Radiation dose in neuroangiography using image noise reduction technology: a population study based on 614 patients

被引:74
作者
Soderman, Michael [1 ,2 ]
Mauti, Maria [3 ]
Boon, Sjirk [3 ]
Omar, Artur [4 ]
Marteinsdottir, Maria [4 ]
Andersson, Tommy [1 ,2 ]
Holmin, Staffan [1 ,2 ]
Hoornaert, Bart [3 ]
机构
[1] Karolinska Univ Hosp Solna, Karolinska Inst, Dept Clin Neurosci, S-17176 Stockholm, Sweden
[2] Karolinska Univ Hosp Solna, Dept Neuroradiol, S-17176 Stockholm, Sweden
[3] Philips Healthcare, Best, Netherlands
[4] Karolinska Univ Hosp, Sect Imaging Phys, Dept Med Phys, Stockholm, Sweden
关键词
Angiography; Radiation dose; Radiation physics; Imaging technology; Interventional neuroradiology; INTERVENTIONAL NEURORADIOLOGY; EXPOSURE;
D O I
10.1007/s00234-013-1276-0
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The purpose of this study was to quantify the reduction in patient radiation dose by X-ray imaging technology using image noise reduction and system settings for neuroangiography and to assess its impact on the working habits of the physician. Radiation dose data from 190 neuroangiographies and 112 interventional neuroprocedures performed with state-of-the-art image processing and reference system settings were collected for the period January-June 2010. The system was then configured with extra image noise reduction algorithms and system settings, which enabled radiation dose reduction without loss of image quality. Radiation dose data from 174 neuroangiographies and 138 interventional neuroprocedures were collected for the period January-June 2012. Procedures were classified as diagnostic or interventional. Patient radiation exposure was quantified using cumulative dose area product and cumulative air kerma. Impact on working habits of the physician was quantified using fluoroscopy time and number of digital subtraction angiography (DSA) images. The optimized system settings provided significant reduction in dose indicators versus reference system settings (p < 0.001): from 124 to 47 Gy cm(2) and from 0.78 to 0.27 Gy for neuroangiography, and from 328 to 109 Gy cm(2) and from 2.71 to 0.89 Gy for interventional neuroradiology. Differences were not significant between the two systems with regard to fluoroscopy time or number of DSA images. X-ray imaging technology using an image noise reduction algorithm and system settings provided approximately 60% radiation dose reduction in neuroangiography and interventional neuroradiology, without affecting the working habits of the physician.
引用
收藏
页码:1365 / 1372
页数:8
相关论文
共 14 条
  • [1] Patient radiation exposure during diagnostic and therapeutic interventional neuroradiology procedures
    Alexander, M. D.
    Oliff, M. C.
    Olorunsola, O. G.
    Brus-Ramer, M.
    Nickoloff, E. L.
    Meyers, P. M.
    [J]. JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2010, 2 (01) : 6 - 10
  • [2] Patient and staff doses in interventional neuroradiology
    Bor, D
    Çekirge, S
    Türkay, T
    Turan, O
    Gülay, M
    Önal, E
    Çil, B
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 117 (1-3) : 62 - 68
  • [3] D'Ercole L, 2007, AM J NEURORADIOL, V28, P503
  • [4] Proposed local diagnostic reference levels in angiography and interventional neuroradiology and a preliminary analysis according to the complexity of the procedures
    D'Ercole, Loredana
    Thyrion, Federico Zappoli
    Bocchiola, Milena
    Mantovani, Laura
    Klersy, Catherine
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (01): : 61 - 70
  • [5] Adult patient doses in interventional neuroradiology
    Gkanatsios, NA
    Huda, W
    Peters, KR
    [J]. MEDICAL PHYSICS, 2002, 29 (05) : 717 - 723
  • [6] Radiation doses in interventional radiology procedures: The RAD-IR Study - Part I: Overall measures of dose
    Miller, DL
    Balter, S
    Cole, PE
    Lu, HT
    Schueler, BA
    Geisinger, M
    Berenstein, A
    Albert, R
    Georgia, JD
    Noonan, PT
    Cardella, JF
    George, JS
    Russell, EJ
    Malisch, TW
    Vogelzang, RL
    Miller, GL
    Anderson, J
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (06) : 711 - 727
  • [7] The potential for radiation-induced skin damage in interventional neuroradiological procedures: A review of 522 cases using automated dosimetry
    O'Dea, TJ
    Geise, RA
    Ritenour, ER
    [J]. MEDICAL PHYSICS, 1999, 26 (09) : 2027 - 2033
  • [8] Radiation Exposure in Vascular Angiographic Procedures
    Pitton, Michael Bernhard
    Kloeckner, Roman
    Schneider, Jens
    Ruckes, Christian
    Bersch, Anton
    Dueber, Christoph
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2012, 23 (11) : 1487 - 1495
  • [9] Studies of patient doses in interventional radiological examinations
    Sarycheva, Svetlana
    Golikov, Vladislav
    Kalnicky, Sergey
    [J]. RADIATION PROTECTION DOSIMETRY, 2010, 139 (1-3) : 258 - 261
  • [10] Image Noise Reduction Algorithm for Digital Subtraction Angiography: Clinical Results
    Soderman, Michael
    Holmin, Staffan
    Andersson, Tommy
    Palmgren, Charlotta
    Babic, Drazenko
    Hoornaert, Bart
    [J]. RADIOLOGY, 2013, 269 (02) : 553 - 560