A practical demonstration of improved technique factors in paediatric fluoroscopy

被引:13
作者
Fenner, JW
Morrison, GD
Kerry, J
West, N
机构
[1] Royal Hallamshire Hosp, Dept Med Phys & Clin Engn, Sheffield S10 2JF, S Yorkshire, England
[2] Cty Hosp, Dept Med Phys, Lincoln LN2 5QY, England
[3] Derbyshire Royal Infirm NHS Trust, Derby DE1 2QY, England
关键词
D O I
10.1259/bjr.75.895.750596
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Dose incurred with fluoroscopic procedures accounts for a significant proportion of medically induced diagnostic exposure. Children are particularly vulnerable and it is therefore important to minimize exposure where practicable, A recent theoretical study has highlighted the potential for X-ray equipment to produce significant dose savings during paediatric fluoroscopy without incurring loss of diagnostic image quality. This is achieved by hardening the beam with additional copper (Cu) filtration (similar to0.2 mm Cu) and biasing exposure factors towards low tube potential, high tube current output. In practice. this method will have limited applicability because the high powered and programmable generator characteristics required are not commonly available in installations used for paediatric imaging. However, we describe a simple experiment in which our clinical equipment was modified to approximate desired low dose performance by altering the filtration and automatic exposure control characteristics of ordinary clinical equipment in the Sheffield Children's Hospital. This enabled us to obtain significant savings in dose. We performed a comparative study (normal dose vs low dose) using water phantoms to simulate patient attenuation in the age range 0-15 years. The Leeds N2 contrast sensitivity phantom was used to provide a measure of image quality. Dosimetric Measurements recorded up to 40% reduction in dose rate with only marginal loss of image quality when 0.1-0.2 min Cu filtration was used with the modified settings. This is a strong indication that significant dose reduction is achievable on routine clinical equipment without compromising image quality. Such simple and cost effective methods of dose reduction should be considered for wider implementation.
引用
收藏
页码:596 / 602
页数:7
相关论文
共 14 条
  • [11] Radiation dose at small intestinal biopsies in children: Results of a national study
    Persliden, J
    Pettersson, HBL
    FalthMagnusson, K
    [J]. ACTA PAEDIATRICA, 1996, 85 (09) : 1042 - 1046
  • [12] PERSLIDEN J, 1993, ACTA RADIOL, V34, P92
  • [13] Radiation burden to paediatric patients due to micturating cystourethrography examinations in a Dutch children's hospital
    Schultz, FW
    Geleijns, J
    Holscher, HC
    Weststrate, J
    Zonderland, HM
    Zoetelief, J
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1999, 72 (860) : 763 - 772
  • [14] A search for improved technique factors in paediatric fluoroscopy
    Tapiovaara, MJ
    Sandborg, M
    Dance, DR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (02) : 537 - 559