Segmental and global left ventricular function assessment using gated SPECT with a semiconductor Cadmium Zinc Telluride (CZT) camera: Phantom study and clinical validation vs cardiac magnetic resonance

被引:19
作者
Bailliez, Alban [1 ,4 ]
Blaire, Tanguy [1 ,4 ]
Mouquet, Frederic [2 ]
Legghe, R. [1 ]
Etienne, B. [1 ]
Legallois, Damien [3 ,4 ]
Agostini, Denis [3 ,4 ]
Manrique, Alain [3 ,4 ]
机构
[1] Polyclin Bois, IRIS, Dept Nucl Med, F-59000 Lille, France
[2] Polyclin Bois, F-59000 Lille, France
[3] CHU Caen, Dept Nucl Med, F-14000 Caen, France
[4] Normandie Univ, EA4650, Caen, France
关键词
CZT; SPECT; myocardial perfusion imaging; solid-state cardiac camera; left ventricular function; wall thickening; wall motion; cardiac magnetic resonance; tetrofosmin; MYOCARDIAL-PERFUSION SPECT; EJECTION FRACTION; QUANTIFICATION; REPRODUCIBILITY; VOLUME; TECHNETIUM-99M-MIBI; TOMOGRAPHY; CARDIOLOGY; RADIATION; SESTAMIBI;
D O I
10.1007/s12350-014-9899-z
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We evaluated gated-SPECT using a Cadmium-Zinc-Telluride (CZT) camera for assessing global and regional left ventricular (LV) function. A phantom study evaluated the accuracy of wall thickening assessment using systolic count increase on both Anger and CZT (Discovery 530NMc) cameras. The refillable phantom simulated variable myocardial wall thicknesses. The apparent count increase (%CI) was compared to the thickness increase (%Th). CZT gated-SPECT was compared to cardiac magnetic resonance (CMR) in 27 patients. Global and regional LV function (wall thickening and motion) were quantified and compared between SPECT and CMR data. In the phantom study using a 5-mm object, the regression between %CI and %Th was significantly closer to the line of identity (y = x) with the CZT (R (2) = 0.9955) than the Anger (R (2) = 0.9995, P = .03). There was a weaker correlation for larger objects (P = .003). In patients, there was a high concordance between CZT and CMR for ESV, EDV, and LVEF (all CCC > 0.80, P < .001). CZT underestimated %CI and wall motion (WM) compared to CMR (P < .001). The agreement to CMR was better for WM than wall thickening. The Discovery 530NMc provided accurate measurements of global LV function but underestimated regional wall thickening, especially in patients with increased wall thickness.
引用
收藏
页码:712 / 722
页数:11
相关论文
共 29 条
  • [1] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [2] A fast cardiac gamma camera with dynamic SPECT capabilities: design, system validation and future potential
    Bocher, Moshe
    Blevis, Ira M.
    Tsukerman, Leonid
    Shrem, Yigal
    Kovalski, Gil
    Volokh, Lana
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 (10) : 1887 - 1902
  • [3] Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart - A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association
    Cerqueira, MD
    Weissman, NJ
    Dilsizian, V
    Jacobs, AK
    Kaul, S
    Laskey, WK
    Pennell, DJ
    Rumberger, JA
    Ryan, T
    Verani, MS
    [J]. CIRCULATION, 2002, 105 (04) : 539 - 542
  • [4] Optimal reproducibility of gated sestamibi and thallium myocardial perfusion study left ventricular ejection fractions obtained on a solid-state CZT cardiac camera requires operator input
    Cherk, Martin H.
    Ky, Jason
    Yap, Kenneth S. K.
    Campbell, Patrina
    McGrath, Catherine
    Bailey, Michael
    Kalff, Victor
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2012, 19 (04) : 713 - 718
  • [5] Absolute Quantification of Left Ventricular Global and Regional Function at Nuclear MPI Using Ultrafast CZT SPECT: Initial Validation Versus Cardiac MR
    Cochet, Hubert
    Bullier, Eric
    Gerbaud, Edouard
    Durieux, Muriel
    Godbert, Yann
    Lederlin, Mathieu
    Coste, Pierre
    Barat, Jean-Louis
    Laurent, Francois
    Montaudon, Michel
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2013, 54 (04) : 556 - 563
  • [6] Visual assessment of left ventricular perfusion and function with electrocardiography-gated SPECT has high intraobserver and interobserver reproducibility among experienced nuclear cardiologists and cardiology trainees
    Danias, PG
    Ahlberg, AW
    Travin, MI
    Mahr, NC
    Abreu, JE
    Marini, D
    Mann, A
    Mather, JF
    Boden, WE
    Heller, GV
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2002, 9 (03) : 263 - 270
  • [7] Progress in the Development of CdTe and CdZnTe Semiconductor Radiation Detectors for Astrophysical and Medical Applications
    Del Sordo, Stefano
    Abbene, Leonardo
    Caroli, Ezio
    Mancini, Anna Maria
    Zappettini, Andrea
    Ubertini, Pietro
    [J]. SENSORS, 2009, 9 (05) : 3491 - 3526
  • [8] Reduced isotope dose and imaging time with a high-efficiency CZT SPECT camera
    Duvall, W. Lane
    Croft, Lori B.
    Ginsberg, Eric S.
    Einstein, Andrew J.
    Guma, Krista A.
    George, Titus
    Henzlova, Milena J.
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2011, 18 (05) : 847 - 857
  • [9] Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: multicenter comparison with standard dual detector cameras
    Esteves, Fabio P.
    Raggi, Paolo
    Folks, Russell D.
    Keidar, Zohar
    Askew, J. Wells
    Rispler, Shmuel
    O'Connor, Michael K.
    Verdes, Liudmilla
    Garcia, Ernest V.
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2009, 16 (06) : 927 - 934
  • [10] Exposure to Low-Dose Ionizing Radiation from Medical Imaging Procedures.
    Fazel, Reza
    Krumholz, Harlan M.
    Wang, Yongfei
    Ross, Joseph S.
    Chen, Jersey
    Ting, Henry H.
    Shah, Nilay D.
    Nasir, Khurram
    Einstein, Andrew J.
    Nallamothu, Brahmajee K.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2009, 361 (09) : 849 - 857