distortions produced by standard Fourier reconstruction techniques on low field permanent magnet MRI systems with strong B-0 inhomogeneity and gradient field nonlinearities. Materials and methods Conventional image distortion correction algorithms require accurate Delta B-0 maps which are not possible to acquire directly when the B-0 inhomogeneities also produce significant image distortions. Here we use a readout gradient time-shift in a TSE sequence to encode the B-0 field inhomogeneities in the k-space signals. Using a non-shifted and a shifted acquisition as input, Delta B-0 maps and images were reconstructed in an iterative manner. In each iteration, Delta B-0 maps were reconstructed from the phase difference using Tikhonov regularization, while images were reconstructed using either conjugate phase reconstruction (CPR) or model-based (MB) image reconstruction, taking the reconstructed field map into account. MB reconstructions were, furthermore, combined with compressed sensing (CS) to show the flexibility of this approach towards undersampling. These methods were compared to the standard fast Fourier transform (FFT) image reconstruction approach in simulations and measurements. Distortions due to gradient nonlinearities were corrected in CPR and MB using simulated gradient maps. Results Simulation results show that for moderate field inhomogeneities and gradient nonlinearities, Delta B-0 maps and images reconstructed using iterative CPR result in comparable quality to that for iterative MB reconstructions. However, for stronger inhomogeneities, iterative MB reconstruction outperforms iterative CPR in terms of signal intensity correction. Combining MB with CS, similar image and Delta B-0 map quality can be obtained without a scan time penalty. These findings were confirmed by experimental results. Discussion In case of B-0 inhomogeneities in the order of kHz, iterative MB reconstructions can help to improve both image quality and Delta B-0 map estimation.
引用
收藏
页码:631 / 642
页数:12
相关论文
共 19 条
  • [1] Image distortion correction for MRI in low field permanent magnet systems with strong B0 inhomogeneity and gradient field nonlinearities
    Kirsten Koolstra
    Thomas O’Reilly
    Peter Börnert
    Andrew Webb
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2021, 34 : 631 - 642
  • [2] Joint B0 and image estimation integrated with model based reconstruction for field map update and distortion correction in prostate diffusion MRI
    Usman, Muhammad
    Kakkar, Lebina
    Matakos, Antonis
    Kirkham, Alex
    Arridge, Simon
    Atkinson, David
    MAGNETIC RESONANCE IMAGING, 2020, 65 : 90 - 99
  • [3] A quadrature volume RF coil for vertical B0 field open MRI systems
    Tomanek, Boguslaw
    Volotovskyy, Vyacheslav
    Tyson, Randy
    Yin, Donghui
    Sharp, Jonathan
    Blasiak, Barbara
    CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2016, 46 (03) : 118 - 122
  • [4] Joint B0 and Image Reconstruction in Low-Field MRI by Physics-Informed Deep-Learning
    Schote, David
    Winter, Lukas
    Kolbitsch, Christoph
    Rose, Georg
    Speck, Oliver
    Kofler, Andreas
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2024, 71 (10) : 2842 - 2853
  • [5] A method for measuring B0 field inhomogeneity using quantitative double-echo in steady-state
    Barbieri, Marco
    Chaudhari, Akshay S.
    Moran, Catherine J.
    Gold, Garry E.
    Hargreaves, Brian A.
    Kogan, Feliks
    MAGNETIC RESONANCE IN MEDICINE, 2023, 89 (02) : 577 - 593
  • [6] Correction of Artifacts Induced by B0 Inhomogeneities in Breast MRI Using Reduced-Field-of-View Echo-Planar Imaging and Enhanced Reversed Polarity Gradient Method
    Rodriguez-Soto, Ana E.
    Fang, Lauren K.
    Holland, Dominic
    Zou, Jingjing
    Park, Helen H.
    Keenan, Kathryn E.
    Bartsch, Hauke
    Kuperman, Joshua
    Wallace, Anne M.
    Hahn, Michael
    Ojeda-Fournier, Haydee
    Dale, Anders M.
    Rakow-Penner, Rebecca
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2021, 53 (05) : 1581 - 1591
  • [7] Patient-to-patient variation of susceptibility-induced B0 field in bilateral breast MRI
    Lee, Seung-Kyun
    Hancu, Ileana
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2012, 36 (04) : 873 - 880
  • [8] Characterization of displacement forces and image artifacts in the presence of passive medical implants in low-field (<100 mT) permanent magnet-based MRI systems, and comparisons with clinical MRI systems
    Van Speybroeck, C. D. E.
    O'Reilly, T.
    Teeuwisse, W.
    Arnold, P. M.
    Webb, A. G.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 84 : 116 - 124
  • [9] Optimal echo times for multi-gradient echo-based B0 field-mapping
    Geiger, Yasmin
    Tal, Assaf
    NMR IN BIOMEDICINE, 2020, 33 (07)
  • [10] Image Distortion Correction of Flash Camera Systems Based on Field-of-View Division Model
    Ren, Yuhang
    Li, Binkang
    Zhou, Errui
    Yang, Shaohua
    Yan, Ming
    Guo, Ming'an
    Li, Gang
    Liu, Lu
    Wang, Jing
    Shi, Mingyue
    LASER & OPTOELECTRONICS PROGRESS, 2024, 61 (24)