A laboratory model to evaluate cutout resistance of implants for pertrochanteric fracture fixation

被引:232
作者
Sommers, MB
Roth, C
Hall, H
Kam, BCC
Ehmke, LW
Krieg, JC
Madey, SM
Bottlang, M
机构
[1] Legacy Clin Res & Technol Ctr, Biomech Lab, Portland, OR 97232 USA
[2] Synthes USA, W Chester, PA USA
[3] Oregon Hlth & Sci Univ, Dept Orthopaed & Rehabil, Portland, OR USA
关键词
hip; cutout model; pertrochanteric fracture; lag screw; fixation;
D O I
10.1097/00005131-200407000-00006
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To establish a laboratory model of implant cutout, which can evaluate the effect of implant design on cutout resistance in a clinically realistic "worst case" scenario. Setting: Orthopaedic biomechanics laboratory. Design: Implant cutout was simulated in an unstable pertrochanteric fracture model, which accounted for dynamic loading, osteoporotic bone, and a defined implant offset. For model characterization, lag screw cutout was simulated in human cadaveric specimens and in polyurethane foam surrogates. Subsequently, foam surrogates were used to determine differences in cutout resistance between 2 common lag screws (dynamic hip screw, Gamma) and 2 novel blade-type implant designs (dynamic helical hip system, trochanteric fixation nail). Main Outcome Measures: Implant migration was continuously recorded with a spatial motion tracking system as a function of the applied loading cycles. In addition, the total number of loading cycles to cutout failure was determined for specific load amplitudes. Results: Implant migration in polyurethane surrogates closely correlated with that in cadaveric specimens, but yielded higher reproducibility and consistent cutout failure. The cutout model was able to delineate significant differences in cutout resistance between specific implant designs. At any of 4 load amplitudes (0.8 kN, 1.0 kN, 1.2 kN, 1.4 kN) dynamic hip screw lag screws failed earliest. The gamma nail lag screw could sustain significantly more loading cycles than the dynamic hip screw. Of all implants, trochanteric fixation nail implants demonstrated the highest cutout resistance. Conclusions: Implant design can significantly affect the fixation strength and cutout resistance of implants for pertrochanteric fracture fixation. The novel cutout model can predict differences in cutout resistance between distinct implant designs.
引用
收藏
页码:361 / 368
页数:8
相关论文
共 29 条
[11]   Internal fixation of femoral neck fractures with posterior comminution: A biomechanical study [J].
Kauffman, JI ;
Simon, JA ;
Kummer, FJ ;
Pearlman, CJ ;
Zuckerman, JD ;
Koval, KJ .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1999, 13 (03) :155-159
[12]   Cutting-out of the lag screw after internal fixation with the Asiatic gamma nail [J].
Kawaguchi, S ;
Sawada, K ;
Nabeta, Y .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1998, 29 (01) :47-53
[13]   Femoral neck fracture after removal of the standard gamma interlocking nail: a cadaveric study to determine factors influencing the biomechanical properties of the proximal femur [J].
Kukla, C ;
Pichl, W ;
Prokesch, R ;
Jacyniak, W ;
Heinze, G ;
Gatterer, R ;
Heinz, T .
JOURNAL OF BIOMECHANICS, 2001, 34 (12) :1519-1526
[14]   BIOMECHANICAL ANALYSIS OF THE SLIDING CHARACTERISTICS OF COMPRESSION HIP SCREWS [J].
KYLE, RF ;
WRIGHT, TM ;
BURSTEIN, AH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1980, 62 (08) :1308-1314
[15]   PREVENTION OF FRACTURE AT THE DISTAL LOCKING SITE OF THE GAMMA-NAIL [J].
LACROIX, H ;
ARWERT, H ;
SNIJDERS, CJ ;
FONTIJNE, WPJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1995, 77B (02) :274-276
[16]  
LAROS GS, 1980, ORTHOP CLIN N AM, V11, P525
[17]   FIXATION OF UNSTABLE TROCHANTERIC HIP-FRACTURES - A CADAVER STUDY COMPARING 3 DIFFERENT DEVICES [J].
LARSSON, S ;
ELLOY, M ;
HANSSON, LI .
ACTA ORTHOPAEDICA SCANDINAVICA, 1988, 59 (06) :658-663
[18]   Dynamic hip screw with trochanteric stabilizing plate in the treatment of unstable proximal femoral fractures: A comparative study with the Gamma nail and compression hip screw [J].
Madsen, JE ;
Naess, L ;
Aune, AK ;
Alho, A ;
Ekeland, A ;
Stromsoe, K .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1998, 12 (04) :241-248
[19]   IMPLANT FAILURES IN PATIENTS WITH PROXIMAL FRACTURES OF THE FEMUR TREATED WITH A SLIDING SCREW DEVICE [J].
MAINDS, CC ;
NEWMAN, RJ .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1989, 20 (02) :98-100
[20]  
Meislin R J, 1990, J Orthop Trauma, V4, P130, DOI 10.1097/00005131-199004020-00005