Time dependent pre-treatment EPID dosimetry for standard and FFF VMAT

被引:44
作者
Podesta, Mark [1 ]
Nijsten, Sebastiaan M. J. J. G. [1 ]
Persoon, Lucas C. G. G. [1 ]
Scheib, Stefan G. [2 ]
Baltes, Christof [2 ]
Verhaegen, Frank [1 ,3 ]
机构
[1] MUMC, MAASTRO, Dept Radiat Oncol, Maastricht, Netherlands
[2] Varian Med Syst iLab GmbH, Res & Dev, Dattwil Baden, Switzerland
[3] McGill Univ, Dept Oncol, Med Phys Unit, Montreal, PQ, Canada
关键词
VMAT; time dependent; EPID dosimetry; control point; VOLUMETRIC-MODULATED ARC; RESPONSE CHARACTERISTICS; LAG CORRECTION; PORTAL IMAGES; VERIFICATION; IMRT; THERAPY; BEAM; QA; DECONVOLUTION;
D O I
10.1088/0031-9155/59/16/4749
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Methods to calibrate Megavoltage electronic portal imaging devices (EPIDs) for dosimetry have been previously documented for dynamic treatments such as intensity modulated radiotherapy (IMRT) using flattened beams and typically using integrated fields. While these methods verify the accumulated field shape and dose, the dose rate and differential fields remain unverified. The aim of this work is to provide an accurate calibration model for time dependent pre-treatment dose verification using amorphous silicon (a-Si) EPIDs in volumetric modulated arc therapy (VMAT) for both flattened and flattening filter free (FFF) beams. A general calibration model was created using a Varian TrueBeam accelerator, equipped with an aS1000 EPID, for each photon spectrum 6 MV, 10 MV, 6 MV-FFF, 10 MV-FFF. As planned VMAT treatments use control points (CPs) for optimization, measured images are separated into corresponding time intervals for direct comparison with predictions. The accuracy of the calibration model was determined for a range of treatment conditions. Measured and predicted CP dose images were compared using a time dependent gamma evaluation using criteria (3%, 3 mm, 0.5 sec). Time dependent pre-treatment dose verification is possible without an additional measurement device or phantom, using the on-board EPID. Sufficient data is present in trajectory log files and EPID frame headers to reliably synchronize and resample portal images. For the VMAT plans tested, significantly more deviation is observed when analysed in a time dependent manner for FFF and non-FFF plans than when analysed using only the integrated field. We show EPID-based pre-treatment dose verification can be performed on a CP basis for VMAT plans. This model can measure pre-treatment doses for both flattened and unflattened beams in a time dependent manner which highlights deviations that are missed in integrated field verifications.
引用
收藏
页码:4749 / 4768
页数:20
相关论文
共 42 条
[1]   Using an EPID for patient-specific VMAT quality assurance [J].
Bakhtiari, M. ;
Kumaraswamy, L. ;
Bailey, D. W. ;
de Boer, S. ;
Malhotra, H. K. ;
Podgorsak, M. B. .
MEDICAL PHYSICS, 2011, 38 (03) :1366-1373
[2]  
Bell Labs, 1971, UN PROGR MAN UN TIM
[3]   Implementation of EPID transit dosimetry based on a through-air dosimetry algorithm [J].
Berry, Sean L. ;
Sheu, Ren-Dih ;
Polvorosa, Cynthia S. ;
Wuu, Cheng-Shie .
MEDICAL PHYSICS, 2012, 39 (01) :87-98
[4]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[5]   Epid cine acquisition mode for in vivo dosimetry in dynamic are radiation therapy [J].
Fidanzio, Andrea ;
Mameli, Alessandra ;
Placidi, Elisa ;
Greco, Francesca ;
Stimato, Gerardina ;
Gaudino, Diego ;
Rarnella, Sara ;
D'Angelillo, Rolando ;
Cellini, Francesco ;
Trodella, Lucio ;
Cilla, Savino ;
Grimaldi, Luca ;
D'Onofrio, Guido ;
Azario, Luigi ;
Piermattei, Angelo .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2008, 266 (04) :658-666
[6]   Volumetric modulated arc therapy is superior to conventional intensity modulated radiotherapy - a comparison among prostate cancer patients treated in an Australian centre [J].
Fogarty, Gerald B. ;
Ng, Diana ;
Liu, Guilin ;
Haydu, Lauren E. ;
Bhandari, Nastik .
RADIATION ONCOLOGY, 2011, 6
[7]   Current status and future perspective of flattening filter free photon beams [J].
Georg, Dietmar ;
Knoos, Tommy ;
McClean, Brendan .
MEDICAL PHYSICS, 2011, 38 (03) :1280-1293
[8]   Effect of statistical fluctuation in Monte Carlo based photon beam dose calculation on gamma index evaluation [J].
Graves, Yan Jiang ;
Jia, Xun ;
Jiang, Steve B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (06) :1839-1853
[9]   Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry [J].
Greer, PB .
MEDICAL PHYSICS, 2005, 32 (12) :3558-3568
[10]   A recursive correction algorithm for detector decay characteristics in CT [J].
Hsieh, J ;
Gurmen, OE ;
King, KF .
MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 :298-305